Description
1. Commercial Introduction
The combined blood and plasma bank Fiocchetti PLASMA-LABOR 2T 400 TOUCH (ref. Cahnos FIO-DML0402, factory code DML0402) is the dual-temperature solution in a single vertical unit: it stores blood at positive temperature (+2/+6 °C) and frozen plasma at -10/-25 °C (usual setpoint -20 °C) in two completely independent compartments, each with its own refrigerating unit and solid closing door.
With a capacity of 358 litres (200+200 nominal litres), it is certified as a class IIa medical device in accordance with Directive 93/42/CEE and its amendment 2007/47/CEE, it is registered in the Italian Ministry of Health’s Medical Devices Register and complies with standard DIN 13277 and Directive RoHS 2011/65/CE. It is the reference choice for blood banks, transfusion services and laboratories that require rigorous thermal traceability and regulatory compliance in tenders and specifications. It is supplied on order and formal quotation; consult our technical-commercial team to configure the interior.
2. Technical explanation of the equipment
The PLASMA-LABOR 2T series integrates two independent refrigeration circuits in a single vertical cabinet, allowing it to simultaneously maintain refrigerated blood and frozen plasma without risk of cross-contamination of temperatures. The refrigeration is professional forced-air: cold air is distributed homogeneously throughout the chamber, with adaptive automatic defrosting and condensate evaporation, eliminating manual maintenance and reducing thermal fluctuations. The high-density injected polyurethane foam insulation (43 kg/m³, 70 mm at the rear and 50 mm at the sides), CFC-free and with ecological water injection, guarantees a contained energy performance of 4 kWh/24h.
The microprocessor controller ECT-F TOUCH with 7″ touch screen simultaneously displays a real-time graph, the current temperature, the programmed setpoint and the alarm limits (both editable), the energy consumption and the battery status, with a resolution of 0.1 °C and an accuracy of ±0.3 °C. It includes as standard the independent recording module DMLP TOUCH with backup battery, which maintains the temperature and alarm log even during a power outage, the product temperature simulation catheter, the electronic lock with generic or individual user password, dry contacts for remote alarm and RJ45 port with integrated web server and MODBUS TCP/IP protocol. The acoustic and visual alarms, which can be silenced, are automatically recorded with date, time, type, duration and maximum temperature reached. It is always possible to integrate the optional GSM and Wi-Fi modules and local or cloud monitoring systems.
3. Technical Specifications
Structure and refrigeration
| Parameter | Specification |
|---|---|
| Configuration | Combined refrigerator + freezer, 2 independent compartments with 2 refrigerating units and 2 solid doors |
| Temperature range | Compartment A (blood): +2/+6 °C | Compartment B (plasma): -10/-25 °C |
| Capacity | 358 lt (12.64 cu ft) |
| Nominal capacity | 200+200 lt (7.06+7.06 cu ft) |
| Net capacity | 170+160 lt (6.01 + 5.66 cu ft) |
| Internal dimensions | 50 x 48.5 x 68.5 / 50 x 48.5 x 73 cm |
| External dimensions (HxDxW) | 60 x 64.5 x 202 cm |
| Insulation type | High-density injected polyurethane foam (43 kg/m³), CFC-free and with ecological water injection |
| Insulation thickness | 70 mm rear / 50 mm on the sides |
| Structure | Steel with white antibacterial plastified PVC (interior and exterior) or optional interior in 18/10 AISI 304 steel; rounded corners and chamber with bottom |
| Mobility | 4 unidirectional wheels included + 2 front stability feet |
| Internal lighting | Full LED across the full height of the chamber, with intensity adjustment |
| Cooling type | Professional ventilated |
| Defrosting type | Adaptive automatic with condensate evaporation |
| Energy saving | Night & Day function |
Electronic controller ECT-F TOUCH
| Characteristic | Description |
|---|---|
| Display | 7″ touch screen with real-time graph; resolution 0.1 °C and accuracy ±0.3 °C |
| Registration | Temperatures and alarms with date, time, type, duration and maximum temperature; history on SD card |
| Ports | SD card (data history), USB (system download and update) and SIM card (for optional GSM module) |
| Connectivity | Integrated web server for LAN (RJ45) and MODBUS TCP/IP; dry contacts for remote alarm |
| DMLP TOUCH module | Independent recording with backup battery (works during power outage); product temperature simulation catheter |
| Safety | Anti-freezing and anti-overheating devices; electronic lock with generic or per-user password; silencable acoustic and visual alarms |
| Optional accessories | GSM and Wi-Fi modules; local and cloud monitoring systems |
Technical data of the refrigerator
| Parameter | Value |
|---|---|
| Refrigerant gas type | R290 |
| Consumption | 4 kWh/24h |
| Absorbed power | kW 0.17+0.32 / Amp 1+2.3 |
| Starting absorption | Amp 3+5 |
| Noise level | <50 dBA |
| Heat emitted (EN 12900) | 166 W |
| Shelves included / max. | 4 (2+2) / max. 3+3 — dimensions cm 46×45 |
| Drawers max. | 3+3 — with dividers (LxPxH) x number of spaces: cm 11.8x40x7.5 (3); without dividers: cm 39x40x7.5 |
| Stainless steel wire drawers max. | 3+3 — internal dimensions (LxPxH): cm 43.5×40.5×6.2 |
| Classification | Medical device class IIa (Directive 93/42/CEE and amendment 2007/47/CEE); standard DIN 13277; RoHS 2011/65/CE |
Available references
| Model | Fiocchetti code | Controller | Temperature range | Capacity | External dimensions (HxDxW) |
|---|---|---|---|---|---|
| PLASMA-LABOR 2T 400 TOUCH | DML0402 | ECT-F TOUCH | +2/+6°C | -10/-25°C | 358 L (12.64 cu ft) | 60 x 64.5 x 202 cm |
| PLASMA-LABOR 2T 280 TOUCH | DML0282 | ECT-F TOUCH | +2/+6°C | -10/-20°C | 306 lt (10.81 cu ft) | 52 x 66.5 x 177.5 cm |
| PLASMA-LABOR 2T 500 TOUCH | DML0502 | ECT-F TOUCH | +2/+6°C | -10/-25°C | 486 lt (17.16 cu ft) | 72 x 74.5 x 208 cm |
| PLASMA-LABOR 2T 700 TOUCH | DML0702 | ECT-F TOUCH | +2/+6°C | -10/-30°C | 593 L (20.94 cu ft) | 72 x 84 x 203 cm |
4. Equivalences and Cross-References
The factory code for this unit is DML0402 (model PLASMA-LABOR 2T 400 TOUCH), verified in the official technical documentation from Fiocchetti Scientific S.r.l.; the corresponding Cahnos SKU is FIO-DML0402. The variants in the same series and from the same manufacturer are DML0282 (2T 280), DML0502 (2T 500) and DML0702 (2T 700), identical in controller and certification and differing only in capacity and dimensions.
There is no equivalence table validated by the manufacturer linking this model to specific codes from other manufacturers of combined refrigerator-freezer cabinets for blood banks; consult the technical department before using it in a tender. The homologous brands and families in the sector for this category (dual-temperature blood + plasma cabinets, positive range +2/+6 °C and frozen -25/-30 °C, with continuous recording and alarms, for blood bank and transfusion use) include Angelantoni Life Science (blood and plasma banks with dual circuit), Kirsch (medical refrigeration and freezing lines for blood products), Liebherr (Medline range for laboratory and blood bank), Gram and Labcold. They are comparable at category level, not interchangeable 1:1: each manufacturer has different chamber geometry, controller, internal fittings (shelves, drawers, baskets), consumption, refrigerant gas and certifications. Before any replacement, compare field by field the thermal range of both compartments, net capacity, external and internal dimensions, chamber material, type of controller and alarm recording, and the regulatory classification (class IIa) of the actual product, validating it with the Cahnos technical department.
5. Selection tips and variants of interest
Installation space: allow access for a unit measuring 60 x 64.5 x 202 cm (WxDxH) and ensure adequate ventilation at the rear (70 mm insulation). Check doorways, corridors and delivery platform before placing the order.
Traceability and compliance: if your specification requires independent continuous recording and alerts in the event of power cuts, this model includes it as standard with the DMLP TOUCH module with backup battery, product temperature simulation probe, dry contacts and RJ45. For hospital network integration, request the GSM/Wi-Fi modules and cloud monitoring.
Internal configuration: the unit accepts up to 3+3 shelves (46×45 cm) and 3+3 drawers (with or without dividers) and stainless steel wire drawers, allowing you to organise blood bags (red cell concentrates, platelets) and fresh frozen plasma according to your service flow. The racks and accessories are quoted separately.
Variants in the same family: if your patient volume is lower, the PLASMA-LABOR 2T series offers the 280 model (DML0282, 306 litres); if you need more capacity, the 500 (DML0502, 486 litres) and 700 (DML0702, 593 litres) models share the same controller and certification. If you only need to store frozen plasma, see the PLASMA-FREEZER TOUCH series (DMF0142, DMF0252, DMF0402, DMF0702 and DMF1502). Sales are made under a formal quotation; contact our technical-commercial team to size the correct solution and receive the regulatory documentation (technical data sheet, CE/MDR certificate) for tenders and competitions.























Reviews