Description
1. Product Presentation
The Bowie-Dick Helix test pack (Ref. BD125X/20) is the process control solution that sterilisation centres need to quickly and reliably verify the correct removal of air and penetration of saturated steam in vacuum-assisted autoclaves, both large and small. Each box includes 20 single-use units, ready to be incorporated into the daily validation routine of the sterilisation service, hospitals and medical device processing companies.
Manufactured by Pergut Medical and distributed by Cahnos in Spain, this device combines a metal capsule, a polyamide tube of 1.5 m x 2 mm lumen with a protective silicone connector and a self-adhesive paper indicator strip of 6 x 76 mm, reproducing the most challenging hollow load condition according to the guidelines of the ISO 11140-4 standard.
2. Technical explanation of the procedure
The Bowie-Dick test is a steam penetration test used to monitor the air removal efficiency of steam sterilisers with dynamic air extraction (pre-vacuum and vacuum pump). If residual air pockets remain in the chamber, the steam does not reach the load surface under saturation conditions and the sterilisation process may not be effective, even if the recorded temperature and time parameters are correct.
The Helix design reproduces the worst-case hollow load scenario: the steam must travel through a polyamide lumen 1.5 m long with only 2 mm internal diameter before reaching the indicator strip. The metal capsule protects the reading area and the silicone connector ensures a seal and a secure connection to the equipment. The self-adhesive paper strip of 6 x 76 mm, printed with steam-sensitive ink, changes colour only when the air has been removed and the saturated steam has penetrated under correct conditions, allowing the load to be released safely.
This device is classified as a chemical indicator Type 2 (ISO 11140-1) and its use is recommended for the Bowie-Dick test according to the ISO 11140-4 standard.
3. Technical Specifications
Device technical data sheet
| Characteristic | Specification |
| Type | B&D Helix test for steam |
| Reference | BD125X/20 |
| Device dimensions | 1.5 m length x 2 mm lumen |
| Indicator strip | Self-adhesive paper 6 x 76 mm |
| Composition | Metal capsule, polyamide tube with protective silicone connector and self-adhesive paper strip |
| Presentation | Box of 20 units |
| Processing | Steam in large and small vacuum-assisted sterilisation equipment |
| Application | Assessment of air removal and steam penetration |
| Classification | Chemical indicator Type 2 / Type S2 (ISO 11140-1) |
| Storage | No special conditions required; keep in original packaging until use |
| Use | Single-use, disposable |
| Conformity | ISO 11140-1 (Type S2) and ISO 11140-4 |
| CE marking | No (product conforms to the ISO 11140-1/-4 standards) |
Typical Bowie-Dick cycle conditions
| Parameter | Value |
| Reference pre-vacuum cycle | 134 °C for 3.5 minutes |
| Alternative cycle | 132 °C for 4 minutes |
| Parameter evaluated | Air removal and saturated steam penetration |
| Conforming result | Uniform and complete colour change of the indicator strip |
| Non-conforming result | Insufficient or irregular strip change: check vacuum system and process |
4. Equivalences and Cross-References
This pack belongs to the family of Helix process challenge devices from the Spanish manufacturer Pergut Medical. The manufacturer’s technical reference whose dimensions and composition match this device is KH2X15-35B (B&D Helix test for steam), conforming to ISO 11140-1 Type S2 and ISO 11140-4. Within the same family, the following references of interest are found:
| Reference | Description | Presentation |
| KH2X15-35B | B&D Helix test for steam (technical equivalent of this product) | On request |
| KH2X15-35 | Helix Test PCD kit + simulation strips for steam | 250 strips/box |
| BD125X/1 | Bowie-Dick test package with pre-notice sheet | 20 uds/caja |
| 24102 | Bowie-Dick test package for daily control (120 x 120 x 17 mm) | 20 uds/caja |
| KH2X12-P1 | Helix Test PCD Kit + strips for plasma | 100 uds/caja |
| KH2X15F1P | Helix Test PCD Kit + strips for formaldehyde | 250 uds/caja |
For hollow load validation purposes, this device is functionally interchangeable with other Helix-type tests conforming to ISO 11140-4, as it exactly meets the 1.5 m x 2 mm lumen geometry required by the market standard for this type of process control.
5. Selection tips and variants of interest
- Hollow load control: if your autoclave processes long-lumen instruments (endoscopy, laparoscopy, cannulas, etc.), the Helix-type test is the right one. For daily monitoring of large autoclaves with a vacuum pump, the classic reference is the Bowie-Dick pack 24102.
- Frequency of use: perform the B&D test at least once a day, with the chamber empty and following the equipment manufacturer’s protocol and the ISO 11140-1 standard.
- Other processes: if you need to validate plasma (hydrogen peroxide vapour) or formaldehyde cycles, the same family offers versions KH2X12-P1 and KH2X15F1P.
- Complementary consumables: for a complete set, combine this pack with B&D simulation strips and steam biological indicators, available in the Cahnos catalogue.
- Storage and handling: keep the product in its original packaging until the moment of use; it is single-use and must not be re-sterilised. Always check the integrity of the packaging before each cycle.
- Professional B2B purchasing: for sterilisation centres, hospitals and processing companies, contact our technical-commercial team for volume and continuous supply conditions.