Description
1. Commercial Introduction
The polypropylene suture Premisut Prime that Cahnos makes available to surgical services, purchasing centres and tenders is the reference choice when a non-absorbable thread with predictable behaviour, high tensile strength and minimal tissue reaction is needed. This presentation in USP 4/0 gauge mounts a half-circle needle with cylindrical body and grooved Taperpoint conical point, 20 mm (HR20 Prime profile), and a highly visible blue strand of 75 cm: a classic and versatile configuration, highly valued in cardiovascular surgery, ophthalmology, neurosurgery and in the approximation and ligation of soft tissues.
It is supplied sterile and ready for single use in a box of 12 units with individual sachets and batch traceability, which facilitates hospital stock management and submission to public tenders. It is marketed under the factory reference P732HR2040 (SKU Cahnos AMP732HR2040), maintained without altering characters.
2. Technical Explanation
Polypropylene is a synthetic linear polyolefin (propylene polymer) presented as an uncoated monofilament, dyed blue for maximum visibility in the surgical field. As it has no interstices and no wicking effect, it offers very smooth passage through tissue, does not favour bacterial adherence or the formation of sinuses in wounds at risk of contamination, and minimises suture extrusion. It is not absorbed or degraded by tissue enzymes: it is gradually encapsulated by fibrous connective tissue and maintains its tensile strength for more than three months, comfortably covering the critical healing phase.
The HR20 Prime needle (half circle, cylindrical body) with grooved Taperpoint conical point separates and displaces tissue fibres without cutting them, making it ideal for soft, friable tissues and thin-walled vascular structures requiring controlled penetration and minimal injury. The Prime finish provides a grooved needle body and improved penetration, facilitating grip in the needle holder and smooth, consistent passage during knot tying. Sterilisation by ethylene oxide (EO); single-use product, not resterilisable.
3. Technical Specifications
3.1. General characteristics
| Parameter | Specification |
| Brand / Line | Premisut Prime (polypropylene, premium range) |
| Thread material | Polypropylene monofilament, uncoated, dyed blue (synthetic linear polyolefin) |
| Thread structure | Ribbed monofilament of a single filament, without coating or braiding |
| Absorption | Non-absorbable; gradual encapsulation; wound support for over 3 months |
| Gauge / USP size | USP 4/0 (metric EP 1.5); conforms to USP and European Pharmacopoeia gauges |
| Approximate thread diameter | 0.150 – 0.199 mm |
| Thread length | 75 cm (1 thread per sachet) |
| Thread colour | Grooved blue identifier, high visibility in the surgical field |
| Needle shape | ½ circle (1/2 circle) |
| Needle body | Cylindrical |
| Tip type | Conical Taperpoint, ribbed (HR20 Prime profile) |
| Needle size | 20 mm (HR20 Prime) |
| Manufacturing quality | Prime (premium finish, improved penetration) |
| Sterilisation | Ethylene oxide (EO) – sterile |
| Use | Single-use, sterile. Do not re-sterilise |
| Indications | Approximation and ligation of soft tissues; cardiovascular surgery, ophthalmology and neurosurgery |
| Commercial presentation | Box of 12 units (sterile individual sachets); batch traceability |
| Storage | Below 30 °C, protected from moisture and direct sunlight |
| Factory reference | P732HR2040 |
| Reference Cahnos (SKU) | AMP732HR2040 |
3.2. Gauge equivalence according to USP/EP Pharmacopoeia (synthetic non-absorbable suture)
| USP size | Metric measurement (EP) | Approximate thread diameter |
| 4/0 | 1.5 | 0.150 – 0.199 mm |
3.3. Premisut HR20 family (½ circle, cylindrical body, Taperpoint conical point) available at Cahnos
| Needle profile | USP size | Presentation | Reference |
| HR20 Prime | 4/0 | 1 × 75 cm | P732HR2040 / AMP732HR2040 (this product) |
| HR20 (standard) | 4/0 | 1 × 75 cm | 732HR2040 / AM732HR2040 |
| HR20 Prime | 3/0 | 1 × 75 cm | P732HR2030 / AMP732HR2030 |
| 2×HR20 Prime | 4/0 | 2 × 90 cm (double needle) | P733HR2040-2 / AMP733HR2040-2 |
| 2×HR20 Prime VisiBlack | 4/0 | 2 × 90 cm (black needle) | P733HR2040-2VB / AMP733HR2040-2VB |
| HR25 Prime | 4/0 | 1 × 75 cm | P732HR2540 / AMP732HR2540 |
4. Equivalences and Cross-References
Reference P732HR2040 corresponds to the official nomenclature of the Premisut Prime range (HR20 profile with Taperpoint conical point, USP 4/0 gauge, blue 75 cm thread) and Cahnos markets it without altering characters. Blue non-absorbable polypropylene monofilament is one of the most standardised permanent suture families on the market (commercially ‘Prolene’ type). There is no manufacturer-validated equivalence table cross-referencing this reference with other brands; comparisons between brands are functional, at category level, and not interchangeable 1:1 without prior technical validation against each manufacturer’s technical data sheet or IFU.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
| Premisut Prime (Cahnos) | Polypropylene ½ circle HR20 Taperpoint conical point, USP 4/0, blue, 75 cm | P732HR2040 / AMP732HR2040 | Exact reference equivalence (current product) | Official reference of the product analysed in this data sheet; code verified in the Cahnos internal catalogue without altering characters. |
| Premisut (Cahnos) | Polypropylene ½ circle HR20, USP 4/0, blue, 75 cm (standard range) | 732HR2040 / AM732HR2040 | Direct range equivalent | Same HR20 needle, gauge, colour and length; the needle finish changes (standard versus Prime) |
| Premisut Prime (Cahnos) | Polypropylene ½ circle 2×HR20, USP 4/0, blue, 2×90 cm (double needle) | P733HR2040-2 / AMP733HR2040-2 | Configuration equivalent (same system) | Same HR20 needle, gauge and Prime quality; two armed ends of 90 cm for continuous suture or anastomosis |
| Ethicon (J&J MedTech) | PROLENE® – blue polypropylene monofilament, 4-0, 75 cm, SH tapered needle (taper point) of ½ circle, 26 mm | 8831H | Comparable by category (not interchangeable 1:1) | Material, gauge, colour and length match; SH needle of 26 mm versus HR20 of 20 mm. Code verified in official Ethicon catalogue and authorised distribution; presentation in box of 36 units. |
| B. Braun (Aesculap) | Premilene® – blue polypropylene monofilament, USP 4/0 (EP 1.5), 75 cm, DRT26 needle round body conical point | C2090533 (GTIN 04046963633769) | Comparable by category (not interchangeable 1:1) | Material, gauge and length match; DRT26 needle of 26 mm versus HR20 of 20 mm. Verified in AccessGUDID and B. Braun catalogue. |
| Medtronic (Covidien) | Surgipro™ II – polypropylene/polyethylene copolymer, blue, 4-0, 30 inches (≈75 cm), V-20 taper point needle, half circle, 26 mm | VP-831-X | Comparable by category (not interchangeable 1:1) | Same gauge, colour and length; copolymer material (not pure polypropylene) and V-20 needle of 26 mm versus HR20 of 20 mm. Verified in authorised distribution. |
| DemeTECH | DemeLENE™ – blue polypropylene monofilament, USP 4/0, 75 cm, ½ circle needle conical point of 24 mm | PM294024C0P | Comparable by category (not interchangeable 1:1) | Gauge, colour and length match; conical point needle of 24 mm versus HR20 of 20 mm; needle body without grooved channel. Verified in official DemeTECH store. |
Mandatory technical note: there is no manufacturer-validated equivalence table cross-referencing reference P732HR2040 with other brands; comparisons with PROLENE®, Premilene®, Surgipro™ II or DemeLENE™ are functional, at category level, and must be confirmed against each manufacturer’s technical data sheet or IFU before use in tendering or direct substitution. Consult the Cahnos technical department to validate any equivalence before substituting the product.
5. Selection tips and variants of interest
Tissue to suture: the HR20 needle with grooved Taperpoint conical point is indicated for soft, vascular and friable tissues requiring non-cutting penetration (cardiovascular surgery, ophthalmology, neurosurgery). For firm tissues or skin, consider trocar point (HRT) or triangular (DS) profiles from the same Premisut range.
Gauge: USP 4/0 (metric 1.5) offers an excellent balance between strength and manageability for fine tissues and delicate vascular structures. With the same HR20 Prime needle it is also available in gauge 3/0 (ref. AMP732HR2030) for layers requiring greater strength.
Needle size: the 20 mm HR20 is ideal for superficial and medium-thickness surgical fields, very common in ophthalmology and vascular microsurgery. If greater reach is needed in deeper planes, the HR25 is available in the same gauge (ref. AMP732HR2540); for smaller or paediatric fields, consider HR17 or HR15 profiles from the same family.
Presentation: if you prefer to make the most of each opening with two armed ends for continuous suture or anastomosis, choose the double-needle version 2×HR20 of 90 cm (ref. AMP733HR2040-2) or its VisiBlack variant with black needle (ref. AMP733HR2040-2VB). As an economical alternative, the standard Premisut quality (ref. AM732HR2040) maintains needle, gauge, colour and length.
Prime range: the Premisut Prime line maintains the geometry of the standard version with improved needle finish (grooved body and superior penetration), useful when the protocol requires premium quality in delicate tissues.
Prior verification: always check the full reference (HR20 needle, USP size 4/0 and 75 cm length) on the packaging, as well as the integrity of the blister and the expiry date before each use; retain the reference P732HR2040 and the batch number for traceability in the operating theatre. Product sterilised by ethylene oxide and single-use; store below 30 °C, protected from moisture and direct sunlight. For special B2B terms by volume or hospital supply agreements, contact the technical-commercial team at Cahnos.





Reviews