Description
1. Commercial Introduction
The NYLON Suture PREMISUT 1/2 Triangular HS17mm 3/0 Black 75cm is one of the most requested configurations in the range of non-absorbable sutures from Cahnos for B2B supply: operating theatres, purchasing departments, general and specialised surgery services, ophthalmology clinics and neurosurgery units. It combines a nylon monofilament (polyamide 6 or 6.6) of USP 3/0 gauge with a ½ circle needle with a 17 mm reverse cutting triangular point, the classic combination for approximating skin and soft tissues under moderate tension. Sterile, ready-to-use product, presented in a box of 12 units in individual sachets with batch traceability. Ideal for hospital replenishment, tenders and volume contracts seeking a permanent thread with an excellent quality-price ratio.
2. Technical Explanation
Nylon (polyamide 6 or 6.6) is a non-absorbable synthetic polymer that remains in the tissue indefinitely: it gradually becomes encapsulated by fibrous connective tissue and retains high tensile strength for more than three months, comfortably covering the critical healing phase. Being an uncoated monofilament with a grooved finish, the strand slides easily through tissues, reduces the saw effect and bacterial adherence compared to multifilaments, minimises the inflammatory reaction and promotes a firm, stable knot with excellent retention.
The HS17 needle of 17 mm features a ½ circle curvature with a reverse cutting triangular point: the cutting edge cuts in the opposite direction to the wound bed, creating a clean entry hole that preserves the skin edge and prevents the cut from widening, facilitating passage through firm, non-distensible tissues as well as in reduced and deep fields. The USP 3/0 gauge (metric EP 2) provides the optimal balance between fineness and strength for skin, subcutaneous tissue, mucous membranes, ophthalmology and neurosurgery. Sterilisation by ethylene oxide (EO); single-use product, not re-sterilisable.
3. Technical Specifications
3.1. General characteristics
| Parameter | Specification |
| Factory reference | 642HS1730 |
| Reference Cahnos (SKU) | AM642HS1730 |
| Brand / quality | PREMISUT (standard range) |
| Thread material | Nylon monofilament (polyamide 6 / 6.6), uncoated |
| Thread structure | Monofilament, grooved finish |
| Absorption | Non-absorbable; gradual encapsulation, wound support for more than 3 months |
| Gauge / size | USP 3/0 (metric 2 / EP 2), conforming to USP gauges and European Pharmacopoeia |
| Approximate thread diameter | 0.20 – 0.249 mm |
| Thread length | 75 cm (free end, without loop) |
| Thread colour | Black identifier (high visibility in surgical field) |
| Needle shape | ½ circle (1/2 circle), fluted body |
| Tip type | Reverse cutting triangular (Reverse Cutting), triangular body and point |
| Needle size | HS 17 (17 mm), 1 needle per thread |
| Sterilisation | Ethylene oxide (EO); sterile, single-use, non-resterilisable |
| Indications | Approximation of soft tissues, ligatures, ophthalmology and neurosurgery |
| Commercial presentation | Box of 12 units (sterile individual sachets); batch traceability |
| Regulatory standard | Conforms to USP/EP Pharmacopoeia for non-absorbable sutures; CE marking (MDR) |
3.2. Gauge equivalence according to USP/EP Pharmacopoeia (synthetic non-absorbable suture)
| USP size | Metric measurement (EP) | Approximate thread diameter |
| 4/0 | 1.5 | 0.15 – 0.199 mm |
| 3/0 (this product) | 2 | 0.20 – 0.249 mm |
| 2/0 | 3 | 0.30 – 0.339 mm |
The strand diameter complies with the official USP Pharmacopoeia ranges for non-absorbable sutures, without exceeding their limits. Minimal tissue reaction, excellent tensile strength, secure knot and predictable behaviour as there is no polymer degradation in the body.
3.3. PREMISUT HS17 family (½ circle needle, 17 mm, triangular reverse cutting point, black 75 cm thread)
| Needle profile | USP size | Thread | Reference |
| HS 17 (standard) | 3/0 | 1 × 75 cm (black) | 642HS1730 / AM642HS1730 (this product) |
| HS 17 (standard) | 4/0 | 1 × 75 cm (black) | 642HS1740 / AM642HS1740 |
| HS 17 Prime | 3/0 | 1 × 75 cm (black) | P642HS1730 / AMP642HS1730 |
| HS 17 Prime | 4/0 | 1 × 75 cm (black) | P642HS1740 / AMP642HS1740 |
Reference decoding: 642 (black thread of 75 cm) + HS17 (½ circle needle with reverse cutting triangular point of 17 mm) + 30 (USP 3/0 gauge). The 641 series corresponds to 45 cm thread and the 643 series to 90 cm. Cahnos markets the official reference without altering characters: no digit may be added, removed or rearranged when ordering or tendering.
4. Equivalences and Cross-References
The reference 642HS1730 corresponds to the official nomenclature of the PREMISUT range (HS17 profile: ½ circle needle with reverse cutting triangular point of 17 mm, USP 3/0 gauge, black grooved thread of 75 cm). Non-absorbable black nylon monofilament is one of the most standardised permanent suture families on the market (commercial type «Ethilon»). However, there is no equivalence table validated by the manufacturer that cross-references this specific configuration (½ circle HS17 needle of 17 mm, USP 3/0, black, 75 cm) with references from other manufacturers.
During verification in official catalogues, the manufacturer’s store and regulatory databases (GUDID/FDA), functional comparisons were found that do not enter the table because they do not meet all matching fields and, therefore, are not interchangeable 1:1: ETHILON® EH7563H (Ethicon, 3-0, 75 cm, black, but PS-1 needle of 24 mm in ⅜ circle: curvature and needle size differ), DemeLON™ NL263020B0P (DemeTECH, 3/0, black, 75 cm, ½ circle, but DSH-2 needle of 20 mm round body: point family and size differ) and DAFILON® C0935352 (B. Braun, 3/0, 75 cm, but blue and DS24 needle of 24 mm in ⅜ circle: colour, curvature and size differ). None has been included as substitutable; they must be reviewed property by property against the technical data sheet or IFU of each manufacturer before any substitution.
Homologous families of the segment that must be reviewed (material, gauge, ½ circle curvature, reverse cutting point family, needle length of 17 mm within tolerance ±1 mm, thread length of 75 cm, black colour, assembly of 1 needle and regulatory classification CE/MDR) before any substitution: ETHILON® (Ethicon / J&J MedTech), DAFILON® (B. Braun), MONOSOF™ and DERMALON™ (Medtronic), DemeLON™ (DemeTECH), LUXAMID® (LuxSutures) and SERALON® (Serag-Wiessner). The cited brands belong to their respective manufacturers and are mentioned solely as an equivalence reference.
There is no equivalence table validated by the manufacturer; consult the technical department before using in tenders. This needle geometry (HS17, USP 3/0, 75 cm) is only safely interchangeable within the PREMISUT system itself: the direct equivalent within the same manufacturer is the Prime version of identical geometry and gauge (Ref. P642HS1730, SKU AMP642HS1730).
5. Selection tips and variants of interest
- Tissue to suture: the USP 3/0 gauge (EP 2) is the reference size for skin, subcutaneous tissue, mucous membranes and closures under moderate tension. The HS17 reverse cutting needle passes through firm tissues without tearing the puncture bed, being indicated in general surgery, ophthalmology, neurosurgery and ligatures. For finer planes, consider the 4/0 of the same HS17 needle (AM642HS1740 / AMP642HS1740).
- Prime quality: if the protocol requires maximum sharpening precision and batch-to-batch consistency, the same geometry and gauge are available in the PREMISUT Prime range (Ref. P642HS1730, SKU AMP642HS1730), a direct equivalent within the same manufacturer.
- Needle length: the HS17 profile of 17 mm in ½ circle covers reduced and deep fields (ophthalmology, neurosurgery, oral surgery). For wider access, the same family is available in HS26 and HS36 profiles, and the DS family of ⅜ circle for large superficial fields.
- Other presentations: the same nylon range offers threads of 45 cm (series 641) and 90 cm (series 643), as well as spatulated needles (DSP/HSP/VSP) for precision ophthalmic surgery and conical tip versions (HR) for lax or vascular tissues.
- Sterility and storage: product sterile by ethylene oxide and single-use; check the integrity of the blister and the expiry date before each use, and keep the reference 642HS1730 together with the batch number for traceability in the operating theatre. Store below 30 °C, away from humidity and direct sunlight. For B2B conditions by volume or hospital supply agreements, contact the technical-commercial team of Cahnos.





Reviews