Description
1. Introduction
The Asypharma Compact hot cell (manufactured by the European leader in radiation protection Lemer Pax as the Easypharma Compact, ref. 00017860) is the reference tool for nuclear medicine departments and hospital radiopharmacies that need to prepare low- and medium-energy radiopharmaceuticals safely, comfortably and reproducibly. With its Class A (ISO 5) air classification under laminar flow, its 2 glove ports and its lead shielding optimised zone by zone, this cell allows working without a lead apron, with ergonomic comfort and with staff exposure reduced to a minimum in accordance with the ALARA principle. A robust, ultra-compact and easily integrated investment, designed both for newly created radiopharmacies and for the renovation of preparation stations to current radiation protection standards.
2. Technical explanation of the equipment
The Easypharma Compact is a closed-containment hot cell designed for the manual preparation of low- and medium-energy radioisotopes used in gamma camera examinations: elution of technetium generators 99Mo/99mTc, fractionation, radiolabelling, reconstitution of radiopharmaceuticals and kits, and activity measurement. The operator accesses the interior exclusively through 2 glove rings (156 mm diameter), so that the barrier remains closed throughout the manipulation, protecting the professional from both contamination and whole-body exposure to ionising radiation.
Radiation protection is provided by lead screens of variable thickness: 15 mm on the work surface, 50 mm in the shielded generator compartment, 15 mm on the dose calibrator and waste containers, and 20 mm on the lateral airlock. This configuration guarantees a dose rate below 25 µSv/h at 5 cm from the walls with the usual activities of Tc-99m, Tl-201, In-111, I-123 and Lu-177. Air quality is maintained by an autonomous ventilation and filtration system: laminar flow Class A (ISO 5) over the entire work surface using HEPA filters, Class C (ISO 7) quality in the generator compartment and airlocks, and internal negative pressure of -180 Pa (±20 %) with extraction of 120 m³/h through HEPA filter and activated carbon.
The unit incorporates two generators with independent lifts compatible with the Tekcis® and Ultra Technekow™ models (others on request), a dose calibrator compartment (Lemer Pax, Medisystem, Capintec or Veenstra) with optional Posilift pedal-operated lifting system, two shielded waste containers, a lateral airlock with sliding tray and interlocked doors to maintain air class integrity, a multiple internal power outlet (electrical, RJ45 and USB) and germicidal UV-C lamps with timer for effective disinfection.
3. Technical Specifications
3.1. General data
| Reference (SKU) | CAH-LEM17860 |
| Manufacturer reference | Lemer Pax 00017860 |
| Manufacturer | Lemer Pax (France), distributed by Cahnos |
| Product type | Hot cell for manual preparation |
| Configuration | Closed, with 2 glove ports |
| Air quality | Class A (ISO 5) on work surface; Class C (ISO 7) in generators and airlocks |
| Radionuclides | Tc-99m, Tl-201, In-111, I-123, Lu-177 |
| Regulatory standard | Machinery Directive 2006/42/EC, EMC 2004/108/EC, EURATOM Directive 96/29, GPP/GMP |
| External dimensions | L 1,553 / 1,894 (1 or 2 airlocks) x D 959 x H 2,165 mm |
| Empty weight | 2,500 / 2,600 kg (1 or 2 airlocks) |
3.2. Lead shielding (radiological protection)
| Zone | Lead thickness |
| Work surface | 15 mm |
| Generator compartment | 50 mm |
| Activity meter compartment | 15 mm |
| Waste container housings | 15 mm |
| Side airlock (sas) | 20 mm |
3.3. Work surface and ports
| Parameter | Specification |
| Work surface dimensions | L 954 x D 585 x H 565 mm (effective L 954 x D 520 mm) |
| Working height | 995 mm |
| Finish | Stainless steel 316L |
| Glove ports | 2, diameter 156 mm, axis at 1,080 mm |
| Lead glass window | L 535 x H 368 mm, thickness 58 mm (Vision+ full-door option) |
| Lighting | LED > 1,000 lux; germicidal UV-C 2 x 8 W with timer |
| Noise level | < 63 dB(A) |
3.4. Integrated compartments
| Compartment | Specification |
| Generators | 2 generators 99Mo/99mTc (Tekcis®, Ultra Technekow™) with loading trays and 2 independent lifts; internal dimension L 369 x D 390 x H 410 mm; door with safety lock |
| Activity meter | Ø 220 x H 510 mm; compatible with Lemer Pax, Medisystem, Capintec and Veenstra; Posilift option with pedal |
| Waste | 2 housings of Ø 145 x H 210 mm; compatible containers Dispo 2L and Septoeco PBS New 2L (AP Medical); full manual or partial magnetic opening |
| Side airlock (sas) | Internal dimension L 319 x D 182 x H 232 mm; sliding tray with max. load 10 kg; shielded window L 155 x H 195 x 24 mm; pedal opening |
3.5. Ventilation and filtration system
| Parameter | Specification |
| Extraction flow rate | 120 m³/h (1.7 m/s at the cell outlet) |
| Extraction filtration | HEPA filter + activated carbon filter |
| Laminar flow filtration | HEPA filter over the entire work surface |
| Internal negative pressure | -180 Pa (± 20 %) |
| Extraction outlet | Ø 100 mm (independent connection to the ventilation system) |
3.6. Electrical data and installation requirements
| Parameter | Specification |
| Power supply | 230 V, 50 Hz, 16 A (2P+G) |
| Internal sockets | 4 (electrical, RJ45 or USB as required) |
| Floor load | 2,900 kg/m² |
| Door passage width | ≥ 90 cm |
| Compressed air | 6 bar, 20 L/min |
3.7. Maximum manageable activity (dose rate < 25 µSv/h at 5 cm from the walls)
| Radionuclide | Maximum activity |
| Tc-99m | 7.07 GBq |
| Tl-201 | 1.04 GBq |
| In-111 | 0.129 TBq |
| Lu-177 | 0.185 TBq |
| I-123 | 307 MBq |
Note: calculation conditions with point sources exposed in contact with the inner wall of the cell, according to ASN Guide No. 32. Exact dimensions must be confirmed by means of an installation plan and with the technical-commercial team of Cahnos.
4. Equivalences and Cross-References
The Asypharma Compact cell corresponds to the Easypharma Compact from Lemer Pax (ref. 00017860), distributed internationally through the IMA-X network and present in the main nuclear medicine departments in Europe. It shares platform and field of application with the PSMEB Class A Cabinet (ref. CAH-LEM5-900) from the Cahnos catalogue, also manufactured by Lemer Pax, for manual preparation of low and medium energy with semi-open access.
Within Cahnos’s own catalogue, the Asypharma Compact is complemented by the RadSafe Pro Shielded Cell (ref. CAH-CB-20/50-2G) for closed configurations with 20-50 mm of lead, the BioSafe II semi-open Microbiological Cabinet (ref. CAH-CSM-II-10PB) and the RadWork Shield Open Workstation (ref. CAH-PTA-50-30VP) as open-access alternatives, and the SynthShield Pro Cell (ref. CAH-SSP-AE) for high-energy PET production. In the international market, its functionality is similar to the manual preparation hot cells of Comecer and Tema Sinergie, with equivalent lead thicknesses (15-50 mm) and a 2-glove-port configuration.
5. Selection tips and variants of interest
- Validate the energy of your radiopharmaceuticals: this cell is indicated for low and medium energy (Tc-99m, Tl-201, In-111, I-123, Lu-177). If you handle high-energy emitters (F-18, Ga-68, I-131), consider the option of additional 30 mm lead shielding or the SynthShield Pro cell from the catalogue for PET production.
- Custom ergonomics: the working position can be chosen seated or standing. The Vision+ option (fully laminated lead glass front door) and the height lift improve visibility and reduce fatigue during high-load shifts.
- Make the most of daily elution: the two generator compartments with independent lifts allow working with two 99Mo/99mTc generators simultaneously and performing daily elutions without opening the barrier.
- Plan the installation: check the floor load (2,900 kg/m²), the door width (≥ 90 cm), the ventilation outlet of Ø 100 mm and the compressed air supply (6 bar, 20 L/min). Load distribution plates are recommended if the floor slab capacity does not correspond to the weight of the cell.
- Complete your radiation protection range: combine this cell with the Scintidose or Medi 405 activimeters, vial and syringe shields, bench shields, shielded waste containers and the handling accessories available at Cahnos for a safe workflow from start to finish.
- Request a formal quotation: confirm with our technical and commercial team the transfer lock configuration (1 or 2), the generator and activimeter model, and the required options before placing the order.