Description
1. Commercial Introduction
The Fiocchetti PLASMA-LABOR 2T 500 TOUCH combined blood and plasma bank (ref. Cahnos FIO-DML0502, factory code DML0502) is the dual-temperature solution in a single vertical unit: it stores blood at positive temperature (+2/+6 °C) and frozen plasma at -10/-25 °C (usual setpoint -20 °C) in two completely independent compartments, each with its own refrigeration unit and solid closing door.
With a total capacity of 486 litres (250+250 nominal), it is certified as a class IIa medical device in accordance with Directive 93/42/EEC and its amendment 2007/47/EC, registered in the Italian Ministry of Health’s Medical Devices Register, and complies with standard DIN 13277 and Directive RoHS 2011/65/EU. It is the reference choice for blood banks, transfusion services and medium-to-high volume laboratories that demand rigorous thermal traceability and regulatory compliance in tenders and bids. It is supplied on order and formal quotation; consult our technical-commercial team to configure the interior.
2. Technical explanation of the equipment
The PLASMA-LABOR 2T series integrates two independent refrigeration circuits in a single vertical cabinet, allowing refrigerated blood and frozen plasma to be maintained simultaneously without risk of cross-contamination of temperatures. The cooling is professional forced-air: cold air is distributed homogeneously throughout the chamber, with adaptive automatic defrosting and condensate evaporation, eliminating manual maintenance and reducing thermal fluctuations. The high-density injected polyurethane foam insulation (43 kg/m³, 75 mm at the rear and 60 mm at the sides), CFC-free and with ecological water injection, ensures a contained consumption of 4.5 kWh/24h.
The ECT-F TOUCH microprocessor controller with 7″ touchscreen simultaneously displays a real-time graph, current temperature, programmed setpoint and alarm limits (both editable), energy consumption and battery status, with a resolution of 0.1 °C and accuracy of ±0.3 °C. It includes as standard the independent recording module DMLP TOUCH with backup battery that maintains the temperature and alarm record even during a power outage, the product temperature simulation probe (instead of ambient), electronic lock with generic or individual password per user, dry contacts for remote alarm and RJ45 port with integrated web server and MODBUS TCP/IP protocol. The acoustic and visual alarms, which can be silenced, are automatically recorded with date, time, type, duration and maximum temperature reached. It is always possible to integrate the optional GSM and Wi-Fi modules and local or cloud monitoring systems.
3. Technical Specifications
Structure and refrigeration
| Parameter | Specification |
|---|---|
| Configuration | Combined refrigerator + freezer, 2 independent compartments with 2 refrigeration units and 2 solid doors |
| Temperature range | Compartment A (blood): +2/+6 °C | Compartment B (plasma): -10/-25 °C |
| Nominal capacity | 250+250 litres (8.82+8.82 cu ft) |
| Net capacity | 249+230 litres (8.80 + 8.13 cu ft) |
| Internal dimensions | 60 x 58.8 x 70 / 60 x 58.8 x 74 cm |
| External dimensions (HxDxW) | 72 x 74.5 x 208 cm |
| Insulation type | High-density injected polyurethane foam (43 kg/m³), CFC-free, with ecological water injection |
| Insulation thickness | 75 mm rear / 60 mm sides |
| Structure | Steel with white antibacterial plastified PVC (interior and exterior), rounded corners and chamber with base |
| Mobility | 4 unidirectional wheels + 2 front stability feet |
| Internal lighting | Full LED across the full height of the chamber, with intensity adjustment |
| Cooling type | Professional ventilated |
| Defrosting type | Adaptive automatic with condensate evaporation |
| Energy saving | Night & Day function |
ECT-F TOUCH electronic controller
| Characteristic | Description |
|---|---|
| Display | 7″ touchscreen with real-time graph; resolution 0.1 °C and accuracy ±0.3 °C |
| Registration | Temperatures and alarms with date, time, type, duration and maximum temperature; history on SD card |
| Ports | SD card, USB (download and update) and SIM card (for optional GSM module) |
| Connectivity | Integrated web server for LAN (RJ45) and MODBUS TCP/IP; dry contacts for remote alarm |
| DMLP TOUCH module | Independent recording with backup battery (operates during power outage); product temperature simulation probe |
| Safety | Anti-freezing and anti-overheating devices; electronic lock with generic or per-user password; silenceable acoustic and visual alarms |
| Optional accessories | GSM and Wi-Fi modules; local and cloud monitoring systems |
Technical data of the refrigerator
| Parameter | Value |
|---|---|
| Refrigerant gas type | R290 |
| Consumption | 4.5 kWh/24h |
| Absorbed power | kW 0.17+0.32 / Amp 1+2.3 |
| Starting absorption | Amp 3+5 |
| Noise level | <50 dBA |
| Heat emitted (EN 12900) | 187 W |
| Shelves included / max. | 4 (2+2) / max. 3+3 — dimensions cm 56×55 |
| Drawers max. | 3+3 — with dividers (LxDxH): cm 15x50x7.5 (3 spaces); without dividers: cm 49x50x7.5 |
| Classification | Class IIa medical device (Directive 93/42/EEC and amendment 2007/47/EC); standard DIN 13277; RoHS 2011/65/EC |
Available references in the series
| Model | Fiocchetti code | Controller | Temperature range | Capacity | External dimensions (HxDxW) |
|---|---|---|---|---|---|
| PLASMA-LABOR 2T 280 TOUCH | DML0282 | ECT-F TOUCH | +2/+6°C | -10/-20°C | 306 lt (10.81 cu ft) | 52 x 66.5 x 177.5 cm |
| PLASMA-LABOR 2T 400 TOUCH | DML0402 | ECT-F TOUCH | +2/+6°C | -10/-25°C | 358 L (12.64 cu ft) | 60 x 64.5 x 202 cm |
| PLASMA-LABOR 2T 500 TOUCH | DML0502 | ECT-F TOUCH | +2/+6°C | -10/-25°C | 486 lt (17.16 cu ft) | 72 x 74.5 x 208 cm |
4. Equivalences and Cross-References
The manufacturer’s factory code for this equipment is DML0502 (model PLASMA-LABOR 2T 500 TOUCH), verified in Fiocchetti’s official technical documentation; the corresponding Cahnos SKU is FIO-DML0502. The variants of the same series and same manufacturer are DML0282 (2T 280) and DML0402 (2T 400), identical in controller and certification and differing only in capacity and dimensions.
There is no equivalence table validated by the manufacturer relating this model to specific codes from other manufacturers of combined refrigerator-freezer units for blood banks; consult the technical department before using in a tender. The homologous brands and families in the sector for this category (dual-temperature blood + plasma cabinets, positive range +2/+6 °C and frozen -25/-30 °C, with continuous recording and alarms, for blood bank and transfusion use) include Angelantoni Life Science (blood and plasma banks with dual circuit), Kirsch (medical refrigeration and freezing lines for blood products), Liebherr (Medline range for laboratory and blood bank), Gram and Labcold. They are comparable at category level, not interchangeable 1:1: each manufacturer has different chamber geometry, controller, internal equipment (shelves, drawers, baskets), consumption, refrigerant gas and certifications. Before any substitution, compare field by field: thermal range of both compartments, net capacity, external and internal dimensions, chamber material, type of controller and alarm recording, and MDR regulatory classification (class IIa) of the actual product, validating it with the Cahnos technical department.
5. Selection tips and variants of interest
Installation space: provide access for equipment measuring 72 x 74.5 x 208 cm (WxDxH) and ensure adequate ventilation at the rear (75 mm insulation). Check doorways, corridors and delivery platform before placing the order.
Traceability and regulations: if your tender requires independent continuous recording and alerts in the event of power cuts, this model includes it as standard with the DMLP TOUCH module with backup battery, product temperature simulation probe, dry contacts and RJ45. For hospital network integration, request the GSM/Wi-Fi modules and cloud monitoring.
Internal configuration: the equipment accepts up to 3+3 shelves (56×55 cm) and 3+3 drawers with or without dividers, allowing blood bags (red blood cell concentrates, platelets) and fresh frozen plasma to be organised according to the flow of your service. The racks and accessories are quoted separately.
Variants of the same family: if your care volume is lower, the PLASMA-LABOR 2T series offers the 400 model (DML0402, 358 litres) and the 280 model (DML0282, 306 litres) with identical controller and certification. If you only need to store frozen plasma, see the PLASMA-FREEZER TOUCH series (DMF0142, DMF0252, DMF0402, DMF0702 and DMF1502). Sales are made under a formal quotation; contact our technical-commercial team to size the correct solution and receive the regulatory documentation (technical data sheet, CE/MDR certificate) for tenders and bids.