Description
1. Commercial Introduction
The SILK Suture PREMISUT PRIME in USP gauge 2 is the reference of the premium range from Cahnos for closure of soft tissues, ligatures, ophthalmology and skin surgery when a non-absorbable material of natural origin with a firm knot and predictable clinical behaviour is required. Mounted on a ½-circle needle with a 37 mm Taperpoint conical point (HR37 profile) and presented in double black fluted thread 2×75 cm, it combines the tradition of surgical silk with a precision finish and full batch traceability. Sterile product, ready to use, in a box of 12 units in individual pouches.
2. Technical Explanation
Surgical silk is a natural protein fibre (fibroin, from the cocoon of Bombyx mori) processed to remove natural waxes and gums and presented in a braided multifilament construction, coated with a mixture of wax and silicone. This coating reduces the capillary effect and friction when passing through tissue, improves handling and favours knot seating (recommended knotting technique 2 = 1 = 1 = 1). Being a non-absorbable material, the thread does not degrade by hydrolysis: it is gradually encapsulated by connective tissue and maintains tensile strength for more than three months, comfortably covering the critical healing phase.
The 37 mm HR37 Prime needle, ½-circle with cylindrical body and Taperpoint conical point, offers atraumatic penetration ideal for soft and non-distensible tissues, minimising tissue damage during passage. USP gauge 2 (metric EP 5) provides high strength for planes under medium-high tension, and the black fluted colour ensures optimal contrast on the surgical field. The double-thread presentation (2 × 75 cm) allows two strands to be available in the same pouch for continuous sutures or to distribute the material across two fields. Sterilisation by ethylene oxide (EO); single-use product, not re-sterilisable.
3. Technical Specifications
3.1. General characteristics
| Parameter | Specification |
|---|---|
| Brand / Quality | PREMISUT (Prime line) |
| Factory reference | P542RHR3702.2 |
| SKU Cahnos | AMP542HR3702.2 |
| Thread material | 100 % natural silk (fibroin from Bombyx mori), multifilament |
| Thread structure | Braided (multifilament), fluted surface |
| Coating | Wax and silicone |
| Absorption type | Non-absorbable; gradual encapsulation, wound support for more than 3 months |
| Gauge / Size | USP 2 (metric EP 5), conforming to USP and European Pharmacopoeia |
| Approximate thread diameter | 0.50 – 0.599 mm (USP range for non-absorbable suture) |
| Number of threads | 2 strands of 75 cm (2 × 75 cm) |
| Thread length | 2 × 75 cm |
| Thread colour | Grooved black (high visibility) |
| Needle shape | ½ circle (1/2 circular), cylindrical body |
| Tip type | Conical / Taperpoint (grooved) |
| Needle size | 37 mm (HR 37 Prime profile) |
| Swaged | 2 strands of 75 cm, each with a swaged needle |
| Recommended knotting technique | 2 = 1 = 1 = 1 |
| Sterilisation | Ethylene oxide (EO); sterile, single-use, non-resterilisable |
| Indications | Approximation of soft tissues, ligatures, ophthalmology and skin surgery |
| Commercial presentation | Box of 12 units (individual sterile sachets) |
3.2. Gauge equivalence according to USP/EP Pharmacopeia (non-absorbable suture)
| USP size | Metric measurement (EP) | Approximate thread diameter |
|---|---|---|
| 2 (this product) | 5 | 0.50 – 0.599 mm |
| 1 | 4 | 0.40 – 0.499 mm |
| 0 | 3.5 | 0.35 – 0.399 mm |
| 2/0 | 3 | 0.30 – 0.339 mm |
3.3. Family PREMISUT Prime with HR37 needle (½ circle, Taperpoint conical point) in Cahnos catalogue
| USP size | Presentation | Reference (SKU) |
|---|---|---|
| 2 | 2 × 75 cm (double thread) | AMP542HR3702.2 (this product) |
| 2 | 1 × 75 cm | AMP542HR3702 |
| 1 | 2 × 75 cm (double thread) | AMP542HR3701.2 |
| 0 | 2 × 75 cm (double thread) | AMP542HR3710.2 |
| 2/0 | 2 × 75 cm (double thread) | AMP542HR3720.2 |
4. Equivalences and Cross-References
The reference P542RHR3702.2 corresponds to the official nomenclature of the PREMISUT Prime range (HR37 profile: ½-circle needle with cylindrical body and 37 mm Taperpoint conical point, USP gauge 2, black fluted thread 2×75 cm, double strand) and Cahnos markets it without altering characters: no digit should be added, removed or reordered when ordering or tendering.
Non-absorbable black braided silk is one of the most standardised natural suture families on the market (commercial types Silkam®, MERSILK™/PERMA-HAND™ Silk or DemeSILK™). The cross-reference detailed below has been verified against the manufacturer’s official catalogue; as it is another brand, it does not constitute an exact reference equivalence (reserved to the same manufacturer) and must be validated against the IFU before any substitution.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
|---|---|---|---|---|
| B. Braun | Silkam® Black – natural braided coated multifilament silk, black thread | C0760544 – SILKAM BLACK 2 (5) 75CM HR37 (verified in official catalogue B. Braun) | Comparable by category (not interchangeable 1:1) | They match in material (coated braided silk), USP gauge 2 (EP 5), black colour, ½-circle curvature, 37 mm HR37 needle with cylindrical body and conical point, and thread length of 75 cm. They differ in configuration (single strand vs. double thread 2×75 cm of the PREMISUT Prime), exact coating (wax/paraffin vs. wax and silicone), sterilisation method (gamma radiation vs. ethylene oxide) and commercial format (box of 36 vs. 12). Confirm IFU and CE/MDR classification before substituting. |
The MERSILK™/PERMA-HAND™ Silk family from Ethicon (J&J MedTech) offers references in USP gauge 2 with conical point (e.g. W936, 75 cm, 40 mm MO-2 needle); it matches in material, gauge and length, but its 40 mm needle exceeds the ±1 mm tolerance compared to the 37 mm HR37, so it is not listed as comparable and remains pending technical validation.
There is no equivalence table validated by the manufacturer for this reference against the brands cited; the cross-references indicated have been verified against the official catalogue of B. Braun, but must be confirmed with the technical department of Cahnos before use in tendering or direct substitution. Full interchangeability is only guaranteed within the PREMISUT Prime system itself, in its double-thread presentation (Ref. AMP542HR3702.2) and in the range of identical geometry and gauge (Ref. AMP542HR3702 or AM542HR3702).
5. Selection tips and variants of interest
- Tissue to suture: USP gauge 2 (EP 5) provides high strength for subcutaneous tissue, fasciae, aponeuroses and soft tissues with medium-high tension, always with a conical point needle to minimise tissue damage. For finer or paediatric planes, it can be reduced to USP 0 (Ref. AMP542HR3710.2) or USP 2/0 (Ref. AMP542HR3720.2) within the same HR37 geometry and double-thread presentation.
- Double presentation: the 2 × 75 cm format is ideal for continuous sutures without re-threading or to distribute the material across two fields. If the protocol requires a single strand, the same needle and gauge are available in 1 × 75 cm (Ref. AMP542HR3702).
- Needle length: the 37 mm HR37 profile covers accesses of medium depth. For smaller fields, the HR30 family is available, and for fields with greater travel, the HR40 profile (Ref. AM542HR4002, USP 2) within the PREMISUT silk range, allowing supplier unification across the entire family.
- Prime vs. standard quality: the Prime range provides maximum sharpening precision and batch-to-batch consistency; if unit cost is a critical factor, the standard range of identical geometry and gauge (Ref. AM542HR3702) maintains the same dimensional specifications with standard sharpening performance.
- Sterility verification: sterile product by ethylene oxide and single-use; check the integrity of the blister and the expiry date before each use, and keep the reference P542RHR3702.2 together with the batch number for traceability in the operating theatre. Store below 30 °C, away from moisture and direct sunlight.
- B2B conditions: for volume purchases, hospital supply agreements or validation of equivalents from other commercial houses, contact the technical-commercial team of Cahnos.





Reviews