Description
1. Commercial Introduction
The SMART ONE (Ref. 33681) is a portable, lightweight and wireless spirometer that turns your smartphone into a complete respiratory monitoring station. Designed for the follow-up of respiratory diseases and patient self-management, it is especially useful in the management of asthma, COPD, post-lung transplant care, cystic fibrosis and in the context of clinical studies. The device fits in the palm of your hand (49×109×21 mm and just 61 g with batteries) and connects via Bluetooth to your phone, so both the professional and the patient have instant access to a graphical history of FEV1 and peak expiratory flow, with a traffic-light health indicator, symptom score and notes for each session.
2. Technical explanation of the equipment
The SMART ONE uses a bidirectional digital turbine as a flow sensor: the patient blows through the mouthpiece and the rotation of the turbine is converted into a flow and volume signal that the device transmits via Bluetooth Low Energy 4.0 to the smartphone (iOS and Android). From this signal, the two key parameters are calculated: FEV1 (forced expiratory volume in the first second) and PEF (peak expiratory flow). The flow range reaches ±16 L/s, with a peak flow precision of ±10% or 0.33 L/s and a volume precision of ±2.5% or 50 ml, and a dynamic resistance below 0.5 cm H2O/L/s. The app includes an exclusive incentive for adults and children based on exhaled flow and volume, which guides the manoeuvre and facilitates accurate results of clinical value. Data storage resides in the smartphone app, allowing sessions to be accumulated and compared with no practical limit on records.
3. Technical Specifications
Equipment specifications
| Parameter | Specification |
|---|---|
| Power supply | 2 AAA batteries 1.5 V (included) |
| Battery life | 1,000 tests |
| Connectivity | Bluetooth® 4.0 Low Energy |
| Dimensions – Weight | 49×109×21 mm – 61 g (with batteries) |
| Nozzles | Ø 30 mm (1.18 inch) |
| Flow sensor | Bidirectional digital turbine |
| Flow range | ±16 L/s |
| Peak flow precision | ±10% or 0.33 L/s |
| Volume precision | ±2.5% or 50 ml |
| Dynamic resistance | <0.5 cm H2O/L/s |
| Measured parameters | FEV1, PEF |
| Memory | The smartphone app stores the data |
App iOS – Android in: GB, FR, IT, ES, PT, DE, NL*, CZ*, PL, RU, TR, DK*, GR, FI, HR*, NO, SE, HU** · *iOS only · **Android only.
Product data
| Manufacturer | GIMA |
|---|---|
| Manufacturer reference | 33681 |
| SKU Cahnos | GM33681 |
| Sales unit | 1 pc |
| Device type | Medical device (Medical device) |
| Device class | II A |
| CND | Z12150102 |
| EAN | 8023279336818 |
Available references
| Reference | Model | Sales unit |
|---|---|---|
| 33681 | SMART ONE – Peak flow and FEV1 on smartphone | 1 pc |
4. Equivalences and Cross-References
Within the catalogue, the SMART ONE (33681) shares a platform with the SMART ONE OXI (33684), which measures the same parameters and incorporates an oximeter in the same device body; both use the same measurement set and the same app. As an original spare part for the flow circuit, the reusable turbine kit + reusable plastic mouthpiece (Ref. 33900) is indicated, valid for models 33681 and 33684. For routine use between patients, compatible disposable mouthpieces from the same brand are recommended (references 33890, 33892, 33410 and 33411, in different sizes and formats). No validated equivalence table from the manufacturer has been found in reliable sources that cross-references this reference with codes from other manufacturers under 1:1 replacement conditions; the coupling, electronics and app are specific to the equipment, so any replacement must be validated beforehand with the technical department.
5. Selection tips and variants of interest
- Check the scope of the test you need: the SMART ONE is oriented towards monitoring FEV1 and peak expiratory flow (PEF) with a smartphone app; if you need a complete examination with forced vital capacity (FVC), pre/post bronchodilator tests and PC software, consider the SPIROBANK / SPIROLAB family (e.g. SPIROBANK II SMART, Ref. 33512, or its version with oximeter, Ref. 33514).
- For equipment maintenance, have a spare kit of reusable turbine and mouthpiece (Ref. 33900) and renew the measurement circuit in case of wear or loss of precision.
- Combine the device with compatible disposable mouthpieces (Ref. 33890, 33892, 33410 and 33411) and, in studies requiring maximum protection, with bacterial-viral filters following your service’s protocol.
- If you also require oxygen saturation (SpO2) in the same workflow, consult the SMART ONE OXI variant (Ref. 33684).
- For large orders you can request a quote: contact our technical-commercial team for order formalisation, supply conditions and advice on compatible accessories.