Description
1. Commercial Introduction
The combined blood and plasma bank Fiocchetti PLASMA-VISION 2T 700 TOUCH (ref. Cahnos FIO-DMD0702, factory code DMD0702) is the dual-temperature solution in a single vertical unit with two glass doors: it stores blood at positive temperature (+2/+6 °C) and frozen plasma at -10/-20 °C in two completely independent compartments, each with its own refrigerating unit. With a nominal capacity of 350+350 litres, it is certified as a class IIa medical device in accordance with Directive 93/42/EEC and its amendment 2007/47/EC, is registered in the Medical Devices Register of the Italian Ministry of Health and complies with RoHS Directive 2011/65/EU.
The two glass doors allow visual supervision of the contents without opening the unit, minimising cold losses, while the customisable interior (shelves, drawers and baskets) optimises the storage of blood bags, fresh frozen plasma and blood derivatives. It is the reference choice for blood banks, transfusion services and small and large volume laboratories that require rigorous thermal traceability. It is supplied on order and formal quotation; consult our technical-commercial team to configure the interior.
2. Technical explanation of the equipment
The PLASMA-VISION 2T series integrates in a single vertical cabinet two independent refrigeration circuits, allowing simultaneous storage of refrigerated blood (usual +4 °C) and frozen plasma (-20 °C) without risk of cross-contamination of temperatures. The refrigeration is professional forced-air: cold air is distributed homogeneously throughout the chamber, with adaptive automatic defrosting and condensate collection, eliminating manual maintenance and reducing thermal fluctuations.
The insulation of high-density injected polyurethane foam (43 kg/m³, with a thickness of 60 mm), CFC-free and with ecological water injection, guarantees a contained energy performance of 7 kWh/24h, complemented by the Night & Day saving function. The monobloc structure is made of white plasticised steel with antibacterial film, or optional interior in 18/10 AISI 304 stainless steel, with rounded interior corners and 4 height-adjustable feet.
The ECT-F TOUCH microprocessor controller with 7″ touch screen simultaneously displays a real-time multi-trace graph, the current temperature, the programmed setpoint and the alarm limits (both editable), the energy consumption and the battery status, with a resolution of 0.1 °C and an accuracy of ±0.3 °C. It includes as standard the independent recording module DMLP TOUCH with backup battery, which maintains the temperature and alarm record even during a power outage, the product temperature simulation catheter, the electronic lock with generic or individual password per user, dry contacts for remote alarm and RJ45 port with integrated web server and MODBUS TCP/IP protocol. The acoustic and visual alarms, which can be silenced, are automatically recorded with date, time, type, duration and maximum temperature reached. It is always possible to integrate the optional GSM and Wi-Fi modules and local or cloud monitoring systems.
3. Technical Specifications
3.1. Structure and refrigeration
| Configuration | Combined refrigerator + freezer, 2 independent compartments with 2 refrigeration units and 2 glass doors |
| Temperature range | Compartment A (blood): +2/+6 °C | Compartment B (plasma): -10/-20 °C |
| Nominal capacity | 350+350 lt (12.36 + 12.36 cu ft) |
| Net capacity | 620 lt (21.89 cu ft) |
| Internal dimensions | 60 x 68 x 70 / 60 x 68 x 74 cm |
| External dimensions (HxDxW) | 72 x 85.5 x 203 cm |
| Insulation type | High-density injected polyurethane foam (43 kg/m³), CFC-free and with ecological water injection |
| Insulation thickness | 60 mm |
| Structure | Steel with white antibacterial plasticised PVC (interior and exterior) or optional interior in 18/10 AISI 304 steel; rounded corners and chamber with base. 4 height-adjustable feet |
| Internal lighting | Full LED across the full height of the chamber, with intensity adjustment |
| Cooling type | Professional ventilated |
| Defrosting type | Adaptive automatic with condensate collection |
| Energy saving | Night & Day function |
3.2. Electronic controller ECT-F TOUCH and safety
| Display | 7″ touchscreen with real-time graph; resolution 0.1 °C and accuracy ±0.3 °C |
| Registration | Temperatures and alarms with date, time, type, duration and maximum temperature; history on SD card |
| Ports | SD card (data history), USB (download and system update) and SIM card (for optional GSM module) |
| Connectivity | Integrated web server for LAN (RJ45) and MODBUS TCP/IP; dry contacts for remote alarm |
| DMLP TOUCH module | Independent recording with backup battery (operates during power outage); product temperature simulation probe |
| Safety | Anti-freezing and anti-overheating devices; electronic lock with generic or per-user password; silenceable acoustic and visual alarms |
| Optional accessories | GSM and Wi-Fi modules; local and cloud monitoring systems |
3.3. Technical data of the refrigerator
| Parameter | Value |
| Refrigerant gas type | R290 |
| Energy consumption | 7 kWh/24h |
| Absorbed power | kW 0.17+0.32 / Amp 1+2.3 |
| Starting absorption | Amp 3+5 |
| Noise level | <50 dBA |
| Heat emission (EN 12900) | W 291 |
| Number of shelves included | 4 (2+2) |
| Max. no. of shelves | 3+3 |
| Shelf dimensions | cm 56×55 |
| Max. no. of drawers | 5+3 |
| Internal dimension of drawers with dividers (LxDxH) x number of spaces | cm 15x60x7.5 (3) |
| Internal dimension of drawers without dividers (LxWxH) | cm 49x60x7.5 |
| Max. number of stainless steel baskets | 3+3 |
| Internal dimensions of the stainless steel baskets (LxPxH) | cm 53.8x63x14.8 |
| Max. no. of stainless steel wire drawers | 5+3 |
| Internal dimension of stainless steel wire drawers (LxDxH) | cm 53.8x49x6.2 |
| Classification | Medical device class IIa (Directive 93/42/CEE and amendment 2007/47/CEE); RoHS 2011/65/CE |
3.4. Available references of the series
| Model | Fiocchetti code | Controller | Temperature range | Capacity | External dimensions (HxDxW) |
| PLASMA-VISION 2T 700 TOUCH | DMD0702 | ECT-F TOUCH | +2/+6°C | -10/-20°C | 593 L (20.94 cu ft) | 72 x 85.5 x 203 |
4. Equivalences and Cross-References
The factory code of this equipment is DMD0702 (model PLASMA-VISION 2T 700 TOUCH), verified in the manufacturer’s official technical data sheet and on its corporate website; the corresponding Cahnos SKU is FIO-DMD0702. The variants of the same series and from the same manufacturer with glass doors are DMD0502 (2T 500, 486 litres) and DMD0402 (2T 400, 358 litres), identical in controller and certification and differing only in capacity and dimensions. As variants from the same manufacturer with solid door, the PLASMA-LABOR 2T series offers DML0702 (593 litres, -10/-30 °C), DML0502 and DML0402, and the PLASMA-SUPERARTIC 2T series the DMT0702 (-20/-40 °C).
There is no manufacturer-validated equivalence table linking this model to specific codes from other manufacturers of combined blood bank refrigerator-freezer cabinets with glass doors; consult the technical department before use in tenders. The sector’s homologous brands and families for this category (dual-temperature blood + plasma cabinets, positive range +2/+6 °C and frozen -20/-25 °C, with continuous recording and alarms, for blood bank and transfusion use) include Angelantoni Life Science (blood and plasma banks with dual circuit), Kirsch (medical refrigeration and freezing lines for blood products), Liebherr (Medline range for laboratory and blood bank), Gram and Labcold. They are comparable at category level, not interchangeable 1:1: each manufacturer has different chamber geometry, controller, internal fittings (shelves, drawers, baskets), door type, consumption, refrigerant gas and certifications. Before any replacement, compare field by field the thermal range of both compartments, net capacity, external and internal dimensions, chamber material, controller and alarm recording type, and the regulatory classification (class IIa) of the actual product, validating it with the Cahnos technical department.
5. Selection tips and variants of interest
Installation space: allow access for a unit measuring 72 x 85.5 x 203 cm (WxDxH) and ensure adequate ventilation at the rear (60 mm insulation). Check doorways, corridors and delivery platform before placing the order; the wheel kit (2 with brake) is available as an accessory for model 700.
Traceability and compliance: if your specification requires independent continuous recording and alerts in the event of power cuts, this model includes it as standard with the DMLP TOUCH module with backup battery, product temperature simulation probe, dry contacts and RJ45. For hospital network integration, request the GSM/Wi-Fi modules, IoT gateway, humidity probe or PT100 probe, and cloud monitoring.
Internal configuration: the unit supports up to 3+3 shelves (includes 4 as standard), 5+3 drawers with or without dividers, 3+3 stainless steel baskets and 5+3 wire drawers, allowing you to organise blood bags (red blood cell concentrates, platelets) and fresh frozen plasma according to your service flow. Racks and accessories are quoted separately.
Variants in the same family: if your patient volume is lower, the PLASMA-VISION 2T series offers models 500 (DMD0502, 486 litres) and 400 (DMD0402, 358 litres) with the same controller and certification. If you only need to store frozen plasma, see the PLASMA-VISION series (DMV0702 and DMV0402) or the PLASMA-FREEZER TOUCH series; if your centre’s regulations require a solid door or deeper ranges, consider the PLASMA-LABOR 2T series (DML0702, -10/-30 °C). Sales are made under a formal quotation; contact our technical-commercial team to size the correct solution and receive the regulatory documentation (technical data sheet, CE/MDR certificate) for tenders and bids.






















Reviews