Description
1. Commercial Introduction
The Silk Suture PREMISUT Prime with 30 mm HR30 needle is the premium reference in the range of non-absorbable silk sutures from Cahnos for professionals who demand a natural, manageable thread with maximum knot security. It combines a 100% natural braided silk multifilament coated with wax and silicone, in identifying black, with a 75 cm strand mounted on a ½ circle needle with cylindrical body and 30 mm Taperpoint conical point (HR30 Prime profile), in USP 1 gauge (EP metric 4).
Sterile product, ready to use, in a box of 12 individual sachets with full batch traceability. Aimed at general surgery services, ophthalmology units, skin surgery and B2B purchasing centres, with special volume conditions for hospitals, clinics and tenders.
2. Technical explanation of the material and procedure
Surgical silk is a natural protein fibre (fibroin from the cocoon of Bombyx mori) presented in a braided multifilament construction coated with a mixture of wax and silicone. This coating reduces capillarity and bacterial drag, favours thread sliding through tissue and guarantees a firm, secure knot with the recommended tying technique 2 = 1 = 1 = 1, one of the most valued qualities of surgical silk.
Being a non-absorbable material, the thread does not degrade by hydrolysis: the body gradually encapsulates it through connective tissue and maintains tensile strength for more than three months, comfortably covering the critical healing phase. The 30 mm HR30 Prime needle features a ½ circle curvature with cylindrical body and Taperpoint conical point: it separates and displaces tissue fibres without cutting them, minimising trauma to delicate, lax or well-irrigated structures. The USP 1 gauge (EP metric 4) offers high resistance for planes with medium-high tension, and the grooved black colour guarantees immediate contrast on the surgical field. Sterilisation by ethylene oxide (EO); single-use product, not re-sterilisable.
3. Technical Specifications
3.1. General characteristics
| Parameter | Specification |
| Brand / Quality | PREMISUT Prime (premium range, grooved needle line) |
| Factory reference | P542HR3001 |
| Reference Cahnos (SKU) | AMP542HR3001 |
| Thread material | 100% natural silk (fibroin from Bombyx mori), braided multifilament |
| Thread structure | Braided (multifilament), fluted surface |
| Coating | Wax and silicone (anti-capillarity, low tissue drag) |
| Absorption type | Non-absorbable; gradual encapsulation, wound support for more than 3 months |
| Gauge / Size | USP 1 (metric EP 4); conforming to USP Pharmacopoeia and European Pharmacopoeia |
| Approximate thread diameter | 0.40 – 0.499 mm (USP Pharmacopoeia range, non-absorbable suture) |
| Thread length | 75 cm (single strand) |
| Thread colour | Fluted black (high visibility on the surgical field) |
| Needle shape | ½ circle (1/2 circular), cylindrical body |
| Tip type | Taperpoint conical (non-cutting), grooved |
| Needle size | 30 mm (HR 30 Prime profile) |
| Swaged | 1 needle / 1 thread of 75 cm per unit |
| Recommended knotting technique | 2 = 1 = 1 = 1 |
| Sterilisation | Ethylene oxide (EO); sterile, single-use, non-resterilisable |
| Indications | Approximation of soft tissues, ligatures, ophthalmology and skin surgery |
| Commercial presentation | Box of 12 units (sterile individual sachets); batch traceability; minimum order 30 boxes |
| Regulatory standard | Compliant with USP/EP Pharmacopoeia for non-absorbable sutures; CE/MDR classification |
3.2. Gauge equivalence according to USP/EP Pharmacopeia (non-absorbable suture)
| USP size | Metric measurement (EP) | Approximate thread diameter |
| 1 (this product) | 4 | 0.40 – 0.499 mm |
| 0 | 3.5 | 0.35 – 0.399 mm |
| 2/0 | 3 | 0.30 – 0.339 mm |
The thread diameter complies with the official USP Pharmacopoeia ranges for non-absorbable sutures, without extending their limits.
3.3. Family PREMISUT Prime with HR30 needle (½ circle, Taperpoint conical point) in Cahnos catalogue
| USP size | Presentation | Reference (SKU) |
| 1 | 1 × 75 cm | AMP542HR3001 (this product) |
| 1 | 2 × 75 cm (double thread) | AMP542RHR3001.2 |
| 0 | 1 × 75 cm | AMP542HR3010 |
| 2/0 | 1 × 75 cm | AMP542HR3020 |
4. Equivalences and Cross-References
The reference P542HR3001 corresponds to the official nomenclature of the PREMISUT Prime range (HR30 profile: ½ circle needle with cylindrical body and 30 mm Taperpoint conical point, USP 1 gauge, 75 cm grooved black thread, single strand assembly) and Cahnos markets it without altering characters: no digit may be added, removed or reordered when ordering or tendering.
Non-absorbable black braided silk is one of the most standardised natural suture families on the market (commercial types Silkam®, MERSILK®/PERMA-HAND™ Silk or DemeSILK™). The following cross-references have been checked against the manufacturer’s official catalogue, regulatory database (GUDID/FDA) or authorised distribution; as they are another brand, they do not constitute an exact reference equivalence (reserved for the same manufacturer) and must be validated against the IFU before any substitution.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
| B. Braun (Aesculap) | Silkam® Black — natural braided coated multifilament silk, black thread | C0760498 — SILKAM BLACK 1 (4) 75 CM HR30 (verified in official B. Braun catalogue and GUDID) | Comparable by category (not interchangeable 1:1) | They match in material (coated braided silk), USP 1 gauge (EP 4), black colour, ½ circle curvature, 30 mm HR30 needle with conical point, 75 cm thread length and single strand format. They differ in the exact coating (wax/paraffin vs. wax and silicone), sterilisation method (gamma radiation vs. ethylene oxide) and commercial format (box of 36 vs. 12). Confirm IFU and CE/MDR classification before substituting. |
| Ethicon (J&J MedTech) | MERSILK™ / PERMA-HAND™ Silk — coated black braided silk, non-absorbable | W723H — MERSILK USP 1, 75 cm, MO-5 needle of ½ circle conical point (verified in GUDID and official Ethicon EMEA page) | Comparable by category (not interchangeable 1:1) | They match in material (wax-coated braided silk), USP 1 gauge, black colour, 75 cm length, ½ circle curvature and conical point family. The 31 mm MO-5 needle falls within the ±1 mm tolerance compared to the 30 mm HR30. They differ in the exact coating (wax without declared silicone) and commercial format (box of 36). Confirm IFU and CE/MDR classification before substituting. |
There is no manufacturer-validated equivalence table for this specific configuration (black braided silk, 30 mm HR30 Prime needle in ½ circle with conical point, USP 1 gauge and 75 cm strand); consult the technical department before using in a tender. Other homologous families on the market that must be reviewed property by property (material, gauge, curvature, point family, needle and strand length, number of strands, colour, coating and CE/MDR regulatory classification) before any substitution are: DemeTECH DemeSILK™, Medtronic Sofsilk™, LOOK™ Silk (Corza Medical) and Assut Surgisilk™. This needle geometry (HR30 Prime, USP 1, 75 cm black strand) is only fully safely interchangeable within the PREMISUT system itself.
5. Selection tips and variants of interest
Tissue to suture: USP 1 (EP 4) gauge provides high tensile strength for subcutaneous tissue, fasciae, aponeuroses and soft tissues under medium-to-high tension, always with a conical-point needle to minimise tissue damage. For finer planes or paediatric use, it can be stepped down to USP 0 (Ref. AMP542HR3010) or USP 2/0 (Ref. AMP542HR3020) within the same HR30 geometry.
Prime vs. standard quality: the Prime range offers maximum sharpening precision and batch-to-batch consistency; if unit cost is a critical factor, the standard range with identical geometry and gauge (Ref. AM542HR3001) maintains the same dimensional specifications with standard sharpening performance.
Double-thread presentation: if the protocol requires two strands for continuous sutures without re-threading or to split the material across two fields, the same needle and gauge are available in a double thread 2 × 75 cm (Ref. AMP542RHR3001.2).
Needle length: the 30 mm HR30 profile covers shallow to medium-depth access and is ideal for reduced fields in skin surgery, ophthalmology and lax tissues. For fields with greater travel, the same range offers the HR37 (Ref. AMP542HR3701) and HR40 (Ref. AMP542HR4001) profiles, allowing a single supplier across the entire PREMISUT Prime silk family.
Sterility verification: sterile product by ethylene oxide and single-use; check the blister integrity and expiry date before each procedure, keep the reference P542HR3001 together with the batch number for traceability in the operating theatre and store below 30 °C, away from moisture and direct sunlight. For special B2B conditions by volume, hospital supply agreements or validation of equivalents from other commercial houses, contact the technical-commercial team at Cahnos.