Description
1. Introduction
The Surgical Mesh Premisut Mesh Regular 30 x 30 cm (Ref. PPMR3030) is a tissue reinforcement implant made of colourless, non-absorbable polypropylene, designed for surgeons and procurement departments looking for a standard-density mesh (80 g/m²), reliable and with excellent value for money for abdominal wall repair and hernia treatment.
Its woven monofilament structure, laser cutting and multidirectional mechanical properties make it a robust solution for both open and laparoscopic techniques, with the guarantee of a sterile single-use product and specialised technical support from Cahnos.
2. Technical Explanation
The mesh is manufactured from 100% extruded monofilament polypropylene, a biostable polymer that is not absorbed or degraded by enzymatic tissue action: the reinforcement remains permanently in the body (permanent tensile strength, non-absorbable due to its technical characteristics).
The weave interlocks the fibres in multiple directions, providing bidirectional elasticity and a balanced distribution of loads on the abdominal wall, allowing the mesh to be adapted and trimmed to the shape and size of the defect without unravelling. The laser cut seals the edges and prevents fraying, ensuring a precise shape. Its pore size of 0.5 x 0.7 mm promotes controlled tissue integration with a moderate material load.
It is mainly indicated for abdominal wall reinforcement and repair of inguinal, femoral, umbilical and ventral hernias, in open and laparoscopic surgery. Supplied sterile by ethylene oxide; do not resterilise and always check the instructions for use (IFU) before implantation.
3. Technical Specifications
| Reference | PPMR3030 |
| Dimensions | 30 x 30 cm |
| Material | Extruded monofilament polypropylene, colourless and non-absorbable |
| Structure | Woven without coating, natural white, laser cut |
| Tensile strength | Permanent, non-absorbable due to its technical characteristics |
| Pore size | 0.5 x 0.7 mm |
| Weight | 80 g/m² |
| Elongation | 89 % |
| Thickness | 0.50 mm |
| Maximum force | 25 kgf |
| Sterilisation | Ethylene oxide (EO). Do not resterilise |
| Use | Single-use, individual sterile pouch |
Indications: abdominal wall reinforcement and treatment of inguinal, femoral, umbilical and ventral hernias. Suitable for open and laparoscopic techniques.
4. Equivalences and Cross-References
There is no official cross-reference table validated by the manufacturer for this mesh. The references listed below are functional category comparisons (same product type and general clinical indication, with verifiable technical differences) and are not interchangeable 1:1: before substituting in a tender or protocol, they must be validated field by field and the official technical data sheets of each manufacturer must be consulted.
| Manufacturer / Brand | Family / Model | Code / SKU | Equivalence level | Technical differences / Remarks |
| B. Braun | Premilene® Mesh (30 x 30 cm) | 1064455 | Comparable by category, not interchangeable 1:1 | Non-absorbable monofilament polypropylene for hernia repair; higher-density mesh: 82 g/m², pore 0.8 mm, thickness 0.48 mm |
| Johnson & Johnson MedTech (Ethicon) | PROLENE® Polypropylene Mesh (30 x 30 cm) | PML1 | Comparable by category, not interchangeable 1:1 | Non-absorbable monofilament polypropylene, standard-weight mesh for inguinal and ventral hernia; verify weight and pore against the official data sheet |
| Medtronic (Covidien) | Parietene™ Macroporous Mesh (30 x 30 cm) | PPM3030 | Comparable by category, not interchangeable 1:1 | Macroporous monofilament polypropylene mesh for abdominal hernia repair; verify weight, pore and dimensions in the official technical data sheet |
| Cahnos | PremisutMesh Lite 30×30 (Ref. PPML3030) | PPML3030 | Comparable by category, not interchangeable 1:1 | Same 30×30 size, but lightweight line: 45 g/m², pore 1.3 x 1.0 mm, thickness 0.43 mm and maximum force 19 kgf |
Fields to check before any substitution: exact dimensions, weight (g/m²), pore size, thickness, structure (monofilament/multifilament), coating, single-use, CE/MDR classification and sterilisation method. There is no equivalence table validated by the manufacturer; consult the technical department before using in a tender.
5. Selection tips and variants of interest
The 30 x 30 cm format is indicated for large abdominal wall defects, eventrations, incisional hernias and repairs requiring a generous overlap onto healthy tissue (an overlap of 3 to 5 cm is recommended for ventral or incisional hernia). For smaller defects, the same Premisut Mesh Regular line maintains identical technical specifications (80 g/m², pore 0.5 x 0.7 mm, thickness 0.50 mm, maximum force 25 kgf) in other formats: PPMR611 (6×11), PPMR7,515 (7.5×15), PPMR1015 (10×15), PPMR1215 (12×15), PPMR1515 (15×15), PPMR1520 (15×20), PPMR1530 (15×30) and PPMR2030 (20×30).
If the centre’s protocol prefers a lighter, more macroporous mesh to reduce foreign body sensation, the alternative within the Cahnos catalogue is the PremisutMesh Lite line (45 g/m², pore 1.3 x 1.0 mm), also available in 30 x 30 cm (Ref. PPML3030). Remember that this mesh must not be used in direct contact with viscera (intraperitoneal route) and that fixation will be performed with sutures or staples according to the surgeon’s criteria, always following the manufacturer’s IFU.





Reviews
There are no reviews yet.