Description
1. Commercial Introduction
The Silk Suture PREMISUT Prime is the reference of the premium range of non-absorbable silk from Cahnos for the professional who demands a natural-origin thread that is manageable and with predictable clinical behaviour. This configuration combines a 100% natural braided silk multifilament, coated with wax and silicone, in high-visibility black, USP gauge 4/0, mounted on a ½ circle needle with cylindrical body and Taperpoint conical tip of 16 mm (HR16 profile) and a 75 cm strand.
Sterile product, ready to use, presented in a box of 12 units in individual sachets with full batch traceability. Ideal for hospital replenishment, purchasing centres, tenders and volume contracts that seek a permanent high-performance thread, consistent sharpening from batch to batch and excellent value for money.
2. Technical Explanation
Surgical silk is a natural protein fibre —fibroin from the cocoon of Bombyx mori— presented in a braided multifilament construction and coated with a mixture of wax and silicone. This coating reduces capillarity and bacterial drag, favours the sliding of the strand through the tissue and guarantees a firm and secure knot with the recommended tying technique 2 = 1 = 1 = 1, one of the most valued qualities of surgical silk.
Being a non-absorbable material, the thread does not degrade by hydrolysis: the body encapsulates it gradually by connective tissue and maintains tensile strength for more than three months, comfortably covering the critical healing phase. The HR16 needle of 16 mm presents a ½ circle curvature with cylindrical body and Taperpoint conical tip: it separates and displaces tissue fibres without cutting them, minimising trauma and bleeding at the entry point, a decisive advantage in soft, lax and well-vascularised tissues, as well as in reduced surgical fields of ophthalmology and skin surgery. The USP gauge 4/0 (metric EP 1.5) balances fineness and resistance for approximation of soft tissues. Sterilisation by ethylene oxide (EO); single-use product, not re-sterilisable.
3. Technical Specifications
Data verified according to the technical data sheet and instructions for use of the PREMISUT Prime range.
3.1. General characteristics
| Parameter | Specification |
| Factory reference | P542HR1640 (Prime range) |
| Reference Cahnos (SKU) | AMP542HR1640 |
| Brand / Quality | PREMISUT Prime (premium range) |
| Thread material | 100% natural silk (Bombyx mori fibroin), braided multifilament |
| Thread structure | Braided (multifilament), Prime non-grooved finish |
| Coating | Wax and silicone (anti-capillarity, low tissue drag) |
| Absorption type | Non-absorbable; gradual encapsulation, wound support for more than 3 months |
| Gauge / Size | USP 4/0 (EP metric 1.5), in accordance with USP Pharmacopoeia and European Pharmacopoeia |
| Approximate thread diameter | 0.150 – 0.199 mm (USP Pharmacopoeia, non-absorbable suture) |
| Thread length | 75 cm (single strand) |
| Thread colour | Black non-ribbed (high visibility in surgical field) |
| Needle shape | ½ circle (1/2 circular), cylindrical body |
| Tip type | Taperpoint conical (non-cutting) |
| Needle size | HR16 Prime (16 mm), 1 needle per thread |
| Swaged | 1 swaged needle / 1 thread per unit |
| Recommended knotting technique | 2 = 1 = 1 = 1 |
| Sterilisation | Ethylene oxide (EO); sterile, single-use, non-resterilisable |
| Indications | Approximation of soft tissues, ligatures, ophthalmology and skin surgery |
| Commercial presentation | Box of 12 units (sterile individual sachets); minimum order of 30 boxes; batch traceability |
| Regulatory standard | Conforms to USP/EP Pharmacopoeia for non-absorbable sutures; CE marking (MDR) |
3.2. Gauge equivalence according to USP/EP Pharmacopeia (non-absorbable suture)
| USP size | Metric measurement (EP) | Approximate thread diameter |
| 5/0 | 1.0 | 0.10 – 0.149 mm |
| 4/0 (this product) | 1.5 | 0.150 – 0.199 mm |
| 3/0 | 2.0 | 0.20 – 0.249 mm |
The diameter of the strand complies with the official ranges of the USP Pharmacopoeia for non-absorbable sutures, without exceeding its limits.
3.3. Decoding the factory reference
The reference P542HR1640 is decoded as follows: P = Prime quality; 542 = black silk thread of 75 cm (541 = 45 cm); HR16 = ½ circle needle with cylindrical body and Taperpoint conical tip of 16 mm; 40 = USP gauge 4/0 (50 = 5/0, 60 = 6/0). Cahnos markets the official reference without altering characters: no digit may be added, removed or reordered when ordering or tendering.
3.4. PREMISUT family in silk with ½ circle Taperpoint conical point needle and black thread
| Needle profile | USP size | Thread | Reference (SKU) |
| HR12 Prime (taper point, 12 mm) | 6/0 | 1 × 45 cm (black) | P541HR1260 / AMP541HR1260 |
| HR13 Prime (taper point, 13 mm) | 6/0 | 1 × 45 cm (black) | P541HR1360 / AMP541HR1360 |
| HR13 Prime (taper point, 13 mm) | 5/0 | 1 × 75 cm (black) | P542HR1350 / AMP542HR1350 |
| HR13 Prime (taper point, 13 mm) | 4/0 | 1 × 75 cm (black) | P542HR1340 / AMP542HR1340 |
| HR16 Prime (taper point, 16 mm) · this product | 4/0 | 1 × 75 cm (black) | P542HR1640 / AMP542HR1640 |
| HR16 Prime (taper point, 16 mm) | 5/0 | 1 × 75 cm (black) | P542HR1650 / AMP542HR1650 |
| HR16 Prime (taper point, 16 mm) | 6/0 | 1 × 45 cm (black) | P541HR1660 / AMP541HR1660 |
4. Equivalences and Cross-References
The reference P542HR1640 corresponds to the official nomenclature of the PREMISUT Prime range (HR16 profile: ½ circle needle with cylindrical body and Taperpoint conical tip of 16 mm, USP gauge 4/0, black thread of 75 cm and single needle configuration) and Cahnos markets it without altering characters. The coated non-absorbable black braided silk is one of the most traditional and standardised natural suture families on the market.
There is no equivalence table validated by the manufacturer that cross-references this specific configuration with references from other houses; the following row is a functional comparison verified against the manufacturer’s official catalogue and authorised distribution, and does not constitute an exact reference equivalence (reserved to the same manufacturer). It must be validated against the IFU and the CE/MDR classification before any substitution.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
| B. Braun (Aesculap) | Silkam® Black – natural braided and impregnated multifilament silk, black, USP 4/0 (EP 1.5), 75 cm, HR17 needle of ½ circle, round body (taper) | C0760137 / G0760137 | Comparable by category (not interchangeable 1:1) | Code verified in the official catalogue of B. Braun and authorised distribution: they match material (coated braided silk), gauge 4/0 (metric 1.5), black colour, length 75 cm, ½ circle curvature, round/taper body and needle size (17 mm vs 16 mm, within tolerance ±1 mm for needles ≥ 10 mm). They differ in the exact coating (wax/paraffin vs wax and silicone) and sterilisation (gamma radiation vs ethylene oxide). Confirm IFU before substituting. |
This needle geometry (HR16, taper point, USP 4/0, black thread of 75 cm) is only interchangeable with full safety within the PREMISUT system itself. Other homologous families of black braided silk in the sector —Ethicon MERSILK® / PERMA-HAND™ Silk (e.g. reference W582 on a BB needle of 17 mm in ⅜ circle, which differs in curvature), Medtronic Sofsilk™, DemeTECH DemeSILK™, Assut Assusilk®, LuxSutures® LuxSilk, Serag-Wiessner, Unilene, Lorca Marín and Laboratorio Aragó— must be reviewed property by property (material, USP/metric gauge, ½ circle curvature, taper point family, needle size and length, thread length, colour, coating and CE/MDR regulatory classification) before any substitution. Consult with the technical department of Cahnos before using in tenders.
5. Selection tips and variants of interest
Tissue to suture: the USP 4/0 gauge (metric EP 1.5, diameter 0.150 – 0.199 mm) is the precision choice for soft tissue approximation, subcutaneous planes and skin surgery. The Taperpoint HR16 conical needle, 16 mm in a ½ circle, separates the fibres without cutting them, which is why it is especially recommended for lax, well-vascularised tissues of high aesthetic quality, as well as in surgical fields of ophthalmology and skin surgery.
Needle size: the 16 mm HR16 offers an intermediate reach with great rotational control in confined fields. If the protocol requires a shorter needle with the same gauge and conical point, the HR13 4/0 75 cm variant is available (SKU AMP542HR1340); if greater reach is sought while maintaining the conical point, move up to the higher profiles of the HR range.
Portfolio unification: the same HR16 Prime needle is offered in silk in 5/0 gauge, 75 cm (SKU AMP542HR1650) and 6/0, 45 cm (SKU AMP541HR1660), allowing several gauges to be covered with a single supplier and volume B2B conditions. For extreme precision needs, the HR13 Prime family covers gauges 4/0, 5/0 and 6/0.
Sterility verification: product sterilised by ethylene oxide and single-use; check the integrity of the blister and the expiry date before each procedure, keep the reference P542HR1640 together with the batch number for traceability in the operating theatre and store below 30 °C, away from moisture and direct sunlight. For special volume B2B conditions, hospital supply agreements or validation of equivalents from other commercial houses, contact the technical-commercial team at Cahnos.