Description
1. Commercial Introduction
The Fiocchetti PLASMA-VISION 700 TOUCH plasma freezer (reference Cahnos FIO-DMV0702, factory code DMV0702) is a large vertical column-type unit, specifically designed for the safe storage of frozen plasma bags and blood products at temperatures down to -20 °C. Manufactured in Italy by Fiocchetti Scientific S.r.l., it is certified as a class IIa medical device in accordance with Directive 93/42/EEC and its amendment 2007/47/EC, is registered in the Medical Devices Registry of the Italian Ministry of Health and complies with Directive 2011/65/EU (RoHS).
With a net capacity of 620 litres and dimensions of 75 x 87 x 210 cm, it is the reference solution for blood banks, transfusion departments and high-volume laboratories that need to store blood components with full thermal traceability and regulatory compliance. It is supplied under formal quotation; consult our technical-commercial team to configure it with the appropriate internal accessories.
2. Technical explanation of the equipment
The PLASMA-VISION 700 is a freezer with professional forced-air refrigeration that distributes cold evenly throughout the chamber, with adaptive automatic defrosting and condensate evaporation, which eliminates manual maintenance, prevents ice formation and minimises thermal fluctuations. Its high-density injected polyurethane foam insulation (43 kg/m³, 75 mm thick), CFC-free and with ecological water injection, guarantees excellent energy performance together with the Night & Day saving function.
Its tempered glass door according to safety standards IEC 61010-1 offers excellent thermal insulation and allows checking the frozen material from the outside without opening the door; the anodised aluminium handle ensures a secure grip and the opening direction is reversible.
The ECT-F TOUCH microprocessor controller with 7″ touch screen simultaneously displays a real-time graph, the current temperature, the programmed setpoint and the alarm limits (both editable), the energy consumption and the battery status, with a resolution of 0.1 °C and an accuracy of ±0.3 °C. The unit includes as standard the independent recording module DMLP TOUCH with backup battery, which maintains the record of temperatures and alarms even in the event of a power outage, the product temperature simulation probe, the electronic lock with generic or per-user password, dry contacts for remote alarm and RJ45 port with integrated web server and MODBUS TCP/IP protocol. The acoustic and visual alarms, which can be silenced, are automatically recorded with date, time, type, duration and maximum temperature reached. It is always possible to integrate the optional GSM and Wi-Fi modules, as well as local or cloud monitoring systems.
3. Technical Specifications
3.1. General data
| Parameter | Specification |
|---|---|
| Field of application | Storage and preservation of plasma and blood products intended for transfusion |
| Temperature range | -10 / -20 °C (adjustable) |
| Electronic controller | ECT-F TOUCH (7″ touchscreen) |
| Nominal capacity | 700 lt (24.72 cu ft) |
| Net capacity | 620 lt (21.89 cu ft) |
| External dimensions (HxDxW) | 75 x 87 x 210 cm |
| Internal dimensions | 60 x 68 x 150 cm |
3.2. Structure and refrigeration
| Parameter | Specification |
|---|---|
| Door | Tempered glass according to IEC 61010-1, anodised aluminium handle, reversible opening |
| Insulation type | High-density injected polyurethane foam in the walls (43 kg/m³), CFC-free and with ecological water injection |
| Insulation thickness | 75 mm |
| Structure | Steel with white antibacterial plasticised PVC (interior and exterior) or optional interior in 18/10 AISI 304 steel; rounded corners and chamber with bottom |
| Support | 4 height-adjustable feet for maximum stability |
| Internal lighting | Full LED across the full height of the chamber, with intensity adjustment |
| Cooling type | Professional ventilated |
| Defrosting type | Adaptive automatic with condensate evaporation |
| Energy saving | Night & Day function |
3.3. Technical data of the refrigerator
| Parameter | Value |
|---|---|
| Refrigerant gas type | R290 |
| Energy consumption | 6 kWh/24h |
| Power consumption | kW 0.35 / Amp 2.5 |
| Starting absorption | Amp 7 |
| Noise level | <55 dBA |
| Heat emission (EN 12900) | W 250 |
| Shelves included | 3 |
| Max. no. of shelves | 7 |
| Shelf dimensions | cm 56 x 65 |
| Max. no. of drawers | 7 |
| Internal dimension of drawers with dividers (L x D x H) x number of compartments | cm 15 x 60 x 7.5 (3) |
| Internal dimension of drawers without dividers (L x D x H) | cm 49 x 60 x 7.5 |
| Max. number of steel wire drawers | 7 |
| Internal dimension of steel wire drawers (L x D x H) | cm 53.8 x 63 x 14.8 |
| Max. number of steel wire drawers | 7 |
| Internal dimension of steel wire drawers (L x D x H) | cm 53.8 x 49 x 6.2 |
3.4. ECT-F TOUCH controller and safety
| Characteristic | Description |
|---|---|
| Display | 7″ touch screen with real-time graph; resolution 0.1 °C and accuracy ±0.3 °C |
| Registration | Temperatures and alarms with date, time, type, duration and maximum temperature; history on SD card |
| Ports | SD card (data history), USB (system download and update) and SIM card (for optional GSM module) |
| Connectivity | Integrated web server for LAN (RJ45) and MODBUS TCP/IP; dry contacts for remote alarm |
| DMLP TOUCH module | Independent recording with backup battery (operates during power cut); product temperature simulation catheter |
| Safety | Anti-freezing and anti-overheating devices; electronic lock with generic or user-specific password; silenceable acoustic and visual alarms |
| Certifications | Medical device class IIa (Directive 93/42/CEE and amendment 2007/47/CEE); Directive 2011/65/CE (RoHS); door according to IEC 61010-1 |
| Optional accessories | GSM and Wi-Fi modules; local and cloud monitoring systems |
3.5. Available references
| Model | Fiocchetti code | Controller | Temperature range | Capacity | External dimensions (HxDxW) |
|---|---|---|---|---|---|
| PLASMA-VISION 700 TOUCH | DMV0702 | ECT-F TOUCH | -10 / -20°C | 620 lt (21.89 cu ft) | 75 x 87 x 210 |
4. Equivalences and Cross-References
The manufacturer’s official reference for this unit is DMV0702 (model PLASMA-VISION 700 TOUCH), verified in the official technical data sheet of Fiocchetti Scientific S.r.l.; the corresponding Cahnos SKU is FIO-DMV0702. Within the same series with tempered glass door, the manufacturer offers the PLASMA-VISION 400 TOUCH (DMV0402, SKU FIO-DMV0402), with identical controller and certification.
There is no manufacturer-validated equivalence table linking the PLASMA-VISION 700 to specific codes from other plasma freezer manufacturers; consult the technical department before using it in a tender. The comparable brands and families in the sector for this category (plasma and blood product freezers, -20/-30 °C range, with continuous recording and alarms, for blood bank use) include Angelantoni Life Science, Kirsch (Medical/Froster line), Liebherr (Medline range for blood products), Gram and Labcold. They are comparable at category level, not interchangeable 1:1: each manufacturer has different chamber geometry, controller, internal equipment, consumption and certifications. Before any substitution, compare field by field the thermal range, net capacity, dimensions, chamber material, type of controller and alarm recording, refrigerant gas and regulatory classification of the actual product, validating it with the Cahnos technical department.
Within the same manufacturer, the PLASMA-FREEZER TOUCH series (solid door, down to -25 °C) offers capacities from 140 to 1500 litres: DMF0142, DMF0252, DMF0402, DMF0702 and DMF1502; the PLASMA-SUPERARTIC TOUCH series (solid door, down to -40 °C) offers models DMS0252, DMS0402 and DMS0702. For dual-compartment needs (refrigerated blood and frozen plasma), see the PLASMA-LABOR 2T / PLASMA-VISION 2T series.
5. Selection tips and variants of interest
Installation space: allow access for a column-type unit of 75 x 87 x 210 cm (WxDxH) with 4 adjustable feet, and ensure adequate ventilation around the unit taking into account its 75 mm insulation. Check doorways, corridors and delivery platform before placing the order.
Traceability: if your specification requires independent continuous recording and alerts in the event of power cuts, this model includes it as standard with the DMLP TOUCH module with backup battery, product temperature simulation catheter, dry contacts and RJ45 port. For hospital network integration, request the GSM/Wi-Fi modules and local or cloud monitoring.
Internal configuration: the unit accepts up to 7 shelves (includes 3 as standard, 56 x 65 cm) and up to 7 drawers with or without dividers, plus up to 7 steel wire drawers, allowing the organisation of fresh frozen plasma, cryoprecipitate and blood product bags to be adapted to the workflow of your service. Racks and accessories are quoted separately.
Variants in the same family: the PLASMA-VISION TOUCH series with tempered glass door offers the PLASMA-VISION 400 TOUCH (DMV0402) for smaller volumes. If you need freezing down to -25 °C or -40 °C, consider the PLASMA-FREEZER TOUCH and PLASMA-SUPERARTIC TOUCH series from the same manufacturer. Sales are made under a formal quotation; contact our technical-commercial team to size the correct solution and receive the regulatory documentation (technical data sheet, CE certificate) for tenders and competitions.