Description
1. Complete and sterile kit for the percutaneous airway with subglottic suction
The percutaneous dilatation kit Ref. TP02-75 is the integral single-use solution for establishing a tracheal access by puncture and dilatation following the Griggs technique, without the need for forceps. It incorporates an 8 mm tracheostomy tube with fixed cervical flange and dedicated subglottic suction lumen, a feature especially valued in intensive care units for its contribution to the prevention of ventilator-associated pneumonia (VAP).
All the necessary material is presented in a single sterile kit, ready to use: a single order, a single code and the guarantee that the team has all the instruments with complete traceability. Preparation time is reduced, the risk of forgetting material is minimised and standardisation of the procedure is facilitated in ICU, emergency departments and operating theatre.
2. Technical explanation of the equipment and procedure
Percutaneous dilatational tracheostomy (PDT) according to the Griggs technique is an evolution of the Ciaglia technique: it is performed at the bedside, usually under bronchoscopic control, without the need to transfer the critically ill patient to the operating theatre. The procedure begins with tracheal puncture using the 16 G introducer needle, continues with the passage of the J-tip guide wire (atraumatic tip that minimises tissue damage) and ends with progressive dilatation of the stoma using the single 14 Fr dilator and advancement of the pre-mounted cannula over the introducer with locking system, which maintains the position of the tube during insertion until its correct placement in the trachea.
The cannula incorporates an additional subglottic suction lumen with outlet above the balloon (cuff), which allows aspiration of secretions accumulated in the subglottic space and reduces the risk of contaminated micro-aspirations into the lower airway. The fixed cervical flange, flexible and in transparent plastic, facilitates stable fixation and visualisation of the stoma; the balloon is low-pressure and the connector is universal 15 mm (15M). All components are single-use, latex-free and phthalate-free.
3. Technical specifications and kit contents
Kit contents (Ref. TP02-75)
| Component | Specification |
|---|---|
| Tracheostomy tube | 8 mm tube with fixed flange and subglottic suction lumen |
| Tracheostomy tube introducer | With locking system and conical tip for safe insertion |
| Dilator | 14 Fr for progressive dilatation of the stoma |
| Guidewire | J-tip 0.05\”, NiTiNol core resistant to kinking |
| Guide catheter | For guiding the dilator and the tube |
| Scalpel | Safety scalpel No. 15 with retractable protection and fixed blade |
| Syringe | 10 ml |
| Introducer needle | 16 G |
| Fixation band | Adjustable, with Velcro-type closure |
| Instrument table cover | Sterile field cover |
| Tracheal compress | 9 x 9 cm, three-layer foam |
| Gauze swabs | 10 x 10 cm |
| Sterile lubricating gel | Individual pouch (on request) |
| Dilatation forceps | Not included (kit without forceps) |
Technical features of the tube
| Parameter | Value |
|---|---|
| Internal tube diameter | 8.0 mm |
| Reference outer diameter | 11.9 mm |
| Flange type | Fixed, cervical, flexible, transparent plastic |
| Subglottic suction | Yes, dedicated lumen |
| Balloon (cuff) | Low pressure |
| Connector | Universal 15 mm (15M) |
| Radiological control | Radiopaque line |
| Use | Single-use (sterile) |
| Sterilisation | Ethylene oxide (ETO), process validated according to EN 11135-1; desorption in accordance with ISO 10993-7 |
| Materials | Latex-free and phthalate-free |
| Classification | CE medical device, class IIa |
| Indication | Percutaneous tracheostomy, cricothyrotomy, coniotomy or laryngotomy |
Family of Griggs kits with fixed flange and subglottic suction (TP02-7x series)
| Reference | Diameter | Flange | Subglottic suction | Forceps |
|---|---|---|---|---|
| TP02-74 | 7 mm | Fixed | Yes | No |
| TP02-75 (this product) | 8 mm | Fixed | Yes | No |
| TP02-76 | 9 mm | Fixed | Yes | No |
| TP02-64 | 7 mm | Fixed | Yes | Yes |
| TP02-65 | 8 mm | Fixed | Yes | Yes |
| TP02-66 | 9 mm | Fixed | Yes | Yes |
4. Equivalences and Cross-References
The SKU TP02-75 corresponds to a proprietary reference of Cahnos within the family of percutaneous kits for tracheotomy, according to the manufacturer’s technical data sheet (tracheostomy kits with fixed flange and suction line). For range purposes, it shares the exact configuration (fixed flange with suction, without forceps) with references TP02-74 (7 mm) and TP02-76 (9 mm), which differ only in the diameter of the cannula and, therefore, are not interchangeable 1:1 depending on the patient: they are direct range equivalents, not identical.
At category level, the set is comparable to percutaneous dilatational tracheostomy systems with subglottic suction from other manufacturers:
- Comparable at category level: Portex® (GRIGGS®) Percutaneous Dilation Tracheostomy Kit with Blue Line Ultra® Suctionaid® tube (ICU Medical / Smiths Medical). Reference 100/893/070 (7.0 mm, with suction lumen, without forceps) is confirmed in authorised distributor catalogue; for the 8.0 mm calibre, the equivalent code remains pending confirmation with the manufacturer. Same functional family (Griggs technique, tube with subglottic suction, without forceps), but different manufacturer and component geometry: not interchangeable 1:1.
- Comparable at category level: TRACOE experc kit (REF 520) completed with a cannula with subglottic suction device. Same clinical indication and suction function, different component configuration: not interchangeable 1:1.
- Comparable at category level: Cook Medical percutaneous kits (Ciaglia and Melker families) and Portex single-dilator kits. The basic technique may differ (single dilator type Ciaglia versus Griggs with forceps) and the cannula configuration also: they should not be considered exact equivalents.
There is no equivalence table validated by the manufacturer that allows cross-referencing this reference with specific codes from other suppliers; consult the technical department of Cahnos before using this data sheet in tenders or stock replacement protocols.
5. Selection tips and variants of interest
The cannula diameter should be selected according to the patient’s anatomy and ventilation requirements: the 7 mm variant (TP02-74) is suitable for smaller-built adults, the 8 mm variant (this kit, TP02-75) is the reference option for medium-built adults, and the 9 mm variant (TP02-76) is reserved for larger builds or patients with greater ventilatory flow demand.
If the patient will require prolonged mechanical ventilation, the subglottic suction of this series is especially recommended for its role in preventing ventilator-associated pneumonia. If the department prefers to complete the dilation with forceps under endoscopic control, choose the versions with dilation forceps (TP02-64, TP02-65 and TP02-66). If an adjustable flange is preferred, consult the Griggs series with adjustable flange available in the store. The kits in the family are single-use: check the integrity of the sterile packaging and the expiry date before use and dispose of them after the procedure as clinical waste in accordance with current local regulations.