Description
1. Introduction
Maintain the highest light quality in your diagnostic, laboratory and production equipment with the Cermax PE175BFA 175 W xenon lamp, manufactured by Excelitas Technologies (formerly PerkinElmer). It is the reference replacement in the Cermax family for equipment that does not require the intensity of the 300 W versions: stable daylight white light with a continuous spectrum and excellent colour rendering. A reliable and durable solution for hospital maintenance services, R&D departments, endoscope fleets and medical and industrial fibre optic lighting systems.
2. Technical explanation
The PE175BFA lamp uses short-arc xenon discharge technology: light is generated in a high-pressure electric arc burning in inert xenon gas inside a ceramic body with a ceramic-metal seal, a design that is much safer, more compact and more stable than conventional quartz lamps. Its integrated 1-inch parabolic reflector concentrates the beam into a very small focal point, optimising coupling to light guides and optical fibres with minimal losses. It operates on direct current (DC) at 175 W nominal (range 150–200 W), with a voltage of 12.5 V and a current draw of 14 A, and offers a visible flux of 2,200 lumens with a colour temperature of 5,900 K, very close to daylight. It is built without mercury or problematic materials (ceramic, metal and sapphire) and supports instant on/off with an ignition voltage of 23–35 kV.
3. Technical Specifications
Validated technical features of the Cermax PE175BFA lamp (Excelitas factory data):
| Characteristic | Value |
|---|---|
| Type | Short-arc xenon lamp (discharge) |
| Brand / Series | Excelitas Technologies (Cermax series) |
| Model | PE175BFA |
| Power | 175 W (range 150–200 W) |
| Voltage | 12.5 V DC (range 11–14 V) |
| Amperage | 14 A DC (range 12–16 A) |
| Ignition voltage | 23–35 kV |
| Luminous flux (390–770 nm) | 2,200 lumens |
| Colour temperature | 5900 K |
| Radiant output | 25 W |
| UV output (<390 nm) | 1.2 W |
| IR output (>770 nm) | 14 W |
| Peak intensity | 350,000 candelas |
| Stability | 4 % peak instability |
| Optics | Parabolic reflector integrated in ceramic body |
| Window diameter | 25.4 mm (1 inch) |
| Beam geometry | 5° / 6° / 7° |
| Focused output (f/1.0, aperture 3 mm) | 830 lumens / 8 W |
| Focused output (f/1.0, aperture 6 mm) | 1,400 lumens / 13 W |
| Average service life | 1,000 hours |
| Maximum working temperature | 150 °C |
| Weight | 131 g |
| Construction | Ceramic, metal and sapphire; mercury-free |
| Operation | Direct current (DC), instant on/off |
| Presentation | Unit (ref. Cahnos 50003403) |
Main applications:
- Fibre optic illuminators for medical and industrial use
- Microscopy and spectroscopy
- UV polymerisation, video projection and solar simulation
- Machine vision, spotlights and beacons of visible and infrared light
- Wafer inspection (semiconductors)
4. Equivalences and Cross-References
The Cermax PE175BFA lamp is marketed under multiple codes depending on the manufacturer and distribution channel. The most common functional equivalents are:
| Code | Origin / destination |
|---|---|
| PE175BFA | Excelitas Cermax reference (general distribution) |
| UXR-175 | Ushio equivalent |
| CL70023 | Carley equivalent |
| XBO C 175W/BF (55202) | OSRAM equivalent of the XBO C series in ceramic |
It is the lower-power version of the Cermax short-arc family, designed for equipment that does not require the intensity of its 300 W sibling. Within the Cahnos catalogue it coexists with the EXCELITAS Cermax PE300BFA (ref. 50003401), the EXCELITAS J2022 for Olympus CLV-190 (ref. 50014955), the Y1964 xenon lamp (ref. 50021071) and the OSRAM XBO R 180W/45 C OFR (ref. 50002956 and 50002957).
5. Selection tips and variants of interest
Before replacing the lamp, note the hours recorded on the light source counter and carry out the preventive replacement when approximately the recommended 1,000 hours are reached to avoid drops in intensity during the procedure. Always handle the lamp with gloves or a clean, dry cloth: fingerprints on the glass window create hot spots and drastically reduce service life. Work with the equipment switched off and cold, verify that the housing is free of dust and respect the polarity, the mounting orientation (never with the window facing upwards at less than 45° from the vertical) and the operating range (150–200 W) indicated by the manufacturer. It is recommended to power it with regulated current or power sources with ripple below 5 % and single ignition pulses, since radiofrequency starters can damage the internal reflector.
If your application requires higher luminous flux, the 300 W Cermax PE300BFA (ref. 50003401) is the natural alternative; if you need the specific reference for Olympus CLV-180/190/290 towers, the EXCELITAS J2022 (ref. 50014955) guarantees the original MAJ-1817 mounting. For scheduled maintenance orders or large volumes, our technical-commercial team can advise you on identifying the exact replacement for your equipment fleet.





Reviews