Description
1. Introduction
The blood bank Fiocchetti EMOTECA TWIN 400 TOUCH (code DMW0402) is the mediana capacity solution in the EMOTECA TWIN series for the safe storage of whole blood, plasma, blood derivatives and donated breast milk at a constant temperature of +4 °C. Designed and certified as a class IIa medical device, it is the ideal equipment for hospitals, blood banks, transfusion services and centres, nursing homes and blood collection centres that require reliability, traceability and regulatory compliance. It is marketed on a quotation basis: Fiocchetti configures each unit to order according to the needs of the centre.
2. Technical explanation of the equipment
The EMOTECA TWIN series is equipped with two independent refrigeration units with alternating function: the dual ECT-F TOUCH controller alternately activates the two compressors in each operating cycle. In the event of failure of one of them, the remaining one is activated with an immediate alarm signal to the operator, maintaining performance at the highest level. Each refrigeration system (A or B) is controlled independently and diagnoses its own faults (sensors, condenser overheating, low evaporator temperature, etc.).
It incorporates as standard the microprocessor controller with 7″ touch screen (resolution of 0.1 °C and accuracy of ±0.3 °C) with temperature and alarm recording, real-time graph, editable setpoint and alarm limits, energy consumption and battery status. It also includes the DMLP TOUCH module with backup battery for independent data and alarm recording even in the event of a power cut, the product temperature simulation catheter, the electronic lock for access control, dry contacts for remote alarm and Ethernet/RJ45 port with integrated web server (LAN and MODBUS TCP/IP). Optional GSM and Wi-Fi modules can be integrated, as well as local and cloud monitoring systems.
The professional ventilated refrigeration ensures a uniform distribution of cold with adaptive automatic defrosting with condensate evaporation and the Night & Day energy-saving function. Its 6 drawers with separators for blood bags facilitate the organisation, access and traceability of the stored product. Fiocchetti blood banks comply with Directive 93/42/EEC and its amendment 2007/47/EC (class IIa), are registered in the Italian Ministry of Health’s Medical Devices Register and comply with the DIN 13277 and 2011/65/EC (RoHS) standards.
3. Technical Specifications
Available references
| Model | Fiocchetti code | Controller | Temperature range | Capacity | External dimensions (HxDxW) |
|---|---|---|---|---|---|
| EMOTECA TWIN 400 TOUCH | DMW0402 | ECT-F TOUCH | +4°C | 347 lt (12.25 cu ft) | 60 x 66 x 196 |
Structure and refrigeration
| Characteristic | Specification |
|---|---|
| Internal dimensions | 50 x 48.5 x 140 cm |
| Blood bag capacity (450 ml) | no. 180 |
| Nominal capacity | 400 lt (14.12 cu ft) |
| Net capacity | 347 lt (12.26 cu ft) |
| Insulation type | High-density injected polyurethane foam in the walls (43 kg/m³), CFC-free and with ecological water injection |
| Insulation thickness | 70 mm rear / 50 mm on the sides |
| Structure | Steel with white antibacterial plastified PVC (interior and exterior) or option of internal structure in 18/10 Aisi 304 steel; rounded corners and chamber with base |
| Stability and mobility | 4 unidirectional castors included + 2 front feet for maximum stability |
| Refrigerating units | No. 2 with alternating function (systems A and B with independent control) |
| Internal lighting | Full LED along the entire height of the chamber with intensity adjustment |
| Cooling type | Professional ventilated |
| Defrosting type | Adaptive automatic with condensate evaporation |
| Energy saving | Night & Day function |
Electronic controller with 7″ touch screen
| Characteristic | Specification |
|---|---|
| Controller | ECT-F TOUCH (microprocessor with 7″ touch screen) |
| Resolution / accuracy | 0.1 °C / ±0.3 °C |
| Ports (window with 3 ports) | SD card (data history), USB (system download and update), SIM (for optional GSM module) |
| Connectivity | Integrated web server for LAN network connection (RJ45) and MODBUS TCP/IP |
| Safety | Anti-freezing and anti-overheating safety device; dry contacts for remote alarm |
| Standard equipment / accessories | Electronic lock, DMLP TOUCH module with backup battery, product temperature simulation catheter, acoustic and visual alarm for power failure; optional GSM and Wi-Fi modules |
Technical data of the refrigerator
| Characteristic | Specification |
|---|---|
| Refrigerant gas type | R290 |
| Consumption | 0.7 kWh/24h |
| Absorbed power | kW 0.17 / Amp 1 |
| Starting absorption | Amp 3 |
| Noise level | 45 dBA |
| Heat output EN 12900 | W 30 |
| No. of drawers included | 6 |
| Maximum number of drawers | 6 |
| Internal dimension of drawers with dividers (LxPxH) (x number of spaces) | 11 x 40 x 7.5 cm (3) |
4. Equivalences and Cross-References
There is no equivalence table validated by the manufacturer Fiocchetti against other brands; it is recommended to consult with the technical department of Cahnos before its use in tenders. The references of other manufacturers indicated below are comparable by category (blood bank refrigerators at +4 °C), not interchangeable 1:1, and require verification of specifications (capacity, dimensions, controller and regulations) before any substitution.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
|---|---|---|---|---|
| B Medical Systems | Blood bank refrigerator B401 | B401 | Comparable by category (not interchangeable 1:1) | Single-chamber refrigerator, gross volume 478 L / storage 390 L, 270 bags of 450 ml, +4 °C, R600a gas, compliant with AABB and EU MDR class IIa. It differs in capacity, refrigerant and in not having a dual refrigeration unit. |
| Helmer Scientific | i.Series Blood Bank Refrigerator (double door) | iBR245-GX / iBR256-GX | Comparable by category (not interchangeable 1:1) | Double-door refrigerators for blood at +4 °C with i.C3 touch screen monitor, AABB certified. They differ in capacity and refrigeration technology. |
Within the Fiocchetti range itself, the EMOTECA TWIN series with two independent refrigeration units offers the following variants of the same architecture:
| Model | Fiocchetti code | Net capacity | 450 ml bags | External dimensions (HxDxW) |
|---|---|---|---|---|
| EMOTECA TWIN 250 TOUCH | DMW0252 | 264 lt (9.32 cu ft) | 150 | 60 x 66 x 163.5 cm |
| EMOTECA TWIN 400 TOUCH | DMW0402 | 347 lt (12.25 cu ft) | 180 | 60 x 66 x 196 cm |
| EMOTECA TWIN 700 TOUCH | DMW0702 | 620 lt (21.89 cu ft) | 366 | 75 x 90 x 205.5 cm |
| EMOTECA TWIN 1500 TOUCH | DMW1502 | 1355 lt (47.85 cu ft) | 672 | 144 x 83 x 204 cm |
5. Selection tips and variants of interest
Before choosing, consider the number of bags to be stored (this model holds 180 bags of 450 ml), the available surface area (60 x 66 x 196 cm), the noise level and consumption (45 dBA / 0.7 kWh per 24 h) and the need for remote monitoring and access control, elements that this equipment covers as standard with the touch controller, the DMLP TOUCH module, the electronic lock and the RJ45 port. The dual refrigeration unit with alternating function makes it especially recommended for centres where continuity of cold is critical.
If the storage volume is smaller, the EMOTECA TWIN series offers the 250 model (DMW0252); for large capacity, the EMOTECA TWIN 700 (DMW0702) and EMOTECA TWIN 1500 (DMW1502) models. To separate products in independent chambers, the double-chamber and double-door EMOTECA 2T range (DMB0282, DMB0402, DMB0502) is the indicated alternative. All incorporate the ECT-F TOUCH touch controller; always check compatibility with your local or cloud monitoring system via the RJ45 port. The equipment is configured to order and is marketed on a quotation basis.