Description
1. Introduction
The preformed nasal tracheal tube with cuff and Murphy eye ID 5.0 mm (reference ETPN50C02) is the reference solution for nasal airway management in anaesthesia operating theatres, intensive care units, emergency and resuscitation departments. Its preformed geometry reproduces the nasopharyngeal anatomical curvature, facilitating insertion, reducing tissue trauma and keeping the surgical field clear in head and neck procedures, oral or dentomaxillary surgery. Manufactured by DEAS S.r.l. (Deaflux Respiratory Products, Italy) and distributed in Spain by Cahnos, it is supplied in a box of 10 individually packaged, sterile units, ready for immediate use.
2. Technical explanation of the device and procedure
In nasotracheal intubation, the tube is inserted through a nostril, advances through the nasopharynx, passes through the glottis and sits in the trachea, where it conducts gases from the ventilator or anaesthesia circuit to the lungs. The preformed south-facing curvature (curved towards the chin) follows the anatomy of the upper airway: it facilitates tube fixation, minimises the risk of kinking and prevents the circuit from interfering with the surgical field.
The low-pressure, high-volume cuff, inflated through the pilot balloon with self-sealing valve, seals the trachea and allows positive-pressure ventilation without significant leaks, with lower pressure on the mucosa than high-pressure cuffs. The Murphy eye, a distal lateral orifice smoothly formed and without sharp edges, ensures an alternative gas passage if the tube tip were obstructed by secretions or by resting against the tracheal wall. The longitudinal radiopaque line, the double depth marker and the graduation every centimetre allow correct placement to be verified radiographically and clinically. The atraumatic, rounded tip minimises damage during intubation and extubation.
3. Technical Specifications
Technical data verified against the official technical data sheet of the ETPN range from manufacturer DEAS S.r.l. (Deaflux), document ref. 04906-3e.doc Rev.3.
| Parameter | Specification |
|---|---|
| Reference | ETPN50C02 |
| Type | Preformed nasal tracheal tube with cuff and Murphy eye |
| Access route | Nasal (nasopharyngeal), preformed south-facing design |
| Internal diameter (ID) | 5.0 mm |
| External diameter (OD) | 6.7 mm |
| Total length | 245 mm |
| Inflated cuff diameter | 17 mm |
| Sleeve | Sealing, low pressure and high volume, with pilot balloon and self-sealing valve |
| Murphy eye | Yes, smoothly formed, distal lateral safety orifice |
| Material | Transparent, thermosensitive medical-grade polyvinyl chloride (PVC) |
| Connector | Universal 15 mm conforming to ISO 5356-1 |
| Marking | Longitudinal radiopaque line, double depth marker, graduation every cm and cutting indicator |
| Latex | Natural rubber latex-free |
| Phthalates | DEHP/DOP-free |
| Sterility | Sterile, individually packaged (ethylene oxide per EN 11135-1; residues per ISO 10993-7) |
| Use | Single-use, non-reusable |
| CE classification | Medical device class IIa (Directive 93/42/EEC, notified body TÜV SÜD) |
| Presentation | Box of 10 individually packaged units (carton of 100 units) |
| Expiry | 59 months from the date of production (indicated on the label) |
ETPN range – preformed nasal tracheal tubes with cuff (same design, presentation and manufacturer, Cahnos references):
| Reference | ID (mm) | OD (mm) | Length (mm) | Cuff Ø (mm) |
|---|---|---|---|---|
| ETPN40C02 | 4.0 | 5.3 | 210 | 14 |
| ETPN45C02 | 4.5 | 6.0 | 225 | 14 |
| ETPN50C02 (this product) | 5.0 | 6.7 | 245 | 17 |
| ETPN55C02 | 5.5 | 7.3 | 275 | 17 |
| ETPN60C02 | 6.0 | 8.0 | 285 | 20 |
| ETPN65C02 | 6.5 | 8.7 | 295 | 20 |
| ETPN70C02 | 7.0 | 9.3 | 310 | 25 |
| ETPN75C02 | 7.5 | 10.0 | 315 | 25 |
| ETPN80C02 | 8.0 | 10.7 | 330 | 26 |
Note: the range of diameters, lengths and cuff diameters shown in the table comes from the manufacturer’s official technical data sheet (DEAS S.r.l. / Deaflux) and from the Cahnos ETPN range catalogue; do not extend these values to other sizes without technical confirmation.
4. Equivalences and Cross-References
The reference ETPN50C02 is an own code of the Cahnos range (manufacturer DEAS S.r.l. / Deaflux Respiratory Products, Italy) and must not be confused with the official part number of any third-party manufacturer. There is no exact reference equivalence table validated by the manufacturer for this code; consult the Cahnos technical department before using this reference in specifications or tenders.
Comparable at category level (not 1:1 interchangeable): preformed nasal tubes with cuff and Murphy eye from lines such as Rüsch® AGT/RAE nasal and Sheridan® (Teleflex), Shiley™ RAE nasal with TaperGuard™ cuff (Medtronic) or BOENMED nasal preformed tubes meet the same general clinical indication and RAE nasal-type geometry. Any specific reference must be individually checked against the manufacturer’s official technical data sheet, since cuff technology, curve geometry, balloon material and connection system may differ; the ID 5.0 mm size of each brand must be verified before proposing stock substitution.
Comparable by category within the Cahnos catalogue (same ID 5.0 mm, other configurations):
| Reference | Description |
|---|---|
| ETAN50C02 | Oral endotracheal tube (Magill type) with cuff and Murphy eye |
| ETPO50C02 | Preformed oral endotracheal tube with cuff and Murphy eye |
Verification of this reference has been carried out against the manufacturer’s official technical data sheet (DEAS S.r.l. / Deaflux, ref. 04906-3e.doc Rev.3) and the Cahnos catalogue; there is no equivalence table validated by the manufacturer; consult the technical department before using in a tender.
5. Selection tips and variants of interest
- Patient selection: the ID 5.0 mm cuffed size is usually used in older paediatric patients, adolescents or adults with a narrow airway; the final choice must always be based on the department protocol and anaesthetic assessment (age, weight and estimated airway and nostril size).
- Cuff pressure: inflate through the valve with a syringe and maintain a sealing pressure of 20–30 cmH₂O (usual clinical practice range, not a manufacturer specification) to minimise tracheal mucosal damage; monitor with a pressure gauge if available in the department.
- Position verification: check by bilateral auscultation and, if possible, radiography using the radiopaque line and the tube’s double depth marker.
- Preferred indications: oral, maxillofacial, ENT and ophthalmic surgery, where it is important to keep the circuit out of the operative field, as well as mechanical ventilation in intensive care.
- Variants of interest in the catalogue: if you require another size, the ETPN series covers from 4.0 mm to 8.0 mm (ETPN40C02 to ETPN80C02). For oral intubation, consider the ETAN (Magill type) and ETPO (oral preformed) ranges. Contact the technical department at Cahnos for samples, volume pricing and hospital supply lead times.





Reviews