SafeAir MR105D. Ambient Oxygen Monitor for Magnetic Resonance Imaging: Detection and Alarm of Reduced O₂ Levels in the Examination Room

Price on request

The SafeAir ambient oxygen monitor for magnetic resonance imaging (original reference MR105D) is the essential safety device in any MRI examination room: it quickly and continuously detects any decrease in oxygen content (normal reference 21%) caused by a possible cryogenic helium leak or a failure of the ventilation system. It is installed in the control room without needing to switch off or demagnetise the scanner, requires no radiofrequency filtering and can be connected to the room’s extraction system for its automatic activation in the event of reduced oxygen levels. It includes a touchscreen with configurable high/low alarm limits, audible and visual alarm, 30 m sampling hose, replaceable HEPA filter and power supply with internal backup battery.

International shipping. Import duties and taxes are payable by the recipient.

Reference
CAHSAFAIROX

You will also find it in

Description

1. Oxygen safety in your MRI room

In a Magnetic Resonance (MR) service, the protection of staff and patients is an absolute priority. One of the most silent risks is the possible leak of cryogenic helium or other gases that displace ambient oxygen and reduce its concentration below the breathable level. The SafeAir monitor (MR105D) is designed precisely to provide a reliable and early warning of reduced oxygen levels in the scanning room: it is installed in the control room, is easy to mount during scanner installation or afterwards, and does not require switching off the MR equipment. A simple, robust and highly valuable solution for radiologists, radiographers, hospital engineers and facility safety officers.

2. How the SafeAir oxygen monitor works

The system performs continuous sampling of the oxygen content from one or more locations in the scanning room: a suction pump drives a constant air flow through a 6 mm flexible PVC hose to an electrochemical oxygen fuel cell detector (Envitec system), which measures the concentration at a predetermined flow rate. The normal oxygen concentration should be 21%; any variation is detected quickly. The monitor incorporates a touchscreen display showing the oxygen level and confirming the upper and lower limits, alarm configuration keys, calibration option (ambient air at 21% or O₂ source at 100%), battery and alarm test, and automatic low battery warning. In addition, it has a pair of internal contacts accessible via a four-pin connector on the rear that allow automatic activation of the ventilation or emergency extraction system of the room when the low oxygen alarm sounds. In the event of mains power loss, an internal backup battery keeps the alarms and ventilation relay operational.

3. Technical Specifications

The main specifications of the ambient oxygen monitor SafeAir MR105D (manufacturer Wardray Premise, Envitec measurement system) are detailed below:

General characteristics

Parameter Specification
Original reference MR105D (SafeAir® Room Oxygen Monitor)
Manufacturer Wardray Premise (Envitec O₂ sensor and electronics)
Set type Ambient oxygen monitor for MRI rooms
Main function Warning of reduced oxygen levels in the scanning room
Installation location Control room (the equipment must not be taken into the MRI room; classified MR Unsafe)
Measurement Continuous sampling of oxygen content from one or more locations
Reference concentration 21% oxygen in ambient air (must not be below 18% for breathable air)
Alarms Visual and acoustic, with configurable high/low limits, alarm silence and low battery warning
Interface Touchscreen with on/off, oxygen % display, calibration and limit adjustment
Ventilation output Switched contact pair (rear 4-pin connector) to activate emergency extraction
RF requirement No radiofrequency filtering required

Dimensions, weight and power supply

Parameter Value
Height 360 mm
Width 120 mm (max.)
Depth 230 mm (incl. cable)
Weight 4.5 kg
Power supply Mains adapter 100–240 V AC, 50/60 Hz, 1.4 A max. (output 18 V DC, 3.62 A)
Backup battery Internal dryfit battery for alarms and ventilation relay in case of mains loss
Ventilation contacts 230 V AC 3 A max. / 115 V AC 6 A max.

Supply contents and maintenance

Item Description
SafeAir unit Complete MR105D oxygen monitor
Transformer 230 V AC – 18 V DC (ref. MR40-LR), with European and UK plugs
Sampling hose Flexible PVC tube, 6 mm, 30 metres long
Ventilation connector 4-pin Vent Relay / Emergency Extract Connector
Filter and sensor User-replaceable HEPA filter and oxygen fuel cell
Recommended maintenance Replacement of HEPA filter and cell every 6 months (replacement kit ref. MR106)
Documentation Instruction manual (IFU) and Envitec PC software

4. Equivalences and Cross-References

The SafeAir monitor is the MR105D model from British manufacturer Wardray Premise, specialised in shielding and safety accessories for magnetic resonance imaging. The detection system relies on the electronics and oxygen cell from Envitec, a well-established reference in gas monitoring. As official consumables, the manufacturer markets the replacement kit MR106 (oxygen fuel cell + HEPA line filter), recommended every 6 months. Functionally, this device falls within the family of oxygen deficiency detectors for MRI rooms available on the market, such as oxygen detectors for MRI environments from FM Grupo Tecnológico or 1- to 3-channel O₂ monitors for MRI installations supplied together with environmental safety control units. In the Cahnos catalogue, this monitor is complemented by the rest of the MRI safety range (test magnets, metal detectors and room accessories) and differs from the MRI pulse oximeter (patient monitoring) in that its function is to monitor the room atmosphere, not the patient.

5. Selection tips and variants of interest

  • Location: install the unit in the control room, in a visible and audible position, and route the sampling hose through a suitable waveguide to the vicinity of the scanner’s cold head.
  • Calibration: after installation or sensor replacement, calibrate the device with ambient air (21%) or with a pure O₂ source (100%); the touchscreen allows precise adjustment of high and low alarm limits.
  • Safety threshold: remember that the oxygen concentration of breathable air must not be lower than 18%; set the alarm limits according to your centre’s protocol.
  • Integration with extraction: connect the rear 4-pin connector to the room’s emergency ventilation/extraction system to achieve automatic activation in the event of reduced oxygen levels.
  • Maintenance: schedule the six-monthly replacement of the HEPA filter and fuel cell (kit MR106); the airflow failure warning facilitates diagnosis of blockages or saturated filters.
  • Variants of interest: to reinforce the comprehensive safety of your MRI service, combine this monitor with the rest of the Cahnos safety range (test magnet, portable metal detector and room accessories).

Additional information

Manufacturer

Wardray Premise (Envitec measurement system)

Reference model

MR105D (SafeAir Room Oxygen Monitor)

Product type

Ambient oxygen monitor for MRI room

Installation

Control room; do not introduce into the scanning room (MR Unsafe); no RF filtering requirement

Measurement

Continuous sampling of oxygen content from one or more room locations

Oxygen concentration

Normal reference 21%; not below 18% for breathable air

Alarms

Visual and acoustic, configurable high/low limits, alarm silence and low battery warning

Ventilation output

Switched contact pair 230 V AC 3 A max. / 115 V AC 6 A max., rear 4-pin connector

Sampling hose

Flexible PVC tubing 6 mm, 30 metres

Filter

User-replaceable HEPA filter; replacement kit MR106 (with fuel cell) every 6 months

Power supply

Mains adapter 100–240 V AC 50/60 Hz, output 18 V DC 3.62 A; internal backup battery

Dimensions

360 mm (height) x 120 mm (width) x 230 mm (depth, incl. cable)

Weight

4.5 kg

Regulatory standard

EMC Directive 2014/30/EU, Low Voltage Directive 2014/35/EU, RoHS 2011/65/EU

Presentation

Unit with transformer, European and UK plugs, 30 m hose, extraction connector and documentation