Description
1. Disposable linear stapler for digestive tract surgery
The disposable linear surgical stapler WZLS-A-90L is the mechanical suture solution your healthcare facility needs for open surgery procedures on the digestive tract, viscera resection and closure of incisions. Its disposable design eliminates reprocessing costs and guarantees a new, sharp and sterile instrument in each intervention, with the safety and traceability required in the operating theatre. Manufactured with high-resistance titanium staples, it offers a firm and homogeneous seal that reduces postoperative bleeding and shortens suture times compared to manual technique.
2. Technical explanation of the equipment
This device applies two interlocked (staggered) rows of titanium staples that stitch the tissue simultaneously, creating a continuous, uniform suture line without tension differences. The anvil features a narrow design that reduces the risk of damaging peripheral tissue and facilitates access to difficult-to-expose areas. The stapler combines manual and automatic operation with audible signals, allowing the surgeon to control each firing phase and confirm complete closure of the handle before releasing the staples. The 90 mm staple line configuration with blue staples of 3.8 mm height is indicated for normal or thin-wall tissue: closure of incisions, visceral transections, stump closure in gastric or intestinal resections and side-to-side anastomoses. Being single-use, the risk of cross-infection is eliminated and the sharpness of the mechanism is ensured in each application.
3. Technical Specifications
| Product reference | WZLS-A-90L |
| Device type | Disposable linear stapler (open surgery, non-cutting) |
| Staple line length | 90 mm |
| Staple wire diameter | 0.28 mm |
| Staple height | 3.8 mm |
| Staple identification colour | Blue (normal tissue / thin wall) |
| Number of staples per unit | 39 |
| Staple rows | 2 staggered (interlocked) rows of titanium |
| Staple material | Titanium |
| Anvil | Narrow design, protects peripheral tissue |
| Actuation | Manual and automatic with audible signals |
| Use | Single-use (disposable) |
| Sterilisation | Ethylene oxide (EO), individually packaged |
| Application | Closure and suture of incisions, resection of viscera and reconstructive surgery of the digestive tract |
WZLS-A family table (factory references):
| Model | Length (mm) | Staple height (mm) | Colour | Number of staples |
| WZLS-A-30L | 30 | 3.8 | Blue | 13 |
| WZLS-A-30H | 30 | 4.5 | Green | 13 |
| WZLS-A-45L | 45 | 3.8 | Blue | 19 |
| WZLS-A-45H | 45 | 4.5 | Green | 19 |
| WZLS-A-60L | 60 | 3.8 | Blue | 25 |
| WZLS-A-60H | 60 | 4.5 | Green | 25 |
| WZLS-A-90L | 90 | 3.8 | Blue | 39 |
| WZLS-A-90H | 90 | 4.5 | Green | 39 |
4. Equivalences and Cross-References
There is no equivalence table validated by the manufacturer that cross-references the WZLS-A series with the references of major Western brands; consult the technical department before use in tenders. The following references are offered as a functional category comparison verified against each manufacturer’s sources, never as direct 1:1 interchangeables, and their specifications must be validated before any substitution.
| Manufacturer / Brand | Family / Model | Code / SKU | Equivalence level | Technical differences / Remarks |
| Medtronic (Covidien) | TA™ with DST Series™ technology | TA9035S (instrument) / TA9035L (reload) | Comparable by category | 90 mm, blue, titanium, 2 rows; open height 3.5 mm (closed 1.5 mm) versus 3.8 mm; 33 staples versus 39; reloadable instrument up to 8 firings |
| Ethicon (Johnson & Johnson MedTech) | PROXIMATE™ TX | TX 90 series (blue reload, normal tissue) | Not verified 1:1 – comparable by category | 90 mm with blue staples of 3.5 mm for standard tissue; confirm exact reference and number of staples with the manufacturer |
| B. Braun (Aesculap) | LS – Linear Stapler | LS9035 | Comparable by category | 90 mm, blue, double row, open surgery; height 3.5 mm versus 3.8 mm; proprietary structure, verify system compatibility |
| Purple Surgical | Ultimate Reloadable Linear Stapler | PS9LSI90B (instrument) / PS9LSR90B (reload) | Not verified 1:1 – comparable by category | 90 mm, blue 3.5 mm for standard tissue; reloadable up to 7 reloads (8 firings); confirm codes with the manufacturer (the green version PS9LSR90G does appear in public sources) |
| Lantex / Infiniti Medical | Linear Stapler | LS90B (instrument) / LSR90B (reload) | Comparable by category | 90 mm, blue, two rows of titanium staples; proprietary structure and own number of staples; verify system compatibility |
Note on codes: TA9035S and TA9035L appear consistently in Medtronic sources and authorised distributors, and LS9035 appears in the official catalogue of B. Braun (Aesculap); they are cited as control references. The internal code WZLS-A-90L is Cahnos’ own reference (manufacturer WZ); it has not been found in independent public sources, so it should be treated as a distributor/factory code and confirmed with the manufacturer before use in tender documents.
5. Selection tips and variants of interest
- Line length: choose 30 mm (WZLS-A-30), 45 mm (WZLS-A-45), 60 mm (WZLS-A-60) or 90 mm (WZLS-A-90) according to the extent of the tissue to be sutured and the depth of the surgical field.
- Staple height / colour: the blue version (3.8 mm, this model) is suitable for normal or thin-wall tissue; the green version (4.5 mm) is intended for thick, oedematous tissue or with greater compression requirements, such as pulmonary parenchyma or thickened intestine.
- Single-use: verify the integrity of the sterile packaging before opening and do not reuse the device; dispose of it according to the centre’s protocol after each intervention.
- Range variants: if your department requires stapling with simultaneous cutting, consult the family of reloadable linear cutting staplers WZCTS-A (60/80/100 mm) and their spare parts; for laparoscopic approaches, the range of endoscopic linear staplers and articulated spare loads.
- For tenders: since there is no official cross-reference table from the manufacturer, validate each equivalence with the technical department and provide the technical data sheets of the comparables before submitting a bid.










Reviews