Description
1. Safe ventilation in paediatric anaesthesia and intensive care
The double-lumen laryngeal mask in size 2.0 Paediatric M (ref. LM1120015) is the supraglottic device indicated for securing the airway quickly, atraumatically and reliably in paediatric patients weighing 10 to 20 kg. It combines a secure and comfortable seal with a gastric drainage channel, making it the reference solution for paediatric anaesthesia services, intensive care units and emergency teams that need to guarantee ventilation, even when regurgitation risk is anticipated.
Its translucent, soft and anatomical structure, made of PVC, polycarbonate (PC) and polyester, facilitates visual monitoring of the airway throughout the procedure. It is supplied in boxes of 5 units, designed for continuous supply to the clinical service.
2. What it is and how it works
Unlike endotracheal intubation, the laryngeal mask is an extraglottic device that rests on the hypopharynx and seals around the laryngeal inlet with an inflatable cuff, without passing through the glottis. This makes it faster to place, better tolerated at lighter anaesthetic planes and especially useful in difficult airways, without the need for laryngoscopy.
In the double-lumen version, alongside the ventilation tube (standard 15M connector) runs a drainage tube through which a gastric tube can be passed or oesophageal contents evacuated, protecting the airway against aspiration of gastric contents. The paediatric version incorporates a pre-inserted stylet made of malleable aluminium coated with medical plastic that facilitates insertion in the paediatric anatomy. The pilot balloon has a self-sealing valve for controlled inflation and pressure verification, and the ventilation tube is marked with the size, recommended patient weight and cuff inflation volume.
3. Technical Specifications
| Type | Inflatable double-lumen supraglottic laryngeal mask |
| Material | PVC, polycarbonate (PC) and polyester; latex-free and DEHP-free |
| Configuration | Double lumen: ventilation tube + gastric drainage tube |
| Size | 2.0 Paediatric M (recommended patient: children 10 to 20 kg) |
| Maximum inflation volume | 10 ml of air |
| Internal diameter (± 0.5 mm) | 7.0 mm |
| External diameter (± 0.5 mm) | 11.3 mm |
| Radius of curvature (± 30 mm) | 120 mm |
| Connector | Standard 15 mm (15M) |
| Stylet | Pre-inserted, made of malleable aluminium coated with medical plastic |
| Sleeve | Elongated high-volume cuff, with oesophageal seal and improved anatomical fit |
| Use | Single-use |
| Sterility | Sterile, sterilised by ethylene oxide (EO), in accordance with ISO 11135 |
| Presentation | Box of 5 units |
| Reference | LM1120015 |
Double-lumen series table (manufacturer’s data):
| Size | Recommended patient | Max. inflation vol. | Internal Ø (± 0.5 mm) | External Ø (± 0.5 mm) | Radius of curvature (± 30 mm) |
|---|---|---|---|---|---|
| 1 | Neonate < 5 kg | 4 ml | 5.5 mm | 8.3 mm | 100 mm |
| 1½ | Children 5–10 kg | 7 ml | 6.0 mm | 10.0 mm | 120 mm |
| 2 | Children 10–20 kg (this product) | 10 ml | 7.0 mm | 11.3 mm | 120 mm |
| 2½ | Children 20–30 kg | 14 ml | 9.0 mm | 13.3 mm | 140 mm |
| 3 | Children/adults 30–50 kg | 20 ml | 10.0 mm | 15.3 mm | 165 mm |
| 4 | Adults 50–70 kg | 30 ml | 10.0 mm | 15.3 mm | 165 mm |
| 5 | Adults 70–100 kg | 40 ml | 11.5 mm | 16.7 mm | 185 mm |
Technical note: the dimensional parameters correspond to the double-lumen series according to the manufacturer’s technical documentation. The indicated volumes and weights are also engraved on the tube of each unit.
4. Equivalences and Cross-References
Reference LM1120015 corresponds to the manufacturer’s official designation for the paediatric double-lumen laryngeal mask size 2.0 (sterile, box of 5), verified in the manufacturer’s technical catalogue (double-lumen series: LM1115015, LM1120015, LM1125015, LM1130015 and LM1140015).
Comparable at category level, not interchangeable 1:1: there is no equivalence table validated by the manufacturer against double-lumen devices from other brands (e.g. supraglottic masks with drainage from Teleflex LMA, Ambu or Intersurgical); it is recommended to consult the technical department at Cahnos before using this datasheet in tendering processes or stock replacement.
Within the Cahnos catalogue, the same paediatric double-lumen series includes references LM1115015 (size 1.5, 5–10 kg) and LM1125015 (size 2.5, 20–30 kg), with pre-inserted stylet, and adult sizes LM1130015 (3.0) and LM1140015 (4.0). They differ in size and dimensional parameters: they should not be treated as a direct substitute for this size 2.0.
The standard series of laryngeal masks without drainage (ref. LM0020101 for size 2.0, box of 10) meets the same general clinical indication for a supraglottic airway, but without a gastric drainage channel: comparable by category, not a 1:1 functional equivalent. As a market reference, disposable laryngeal masks in size 2 (10–20 kg) from other suppliers respond to the same general clinical indication, but differ in configuration, material and manufacturer, so they should not be considered interchangeable without technical validation.
The devices in this series comply with current EU medical device regulations (MDR 2017/745); always consult the original labelling for the specific product classification.
5. Selection tips and variants of interest
The size should be chosen according to the patient’s weight and not only by age or commercial designation:
| Size | Recommended patient |
|---|---|
| 1.0 | Neonates up to 5 kg |
| 1.5 | Children 5–10 kg |
| 2.0 (this product) | Children 10–20 kg |
| 2.5 | Children 20–30 kg |
| 3.0 | Children/adults 30–50 kg |
| 4.0 | Adults 50–70 kg |
| 5.0 | Adults 70–100 kg |
This size 2.0 with drainage channel is especially recommended in paediatric patients weighing 10 to 20 kg when there is a risk of regurgitation or aspiration. If the procedure does not require gastric access, the standard range of laryngeal masks (refs. LM0010101 to LM0050101, box of 10) offers a lower cost per unit alternative. During use, it is recommended to inflate the cuff to a pressure of 18 to 20 cmH2O, check the integrity of the pilot balloon before insertion and confirm the seal by monitoring the expired volume. Always check the integrity of the sterile packaging and the expiry date before use; the device is single-use and must be disposed of after the procedure as clinical waste in accordance with current local regulations.