Description
1. Introduction
The turbine transducer kit (ref. 07053 / 09890) is the spare part set that keeps the Sibelmed Datospir Touch desktop spirometer operational in its turbine configuration, the most efficient for practices with high patient turnover. It groups the handle, transducer bag and turbine into a single purchase, so you can replace the complete measurement circuit of your device without having to manage multiple references or rely on calibration adjustments: a ready-to-use solution for performing spirometry tests with the accuracy and repeatability required in daily clinical practice.
2. Technical explanation of the equipment and procedure
The turbine transducer measures airflow directly: the air that the patient inhales and exhales rotates a low-inertia turbine, and the device calculates flow and volume from the turbine’s rotation speed. This technology offers a very stable response, minimal flow resistance (below 0.7 cmH2O/l/s at 14 l/s, the lowest of the three Datospir Touch configurations) and a very tight operating cost, as it does not require precision mouthpieces or additional consumables beyond the bacterial/viral filter and the disposable mouthpiece.
In the Datospir Touch spirometer, the transducer is attached to the device via the handle and transducer bag included in this kit. The patient blows through a disposable bacterial/viral filter and a mouthpiece connected to the turbine, and the device records flow-volume (F/V) and volume-time (V/T) curves, calculating more than 40 spirometric parameters (FVC, FEV1, PEF, FEF25-75 %, MVV, etc.). In accordance with the recommendations of the ERS and SEPAR societies, lung function tests should always be performed with a disposable bacterial and viral filter between patient and transducer to avoid cross-contamination; therefore, the reusable turbine must be disinfected according to the device’s instruction manual and protected with a filter for each patient.
3. Technical Specifications
| Parameter | Specification |
| Reference / SKU | 07053 (distributor SKU) / 09890 (manufacturer’s official code for the turbine transducer module) |
| Product type | Turbine transducer kit for Datospir Touch spirometer |
| Kit contents | Handle + transducer bag + turbine, integrated in a single pack |
| Compatibility | Spirometer Datospir Touch (Sibelmed) in turbine configurations (Easy-T and Diagnostic-T) |
| Measurement technology | Turbine (direct flow measurement by rotation) |
| Flow range (BTPS) | 0 to ±16 l/s |
| Volume range (BTPS) | 0 to 10 l |
| Dynamic resistance at 14 l/s | <0.7 cmH2O/l/s (turbine) |
| Volume accuracy (BTPS) | ±2.5 % or 50 ml (whichever is greater) |
| Flow accuracy (BTPS) | ±3 % or 150 ml/s (whichever is greater) |
| PEF accuracy (BTPS) | ±10 % or 170 ml/s |
| Temporal accuracy | 0.50 % |
| Volume resolution | <10 ml |
| Sampling frequency | 100 Hz |
| Use | Reusable transducer; requires disposable bacterial/viral filter between patients |
| Recommended filter | 09698 (28 mm bacterial/viral filter, box of 100 units) |
| Regulatory standard | Medical device compliant with Regulation (EU) 2017/745 (MDR); EN ISO 9001:2015 and EN ISO 13485:2016+A11:2021 |
| Sales unit | Per unit (1 kit) |
4. Equivalences and Cross-References
The code 09890 corresponds to the official Sibelmed turbine transducer module for the Datospir Touch family, verified in the official technical data sheet of the spirometer published by the manufacturer; reference 07053 is the distributor code under which Cahnos markets this kit (handle + bag + turbine). The turbine transducer is a specific component of the measurement circuit of each spirometer: although other manufacturers use digital turbines with the same physical principle (MIR with its FlowMIR turbine in Spirobank and Spirolab, COSMED in Pony FX, or ndd’s ultrasonic measurement in EasyOne), the coupling, connection electronics and consumables are specific to each device, so they are not interchangeable between brands. Within the Datospir Touch family itself, the following official manufacturer references have been verified:
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
| Sibelmed | Fleisch transducer module Datospir Touch (spare head) | 09888 | Comparable by category (not interchangeable 1:1) | Same coupling and Datospir Touch family, but Fleisch differential pressure technology (capillary mesh) with dynamic resistance <1.4 cmH2O/l/s at 14 l/s; requires bacterial/viral filter and disinfection of the circuit. Code verified in the official Datospir Touch technical data sheet. |
| Sibelmed | Disposable transducer module Datospir Touch (spare head) | 09889 | Comparable by category (not interchangeable 1:1) | Same connector and Datospir Touch family, but disposable Lilly differential-pressure principle with dynamic resistance <1.2 cmH2O/l/s at 14 l/s; maximum hygiene without reprocessing. Code verified in the official technical data sheet of the Datospir Touch. |
There is no equivalence table validated by the manufacturer that cross-references this kit with codes from other brands on a 1:1 interchangeable basis; consult the technical department before using in a tender. Within the same Sibelmed family, the official transducer modules for the Datospir Touch are 09888 (Fleisch), 09889 (disposable) and 09890 (turbine, included in this kit); the official bacterial/viral filter is 09698 (28 mm, box of 100 units) and the compatible cardboard mouthpiece for Datospir-Sibelmed is 01555 (28 ext. × 60 mm).
5. Selection tips and variants of interest
Choose the transducer according to your centre’s profile: the turbine in this kit is the option with the lowest flow resistance and lowest running cost, highly recommended for primary care, occupational medicine and high-turnover practices; if you are looking for maximum accuracy for pulmonary function laboratories, the Datospir Touch range also supports the Fleisch transducer (kit ref. 07052) and, for maximum hygiene without disinfection, the disposable Lilly 2-in-1 transducer (kit ref. 07046). In the most recent Datospir Touch models, the turbine configuration is identified as Easy-T (code 09879) or Diagnostic-T (code 09882); confirm the version of your equipment before placing the order.
Remember to replace the bacterial/viral filter for each patient (ref. 09698), keep compatible cardboard mouthpieces (ref. 01555) and a nasal clip (ref. 02692 or 09661) available. For transducer quality control, use the 3-litre calibration syringe (ref. 08290) according to ATS/ERS recommendations, and for paediatric patients attach the paediatric mouthpiece adapter (ref. 01149). The reusable transducer must be disinfected following the equipment’s instruction manual; check availability and request your quote without obligation at Cahnos.