Description
1. Introduction
The PSMEB Shielded Microbiological Safety Cabinet (Ref. CAH-LEM5-900) is the reference solution for nuclear medicine and radiopharmacy services that need to safely prepare low- and medium-energy radiopharmaceutical substances. Designed for the manual preparation of radiopharmaceuticals, this cabinet combines the microbiological containment of a Class A (ISO 5) cabinet with effective radiological protection against ionising radiation, allowing the professional to work with total confidence in generator elution, kit reconstitution, cell labelling and dose fractionation. Manufactured to European standards and distributed by Cahnos, it is the ideal choice both for newly created hospital radiopharmacies and for the renovation of preparation stations with current radioprotection criteria.
2. Technical explanation of the equipment
The PSMEB is a Class II microbiological safety cabinet (according to European standard EN 12469) with integrated shielding for operator protection against ionising radiation. Its filtration system consists of an automatic control fan and two HEPA H14 filters: one supply filter that introduces clean air into the work area and one exhaust filter that expels air to the outside. 70% of the air is recirculated through the main HEPA H14 filter and the remaining 30% is renewed through the HEPA H14 exhaust filter, with an activated carbon filter option, maintaining an air barrier at a minimum velocity of 0.4 m/s.
Radiological protection is achieved by means of lead shielding of variable thickness (from 5 to 25 mm depending on the model) and a movable front lead glass panel (from 30 to 80 mm), which allows the operator to choose their working position, sitting or standing, with full visibility of the operation. The work area, made of 316L stainless steel, achieves ISO 5 (Class A) air quality, making it suitable for the simultaneous and safe handling of pathogenic substances and radiopharmaceuticals, in accordance with the ALARA principle.
3. Technical Specifications
The reference specifications for the PSMEB Class A cabinet from the Cahnos radioprotection range are detailed below:
| Parameter | Specification |
| Reference (SKU) | CAH-LEM5-900 |
| Product type | Shielded microbiological safety cabinet (PSMEB) for manual preparation of low- and medium-energy radiopharmaceuticals |
| Safety class | Class A (ISO 5) in the work area; type II A2 cabinet according to EN 12469 |
| Shielding | Lead from 5 to 25 mm depending on model; lead glass screen from 30 to 80 mm |
| Filtration | 2 HEPA H14 filters (efficiency > 99.995% MPPS) |
| Air recirculation | 70% recirculated / 30% renewed by exhaust filter |
| Ventilation | Automatic control fan; minimum air barrier of 0.4 m/s |
| External dimensions (mm) | Width 1,248 to 2,160 x Depth 870 x Height 2,540 |
| Work surface (mm) | Width 885 to 1,800 x Depth 660 x Height 580 |
| Weight | 1,800 to 2,300 kg depending on configuration |
| Exterior finish | Steel with epoxy paint |
| Interior finish | Stainless steel 316L |
| Lighting | > 1,000 lux; 15 W UV-C germicidal lamp |
| Noise level | < 54 dB(A) |
| Power supply | 220/240 V, 50 Hz, 16 A |
The cabinet is indicated for handling the following low- and medium-energy radionuclides:
| Radionuclide | Energy | Typical application |
| Tc-99m | Low (140 keV) | Generator elution 99Mo/99mTc, scintigraphy and SPECT |
| Tl-201 | Low | Myocardial perfusion studies |
| In-111 | Medium | Cell labelling and scintigraphy |
| I-123 | Medium | Thyroid diagnostics |
| Lu-177 | Medium | Radionuclide therapy (targeted radiotherapy) |
Available options: shielded compartment for two generators (50 mm lead) with double access and independent lifts; shielded compartment for activimeter (15 or 25 mm lead) with Posilift pedal-operated lifting system; shielded compartment for waste container (15 or 25 mm lead); activated carbon filter and dry contact for status return.
4. Equivalences and Cross-References
The PSMEB cabinet belongs to the family of shielded microbiological safety cabinets from the French manufacturer Lemer Pax, European leader in radiological protection. In the distribution catalogue, this unit corresponds to reference LEM5-900 (IMA-X) and to the Lemer Pax PSMEB 10-1200 model (10 mm lead shielding, 1,200 mm length and 40 mm leaded glass window). The same platform is available in other shielding configurations: PSMEB 5-1200 (5 mm lead, 30 mm glass), PSMEB 15-1200 (15 mm lead, 55 mm glass) and PSMEB 25-1200 (25 mm lead, 80 mm glass), as well as the PSMEB LME, PSMEB BME and PSMEB HE variants for specific applications.
Within Cahnos’s own catalogue, it shares its field of application with the BioSafe II Semi-Open Shielded Microbiological Cabinet (Ref. CAH-CSM-II-10PB) for radiopharmaceutical preparation and cell labelling, and with the Asypharma Compact Class A Cell (Ref. CAH-LEM17860) for closed configuration with glove rings. For open-access configurations, the alternative is the RadWork Shield Open Workstation (Ref. CAH-PTA-50-30VP); for high-energy PET production, the SynthShield Pro Cell (Ref. CAH-SSP-AE); and for total containment in closed configuration, the RadSafe Pro Shielded Cell (Ref. CAH-CB-20/50-2G). On the international market, its functionality is similar to the preparation cells and cabinets from Comecer and Tema Sinergie with equivalent lead thicknesses and leaded glass windows.
5. Selection tips and variants of interest
Select the shielding thickness according to the energy and activity you will be handling: for low-energy Tc-99m and Tl-201, 5 or 10 mm lead configurations may be sufficient; for medium-energy I-123, In-111 or Lu-177 with high activities, it is recommended to opt for 15 or 25 mm shielding. Assess the maximum manageable activity to keep the dose rate below 25 µSv/h at 5 cm from the walls, in accordance with your radiation protection programme.
Configure the compartments according to your workflow: if your department works with frequent generator elutions, the shielded compartment for two generators with independent lifts facilitates operation without opening the barrier; if activity verification is critical, the compartment for the activimeter with Posilift system allows measurement with foot pedal control; the waste compartment improves radioactive waste management without interrupting preparation.
Consider ergonomics: the space under the work surface allows alternating between sitting and standing positions, reducing fatigue during high-workload shifts. Verify the installation conditions: access door of at least 90 cm, floor load from 1,600 to 2,240 kg/m² depending on the model, and internal ventilation outlet of 200 mm with a capacity of 500 m³/h. Complete the radiation protection line with syringe shields, shielded syringe holders, shielded waste containers and mini source stores available at Cahnos. Consult our technical-commercial team to confirm the configuration, exact dimensions, options and formalised quotation according to the needs of your centre.