Description
1. Introduction
The Fiocchetti EMOTECA 500 TOUCH blood bank (factory code DME0502) is the high-capacity solution of the EMOTECA series for the safe storage of whole blood, plasma, blood derivatives and donated breast milk at a constant temperature of +4 °C. Designed and certified as a medical device, it is the ideal equipment for hospitals, blood transfusion services and centres, blood banks, nursing homes and blood collection centres that demand reliability, traceability and compliance with the most restrictive health regulations, both national and international. It is sold on a quotation basis: Fiocchetti configures each unit to order according to the centre’s needs.
2. Technical explanation of the equipment
The Fiocchetti EMOTECA series incorporates a microprocessor electronic controller ECT-F TOUCH with a 7″ touch screen that displays in real time the temperature (resolution 0.1 °C and accuracy ±0.3 °C), the programmed set point and editable alarm limits, a real-time graph, energy consumption and battery status. As standard it includes the DMLP TOUCH module with backup battery, which allows independent recording of temperatures and alarms even during a power outage, the product temperature simulation probe, the electronic lock for access control (with generic or user password) and the terminal box for remote alarm dry contacts and RJ45 port with integrated web server (LAN and MODBUS TCP/IP connection). Optional GSM and Wi-Fi modules can be integrated, as well as local and cloud monitoring systems.
The professional forced-air cooling ensures uniform cold distribution, with adaptive automatic defrosting with condensate evaporation and the Night \u0026 Day energy-saving function. The acoustic and visual alarms, which can be silenced, cover all abnormal conditions — also in case of power outage thanks to the backup battery — and are automatically recorded with date, time, type, duration and maximum temperature reached. Its 7 drawers with dividers for blood bags facilitate organisation, access and traceability of the stored product. Fiocchetti blood banks are medical devices that comply with Directive 93/42/EEC and its amendment 2007/47/EC (class IIa), are registered in the Medical Devices Register of the Italian Ministry of Health and comply with standards DIN 13277 and 2011/65/EC (RoHS).
3. Technical Specifications
Available references
| Model | Fiocchetti code | Controller | Temperature range | Capacity | External dimensions (HxDxW) |
|---|---|---|---|---|---|
| EMOTECA 500 TOUCH | DME0502 | ECT-F TOUCH | +4°C | 527 L (18.61 cu ft) | 72 x 76 x 208 cm |
Structure and refrigeration
| Characteristic | Specification |
|---|---|
| Internal dimensions | 60 x 58 x 150 cm |
| Blood bag capacity (450 ml) | No. 273 |
| Nominal capacity | 500 L (17.65 cu ft) |
| Net capacity | 527 L (18.61 cu ft) |
| Insulation type | High-density injected polyurethane foam in the walls (43 kg/m³), CFC-free and with ecological water injection |
| Insulation thickness | 75 mm rear / 60 mm sides |
| Structure | White antibacterial PVC-coated steel (interior and exterior) or option of internal structure in 18/10 AISI 304 steel; rounded corners and chamber with bottom |
| Stability and mobility | 4 unidirectional castors included + 2 front feet for maximum stability |
| Internal lighting | Full LED along the entire height of the chamber with intensity adjustment |
| Cooling type | Professional ventilated |
| Defrosting type | Adaptive automatic with condensate evaporation |
| Energy saving | Night & Day function |
Electronic controller with 7″ touch screen
| Characteristic | Specification |
|---|---|
| Controller | ECT-F TOUCH (microprocessor with 7″ touch screen) |
| Resolution / accuracy | 0.1 °C / ±0.3 °C |
| Ports (window with 3 ports) | SD card (data history), USB (system download and update), SIM (for optional GSM module) |
| Connectivity | Integrated web server for LAN connection (RJ45) and MODBUS TCP/IP |
| Safety | Anti-freezing and anti-overheating safety device; dry contacts for remote alarm |
| Standard equipment | Electronic lock, DMLP TOUCH module with backup battery, product temperature simulation probe and terminal box for dry contacts and RJ45 port |
Technical data of the refrigerator
| Characteristic | Specification |
|---|---|
| Refrigerant gas type | R290 |
| Consumption | 1 kWh/24h |
| Absorbed power | kW 0.17 / Amp 1 |
| Starting absorption | Amp 3 |
| Noise level | 45 dBA |
| Heat output EN 12900 | 40 W |
| No. of drawers included | 7 |
| Maximum number of drawers | 7 |
| Internal dimension of drawers with dividers (LxPxH) (x number of spaces) | 15 x 50 x 7.5 cm (3) |
4. Equivalences and Cross-References
There is no equivalence table validated by the manufacturer Fiocchetti against other brands; it is recommended to consult with the technical department of Cahnos before using it in a tender. The references of other manufacturers indicated below are comparable by category (blood bank refrigerators at +4 °C), not interchangeable 1:1, and require verification of specifications (capacity, dimensions, controller and standards) before any replacement.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
|---|---|---|---|---|
| B Medical Systems | Blood bank refrigerator B401 | B401 | Comparable by category (not interchangeable 1:1) | Single-chamber refrigerator, gross volume 478 L / storage 390 L, 270 bags of 450 ml, +4 °C, gas R600a, compliant with AABB and EU MDR class IIa. Differs in capacity and refrigerant. |
| Helmer Scientific | i.Series Blood Bank Refrigerator (double door) | iBR245-GX / iBR256-GX | Comparable by category (not interchangeable 1:1) | Double-door blood refrigerators at +4 °C with i.C3 touch screen monitor, AABB certified. Differ in capacity and refrigeration technology. |
5. Selection tips and variants of interest
Before choosing, consider the volume of bags to be stored (this model holds 273 blood bags of 450 ml), the available floor space (72 x 76 x 208 cm), the noise level and consumption (45 dBA / 1 kWh per 24 h) and the need for remote monitoring and access control, elements that this unit covers as standard with the touch controller, the DMLP TOUCH module, the electronic lock and the RJ45 port.
Within the single-chamber EMOTECA family, Fiocchetti offers models with lower and higher capacity to suit each facility: EMOTECA 100 TOUCH (DME0102), EMOTECA 140 TOUCH (DME0142), EMOTECA 170 TOUCH (DME0172), EMOTECA 200 TOUCH (DME0202), EMOTECA 250 TOUCH (DME0252), EMOTECA 400 (DME0402), EMOTECA 700 (DME0702) and EMOTECA 1500 (DME1502). When cold redundancy is critical, the EMOTECA TWIN series (DMW0252, DMW0402, DMW0702, DMW1502) incorporates two independent refrigeration units with alternating function. To separate products into independent chambers, the dual-chamber, dual-door EMOTECA 2T range (DMB0282, DMB0402, DMB0502) is the recommended alternative. All incorporate the ECT-F TOUCH touch controller; always check compatibility with your local or cloud monitoring system via the RJ45 port. The equipment is custom-configured and sold on quotation.






















Reviews