Description
1. Commercial Introduction
The Fleisch transducer kit (ref. 07052) is the replacement set that keeps the Datospir Touch by Sibelmed bench spirometers operational in Fleisch configuration, the most demanding in accuracy and repeatability and the preferred one in pulmonary function units and respiratory diagnostic laboratories. It brings together in a single purchase the entire measurement circuit of the device: handle, transducer bag, Fleisch flow transducer and bacterial filter. The included transducer corresponds to the manufacturer’s official Fleisch transducer module (Code 09888, verified in the Datospir Touch technical data sheet FG1102_ES), so the kit integrates without modifications into the original device and restores the factory performance of your spirometer.
2. Technical explanation of the equipment and procedure
The Fleisch transducer is a differential pressure pneumotachograph: the air exhaled by the patient passes through a superimposed capillary mesh and the device calculates the flow from the pressure difference generated between both ends of the mesh. It is one of the most accurate and repeatable flow measurement technologies on the market, which is why it is routinely used in pulmonary function units and respiratory medicine departments.
On the Datospir Touch spirometer, the transducer is attached to the device through the handle and the transducer bag included in this kit. The patient blows through a mouthpiece (or a disposable bacterial/viral filter) connected to the transducer, and the device records the flow-volume and volume-time curves and calculates more than 40 spirometric parameters (FVC, FEV1, PEF, FEF25-75 %, etc.). In accordance with the recommendations of the ERS and SEPAR societies, pulmonary function tests should always be performed with a disposable bacterial and viral filter between patient and transducer to prevent cross-contamination; this is why this kit includes the filter, and the reusable transducer must be disinfected following the device’s instruction manual.
3. Technical Specifications
| Reference / SKU | 07052 |
| Product type | Fleisch flow transducer kit for Datospir Touch spirometer |
| Kit contents | Handle, Datospir Touch transducer bag, Fleisch transducer and bacterial filter |
| Compatibility | Datospir Touch spirometer (Sibelmed) in configurations with Fleisch transducer (Easy-F and Diagnostic-F) |
| Measurement technology | Fleisch differential pressure pneumotachograph (capillary mesh) |
| Flow range (BTPS) | 0 to ±16 l/s |
| Volume range (BTPS) | 0 to 10 l |
| Dynamic resistance at 14 l/s | <1.4 cmH2O/l/s |
| Volume accuracy | ±2.5 % or 50 ml (whichever is greater) |
| Flow accuracy | ±3 % or 150 ml/s (whichever is greater) |
| PEF accuracy | ±10 % or 170 ml/s |
| Sampling frequency | 100 Hz |
| Use | Reusable transducer; requires disposable bacterial/viral filter between patients |
| Recommended filter | 09698 (28 mm bacterial/viral filter, box of 100 units) |
| Sales unit | Per unit (1 kit) |
Family references and consumables verified in the manufacturer’s official documentation (Datospir Touch data sheet FG1102_ES and consumables catalogue FG SIB1201_ES):
| Item | Official Sibelmed reference | Observations |
| Datospir Touch Fleisch transducer module (transducer included in this kit) | 09888 | Official code of the Fleisch module, verified in the Datospir Touch technical data sheet |
| Disposable transducer module (Lilly) | 09889 | Disposable configuration for the same device |
| Turbine transducer module | 09890 | Turbine configuration for the same device |
| Bacterial/viral filter 28 mm (box of 100 units) | 09698 | Official consumable for the Fleisch and turbine modules |
| Disposable transducer with filter (box of 100 units) | 09718 | Disposable 2-in-1 solution for maximum hygiene |
| Disposable transducer (bag of 50 units) | 03169 | Alternative disposable consumable |
| Cardboard mouthpiece 28 ext. × 60 mm (box of 100 units) | 01555 | Mouthpiece compatible with the Fleisch and turbine modules |
| 3-litre calibration syringe | 08290 | Quality control according to ATS/ERS |
| Paediatric mouthpiece adapter | 01149 | For paediatric population (turbine and Fleisch modules) |
4. Equivalences and Cross-References
There is no equivalence table validated by the manufacturer that cross-references this kit with codes from other brands on a 1:1 interchangeable basis; consult the technical department before using it in a tender. The Fleisch transducer is a component specific to each spirometer: although the differential pressure pneumotachographs from other manufacturers (Vyaire/Jaeger, MIR, nSpire, COSMED or Vitalograph) share the same measurement principle, the coupling, the mesh geometry and the connection electronics are specific to each device, so the transducers are not interchangeable between brands. Within the kit itself, only the filter element supports a documented comparison with a third party, verified in the official spirometry filter catalogue of AB Medica Group (Lessa):
| Manufacturer / Brand | Family / Model | Code / SKU | Equivalence level | Technical differences / Remarks |
| AB Medica Group (Lessa) | Spirometry filter for Sibelmed Datospir spirometers | 9984009 | Comparable by category (not interchangeable 1:1) | Code verified in the official Lessa filter catalogue: equipment-side connection of 27.4–28.3 mm, internal 25.3 mm, designed for the Datospir-Sibelmed family (100, 120, 200, 600, 92, 709, Micro C). It functionally replaces only the filter component of the kit; it does not include the handle, bag or Fleisch transducer, so it is not interchangeable 1:1 with the complete reference 07052. |
There is no equivalence table validated by the manufacturer for the complete kit; the above cross-reference is limited to the filter element and must be confirmed against the manufacturer’s IFU before any substitution. For the Fleisch transducer itself, no substitute from another brand has been found, not even at category level, that guarantees the coupling and electronics of the Datospir Touch: always confirm with the technical department.
5. Selection tips and variants of interest
Choose the transducer according to your centre’s profile: the Fleisch in this kit (ref. 07052) is the maximum precision option, recommended for pulmonary function and diagnostics units; if you are looking for lower running costs, the Datospir Touch range also supports the turbine transducer kit (ref. 07053, module 09890) and the disposable Lilly kit (ref. 07046, 25 units), as well as the disposable transducer with filter (ref. 09718).
Remember to renew the bacterial/viral filter for each patient (ref. 09698, 28 mm, box of 100 units) and have compatible cardboard mouthpieces (ref. 01555) and a nasal clip available. For transducer quality control, use the 3-litre calibration syringe (ref. 08290) in accordance with ATS/ERS recommendations, and for paediatric patients attach the paediatric mouthpiece adapter (ref. 01149). The reusable transducer must be disinfected according to the equipment’s instruction manual; confirm the configuration of your Datospir Touch (Easy-F or Diagnostic-F) before purchase to ensure the correct spare part.