Description
1. Commercial Introduction
The combined blood and plasma bank Fiocchetti PLASMA-LABOR 2T 280 TOUCH (ref. Cahnos FIO-DML0282, factory code DML0282) is the dual-temperature solution in a single vertical unit: it stores blood at positive temperature (+2/+6 °C) and frozen plasma at -10/-20 °C (usual setpoint -20 °C) in two completely independent compartments, each with its own refrigeration unit and solid closing door.
With a total capacity of 306 litres (140+140 nominal), it is certified as a class IIa medical device in accordance with Directive 93/42/EEC and its amendment 2007/47/EC, is registered in the Medical Devices Register of the Italian Ministry of Health and complies with standard DIN 13277 and Directive RoHS 2011/65/EC. It is the reference choice for blood banks, transfusion services, small and large laboratories that need to store blood derivatives with full thermal traceability and regulatory compliance in tenders and specifications. It is supplied on order and formal quotation; consult our technical-commercial team to configure the interior.
2. Technical explanation of the equipment
The PLASMA-LABOR 2T series integrates two independent refrigeration circuits in a single vertical cabinet, allowing refrigerated blood and frozen plasma to be maintained simultaneously without the risk of cross-contamination of temperatures. The cooling is professional forced-air: cold air is distributed homogeneously throughout the chamber, with adaptive automatic defrosting and condensate evaporation, eliminating manual maintenance and reducing thermal fluctuations. The Night & Day function contributes to energy savings, and the high-density injected polyurethane foam insulation (43 kg/m³, 40 mm thick), CFC-free and with ecological water injection, guarantees a contained consumption of 4.6 kWh/24h.
The microprocessor controller ECT-F TOUCH with 7″ touchscreen simultaneously displays a real-time graph, current temperature, programmed setpoint and alarm limits (both editable), energy consumption and battery status, with a resolution of 0.1 °C and accuracy of ±0.3 °C. It includes as standard the independent recording module DMLP TOUCH with backup battery, which maintains the temperature and alarm log even during a power outage, the product temperature simulation probe, the electronic lock with generic or per-user password, dry contacts for remote alarm and RJ45 port with integrated web server and MODBUS TCP/IP protocol. The acoustic and visual alarms, which are silenceable, are automatically recorded with date, time, type, duration and maximum temperature reached. It is always possible to integrate the optional GSM and Wi-Fi modules and local or cloud monitoring systems.
3. Technical Specifications
Structure and refrigeration
| Parameter | Specification |
|---|---|
| Configuration | Combined refrigerator + freezer, 2 independent compartments with 2 refrigeration units and 2 solid doors |
| Temperature range | Compartment A (blood): +2/+6 °C | Compartment B (plasma): -10/-20 °C |
| Capacity | 306 lt (10.81 ft³) |
| Nominal capacity | 140+140 lt (4.94 + 4.94 ft³) |
| Net capacity | 128+128 lt (4.53 + 4.53 ft³) |
| Internal dimensions | 44 x 51.6 x 56 cm (x 2) |
| External dimensions (HxDxW) | 52 x 66.5 x 177.5 cm |
| Insulation type | High-density injected polyurethane foam (43 kg/m³), CFC-free and with ecological water injection |
| Insulation thickness | 40 mm |
| Structure | White antibacterial plasticised PVC steel (interior and exterior) or optional interior in 18/10 AISI 304 steel; rounded corners and chamber with base |
| Mobility | 4 unidirectional castors included + 2 front feet for maximum stability |
| Internal lighting | Full LED across the full height of the chamber, with intensity adjustment |
| Cooling type | Professional ventilated |
| Defrosting type | Adaptive automatic with condensate evaporation |
| Energy saving | Night & Day function |
ECT-F TOUCH electronic controller
| Characteristic | Description |
|---|---|
| Display | 7″ touchscreen with real-time graph; resolution 0.1 °C and accuracy ±0.3 °C |
| Registration | Temperatures and alarms with date, time, type, duration and maximum temperature; history on SD card |
| Ports | SD card (data history), USB (system download and update) and SIM card (for optional GSM module) |
| Connectivity | Integrated web server for LAN (RJ45) and MODBUS TCP/IP; dry contacts for remote alarm |
| DMLP TOUCH module | Independent recording with backup battery (works during power failure); product temperature simulation probe |
| Safety | Anti-freezing and anti-overheating devices; electronic lock with generic or per-user password; silenceable acoustic and visual alarms |
| Optional accessories | GSM and Wi-Fi modules; local and cloud monitoring systems |
Technical data of the refrigerator
| Parameter | Value |
|---|---|
| Refrigerant gas type | R290 |
| Consumption | 4.6 kWh/24h |
| Absorbed power | kW 0.14+1.18 / Amp 0.9+1.2 |
| Starting absorption | Amp 2.5+3.5 |
| Noise level | <45 dBA |
| Heat emitted (EN 12900) | W 191 |
| No. of shelves included | 4 (2+2) |
| Maximum number of shelves | 3+3 |
| Shelf dimensions | cm 40×49 |
| Maximum number of drawers | 5+3 |
| Internal dimension of drawers with dividers (LxDxH) x number of spaces | cm 10x45x7.5 (3) |
| Internal dimension of drawers without dividers (LxDxH) | cm 33x45x7.5 |
| Classification | Class IIa medical device (Directive 93/42/EEC and amendment 2007/47/EC); standard DIN 13277; RoHS 2011/65/EC |
Available references
| Model | Fiocchetti code | Controller | Temperature range | Capacity | External dimensions (HxDxW) |
|---|---|---|---|---|---|
| PLASMA-LABOR 2T 280 TOUCH | DML0282 | ECT-F TOUCH | +2/+6°C | -10/-20°C | 306 lt (10.81 cu ft) | 52 x 66.5 x 177.5 cm |
| PLASMA-LABOR 2T 400 TOUCH | DML0402 | ECT-F TOUCH | +2/+6°C | -10/-25°C | 358 L (12.64 cu ft) | 60 x 64.5 x 202 cm |
| PLASMA-LABOR 2T 500 TOUCH | DML0502 | ECT-F TOUCH | +2/+6°C | -10/-25°C | 486 lt (17.16 cu ft) | 72 x 74.5 x 208 cm |
| PLASMA-LABOR 2T 700 TOUCH | DML0702 | ECT-F TOUCH | +2/+6°C | -10/-30°C | 593 L (20.94 cu ft) | 72 x 84 x 203 cm |
4. Equivalences and Cross-References
The manufacturer’s factory code for this equipment is DML0282 (model PLASMA-LABOR 2T 280 TOUCH), verified in the official technical documentation of Fiocchetti Scientific S.r.l.; the corresponding Cahnos SKU is FIO-DML0282. The variants of the same series and from the same manufacturer are DML0402 (2T 400), DML0502 (2T 500) and DML0702 (2T 700), identical in controller and certification and differing only in capacity, thermal range of the freezing compartment and dimensions.
There is no equivalence table validated by the manufacturer linking this model to specific codes from other manufacturers of combined refrigerator-freezer cabinets for blood banks; consult the technical department before using in tenders. The equivalent brands and families in the sector for this category (dual-temperature blood + plasma cabinets, positive range +2/+6 °C and frozen -20/-30 °C, with continuous recording and alarms, for use in blood banks and transfusion) include Angelantoni Life Science (blood and plasma banks with dual circuit), Kirsch (medical refrigeration and freezing lines for blood products), Liebherr (Medline range for laboratory and blood bank), Gram and Labcold. They are comparable at category level, not interchangeable 1:1: each manufacturer has different chamber geometry, controller, internal fittings (shelves, drawers, baskets), consumption, refrigerant gas and certifications. Before any substitution, compare field by field the thermal range of both compartments, net capacity, external and internal dimensions, chamber material, type of controller and alarm recording, and the regulatory classification (class IIa) of the actual product, validating it with the technical department of Cahnos.
5. Selection tips and variants of interest
Installation space: provide access for a unit measuring 52 x 66.5 x 177.5 cm (W x D x H) and ensure adequate ventilation at the rear (40 mm insulation). Check doorways, corridors and delivery platform before formalising the order.
Traceability and regulations: if your specification requires independent continuous recording and alerts in the event of power cuts, this model includes it as standard with the DMLP TOUCH module with backup battery, product temperature simulation probe, dry contacts and RJ45. For integration into the hospital network, request the GSM/Wi-Fi modules and cloud monitoring.
Internal configuration: the unit supports up to 4 shelves as standard (40×49 cm), up to 3+3 shelves and up to 5+3 drawers with or without dividers, allowing you to organise blood bags (red blood cell concentrates, platelets) and fresh frozen plasma according to the flow of your service. Racks and accessories are quoted separately.
Variants in the same family: if your patient volume is higher, the PLASMA-LABOR 2T series offers models 400 (DML0402, 358 litres), 500 (DML0502, 486 litres) and 700 (DML0702, 593 litres), with identical controller and certification. If you only need to store frozen plasma, see the PLASMA-FREEZER TOUCH series (DMF0142, DMF0252, DMF0402, DMF0702 and DMF1502). Sales are made under a formal quotation; contact our technical-commercial team to size the correct solution and receive the regulatory documentation (technical data sheet, CE certificate) for tenders and competitions.