Description
1. Commercial Introduction
The digital peak expiratory flow meter PF100 is the professional and home solution for effective control of asthma and other respiratory conditions. Its portable design, 240-record memory and USB connection make it the perfect ally for both the patient who needs daily self-monitoring and the healthcare professional who requires reliable and documented data. It includes clinical analysis software to interpret the patient’s progress and facilitate communication with the specialist.
2. Technical explanation of the equipment
The PF100 is an electronic asthma monitor that measures two key parameters of lung function: the peak expiratory flow (PEF), which is the maximum speed at which air leaves the lungs after a full inspiration, and the forced expiratory volume in the first second (FEV1), the volume of air exhaled during the first second of a forced expiration. Both indicators allow detection of airway obstructions and objective monitoring of the disease.
Measurement is performed by blowing forcefully through the mouthpiece; a rotating vane sensor records the flow and the device displays the PEF result for about 3 seconds, followed by the FEV1 value. The device incorporates a traffic light system (green, yellow and red) with a reference value of 500 l/min for intuitive self-assessment: green above 400 l/min, yellow between 300 and 400 l/min and red below 300 l/min. It stores up to 240 measurements with date and time, and when the memory is full, the oldest records are automatically overwritten. Data can be transferred to the computer via the USB port and analysed with the included software, which allows the traffic light limits to be configured individually. The mouthpiece and measuring tube are removable and washable to ensure hygiene between users.
3. Technical Specifications
| Brand / model | Microlife PF 100 |
| Measured parameters | PEF (peak expiratory flow) and FEV1 (forced expiratory volume in 1 s) |
| PEF measurement range | 50 – 900 l/min |
| FEV1 measurement range | 0.01 – 9.99 litres |
| Measurement method | Rotating vane (rotating turbine) |
| PEF accuracy | ±20 l/min or 10% of reading, whichever is greater |
| FEV1 accuracy | ±0.1 l or ±5% of reading, whichever is greater |
| Resolution | PEF 1 l/min; FEV1 0.01 l |
| Memory | 240 measurements with date and time (EEPROM) |
| Connectivity | USB port for data transfer to PC |
| Software | Microlife Asthma Analyzer, included (traffic light configuration, reports and monitoring) |
| Self-assessment | Traffic light system: green >400 l/min, yellow 300–400 l/min, red <300 l/min (reference value 500 l/min) |
| Power supply | 2 batteries, 1.5 V, size AAA |
| Battery life | Approx. 1,000 measurements |
| Dimensions | 77 (width) x 144 (length) x 48 (height) mm |
| Weight | 150 g (with batteries) |
| Protection class | IP22 |
| Operating conditions | 10 – 40 °C, 10 – 85% maximum relative humidity |
| Storage conditions | -5 to +50 °C, 10 – 90% maximum relative humidity |
| Regulatory standard | CE, EN 60601-1, EN 60601-1-2, EN ISO 23747, IEC 60601-1-11, updated ATS 1994 standard |
| Expected service life | 5 years or 10,000 measurements |
| Intended use | Paediatric and adult patients, home monitoring and clinical support |
| Package contents | PF100 device, mouthpiece, measuring tube, USB cable, software, asthma control card, wall bracket and carrying case |
4. Equivalences and Cross-References
The PF100 belongs to the family of digital peak expiratory flow meters from Microlife. Its direct equivalents within the same range are the PF 200 model and its Bluetooth variant PF 200 BT, which share the same measurement parameters (PEF and FEV1) and the same analysis software. The Asthma Analyzer software is compatible with all three models and is available on the manufacturer’s website.
In the Cahnos catalogue, it can be complemented with the manual peak flow meter (ref. 0992059, adult and paediatric sizes), ideal as an economical self-monitoring alternative, and with the portable digital spirometers from the Datsopir range (refs. 02072 Micro A and 02073 Micro C), recommended when a complete spirometry with more than 1,000 stored tests is required. For respiratory patient assessment, it can also be combined with mouthpieces, disposable filters or nitric oxide monitors (ref. ANONBA10V1) available at Cahnos.
5. Selection tips and variants of interest
Choose the PF100 when you need a PEF/FEV1 measurement device with memory and data export for scheduled monitoring of the asthmatic patient, both in the clinic and at home. For basic and economical peak flow monitoring, the manual meter (ref. 0992059) is a valid alternative. If your centre requires complete diagnostic spirometry with curves and quality control, opt for the Datsopir Micro A (ref. 02072) or Micro C (ref. 02073) spirometers.
It is recommended to perform three or more measurements per session and record the highest value, preferably in the morning and in the evening before medication; the traffic light limits should be set together with the specialist. To maintain hygiene between patients, replace the mouthpiece and disinfect the measuring tube following the manufacturer’s instructions. The measurement is for self-monitoring and does not replace diagnosis or medication adjustment, which must always be medically supervised.






Reviews