Description
1. Commercial Introduction
The flat mixed sterilisation reel of 200 m x 100 mm is the reference packaging solution for sterilisation departments, operating theatres, dental clinics and practices that need to make custom-sized pouches for instruments with maximum safety and minimal material waste. Its roll format allows cutting, inserting and heat-sealing the packaging to the exact length of each device, optimising the cost per sterilisation cycle. It is a Class I medical device, manufactured with certified high-quality materials and supplied in a box of 6 units.
2. Technical explanation
The sterile barrier system consists of two heat-sealed faces: one of high-grammage medical paper (60 g/m²) that allows the penetration and evacuation of the sterilising agent, and another of a double layer of transparent polyethylene/polypropylene film (55 g/m²) that provides mechanical strength and visibility of the contents without opening the packaging. The assembly is designed in compliance with Regulation (EU) 2017/745 (MDR) to offer high antibacterial protection in medical sterilisation. The reel is compatible with steam sterilisation (autoclave) and ethylene oxide (ETO), and its process indicators show a clear colour change after exposure to the sterilising agent, facilitating verification of each cycle. It guarantees constant sealing during sterilisation, extremely easy opening without fibre release or tearing, and proper maintenance of sterility of the contents up to the point of use.
3. Technical Specifications
3.1. Process indicators
| Sterilisation method | Indicator colour change |
|---|---|
| Steam (autoclave) | Light blue to green |
| Ethylene oxide (ETO) | Pink to orange |
| Formaldehyde | Green to yellow |
3.2. General characteristics
| Parameter | Specification |
|---|---|
| Format | Flat mixed reel for sterilisation |
| External dimensions | 200 m x 100 mm |
| Length | 200 m |
| Width | 100 mm |
| Paper face | High-grammage medical paper, 60 g/m² |
| Film face | Double layer of polyethylene/polypropylene film, 55 g/m² |
| Sterilisation methods | Steam (autoclave) and ethylene oxide (ETO) |
| Sealing | Constant during sterilisation |
| Aperture | Easy, no fibre release or tearing |
| Sterility maintenance | Correct (sterile barrier up to the point of use) |
| Protection | High antibacterial protection |
| Regulatory standard | MDR (EU) 2017/745 |
| Product type | Medical device |
| Device class | I |
| Presentation | Box of 6 units |
| Manufacturer reference | 36002 |
| SKU Cahnos | GM36002 |
3.3. Product data
| Manufacturer | GIMA |
|---|---|
| Sales unit | Box of 6 units |
| Device type | Medical device |
| Device class | I |
| CND | S01020101 |
| GMDN | 13735 |
4. Equivalences and Cross-References
Reference 36002 (SKU Cahnos GM36002) corresponds to the manufacturer’s official factory code, without modifications. There is no official equivalence table validated by the manufacturer that cross-references this reel with other brands on a 1:1 basis; the following comparisons are functional, at category level, and not interchangeable without technical validation. Due to its construction of 60 g/m² medical paper and transparent film with indicators for steam, ethylene oxide and formaldehyde, this reel is comparable to the flat mixed rolls BOP TI from the G2T series available at Cahnos (e.g. G2T075X200, G2T100X200 and G2T150X200), which share the same general clinical indication and peelable packaging composition. The Esterivec mixed reels with Tyvek® base (low-temperature range for plasma and hydrogen peroxide) serve a different purpose and are not equivalent for the steam and ETO cycles specific to this reel. Before any substitution, each field must be checked: actual roll width, grammage and paper type, film composition, number and type of indicators, peelability, sealing and conformity regulations; consult the technical department before use in tenders or stock replacement.
5. Selection tips and variants of interest
Always select a reel width slightly larger than the device to be packaged, leaving a minimum margin for the seals; the 100 mm format is ideal for forceps, scissors, scalpels, clamps and small instrument sets. Seal with an approved heat sealer, respecting the temperature, pressure and time indicated by the equipment manufacturer, and check the integrity of the seal after each cycle. In the steriliser, arrange the pouches plastic-to-plastic and paper-to-paper. Verify the colour change of the process indicator (steam: light blue to green; ETO: pink to orange; formaldehyde: green to yellow) before considering the material suitable, and record batch, date and contents on the labelling. Store the rolls in a cool, dry place, in their original packaging and away from sterilising agents until use. This is a single-use packaging: do not re-sterilise. For institutional volumes, framework agreements or tenders, contact the technical-commercial department of Cahnos for special B2B conditions and delivery times.





Reviews
There are no reviews yet.