Description
1. Introduction
The oral endotracheal tube with cuff, suction lumen and Murphy eye (ref. ETSL60C02) is an advanced airway management device designed for orotracheal intubation in tall paediatric patients or small-built adults. It is the reference solution in intensive care units, anaesthesia operating theatres and emergency departments where a sealed, stable airway with continuous suction capability is required during mechanical ventilation or general anaesthesia. It is supplied in a box of 10 units, individually packaged and sterile, ready for immediate use.
2. Technical explanation of the device
The tube is inserted orally through the glottis into the trachea and is connected at its proximal end to a ventilator, an anaesthesia circuit or a resuscitation bag via the standard 15 mm connector (conforming to ISO 5356-1). Its low-pressure, high-volume cuff, once inflated via the pilot balloon, creates an airtight seal against the tracheal wall that allows positive pressure ventilation and minimises the risk of leakage and aspiration of gastric or pharyngeal contents, with lower pressure on the mucosa than high-pressure cuffs.
The integrated suction lumen runs along the wall of the tube and opens into a proximal port connectable to a vacuum system, allowing subglottic secretions accumulated above the cuff to be aspirated without the need to extubate the patient. This function reduces bacterial colonisation of the lower respiratory tract and is a recognised measure for preventing ventilator-associated pneumonia (VAP). The Murphy eye, a lateral opening at the distal end, ensures airflow even if the tip of the tube becomes occluded by contact with the tracheal wall. The radiopaque line and double depth marker facilitate position verification by radiography and clinical monitoring.
3. Technical Specifications
Technical data verified against the official technical data sheet of the ETSL range from Cahnos:
| Parameter | Specification |
|---|---|
| Access route | Oral (orotracheal) |
| Internal diameter (ID) | 6.0 mm |
| External diameter (OD) | 9.6 mm |
| Total length | 285 mm |
| Inflated cuff diameter | 20 mm |
| Sleeve | Sealing, low pressure and high volume (pilot balloon with size indication) |
| Suction lumen | Yes, integrated for continuous or intermittent subglottic suction |
| Murphy eye | Yes, distal lateral safety opening |
| Material | Transparent, thermosensitive medical-grade polyvinyl chloride (PVC) |
| Connector | Universal 15 mm (ISO 5356-1) |
| Marking | Longitudinal radiopaque line, double depth marker and graduation every cm |
| Latex | Natural rubber latex-free |
| Phthalates | DEHP/DOP-free |
| Sterility | Sterile, individually packaged (ethylene oxide, according to EN ISO 11135-1) |
| Use | Single-use, non-reusable |
| CE classification | Medical device class IIa (Regulation (EU) 2017/745, MDR) |
| Presentation | Box of 10 individually packaged units |
ETSL range – oral endotracheal tubes with cuff and suction lumen (Cahnos references):
| Reference | ID (mm) | OD (mm) | Length (mm) | Cuff Ø (mm) |
|---|---|---|---|---|
| ETSL60C02 (this product) | 6.0 | 9.6 | 285 | 20 |
| ETSL65C02 | 6.5 | 10.2 | 295 | 20 |
| ETSL70C02 | 7.0 | 10.8 | 310 | 25 |
| ETSL75C02 | 7.5 | 11.5 | 315 | 25 |
| ETSL80C02 | 8.0 | 12.2 | 330 | 26 |
| ETSL85C02 | 8.5 | 12.9 | 330 | 26 |
| ETSL90C02 | 9.0 | 13.6 | 330 | 28 |
4. Equivalences and Cross-References
Functional equivalence of category (not interchangeable 1:1). This tube belongs to the family of endotracheal tubes with subglottic suction lumen, a category represented by manufacturers such as Teleflex (Rüsch® Guardian™ with evacuation lumen), Medtronic (Shiley™ Evac with TaperGuard™ technology) and ICU Medical (Portex line with subglottic suction). All share the same general clinical indication —drainage of secretions above the cuff to reduce VAP—, but they differ in cuff material (ultra-thin polyurethane vs PVC), balloon geometry and lumen connection system, so they should not be treated as direct equivalents without individual verification of each reference.
The reference ETSL60C02 is a proprietary code of the Cahnos range and could not be cross-referenced with an official third-party manufacturer code in public sources: there is no equivalence table validated by the manufacturer; consult the technical department of Cahnos before use in tenders or as a replacement for stock from other brands.
Comparables by category within the Cahnos catalogue (same ID 6.0 mm):
| Reference | Description |
|---|---|
| ETAN60C02 | Oral endotracheal tube with cuff and Murphy eye, without suction lumen |
| ETPN60C02 | Preformed nasal tracheal tube with cuff and Murphy eye |
| ETA060C02A2 | Reinforced oral endotracheal tube with cuff |
| ETAM60C02A1 | Reinforced oral endotracheal tube with cuff and guide |
5. Selection tips and variants of interest
The ID 6.0 mm is reserved for tall paediatric patients or small-built adults; as a general guide, in adults tubes of 7.0–9.0 mm are used depending on build, always selecting the largest size compatible with the patient’s anatomy. Before intubation, check the integrity of the cuff, the patency of the suction lumen and the pilot balloon, and control the sealing pressure (usual recommended cuff pressure: 20–30 cmH₂O; common market practice, verify the centre’s protocol).
If your centre works with prolonged intubations or patients at high risk of VAP, the ETSL range with suction lumen is the indicated option. For procedures requiring kink resistance or frequent repositioning, consider the reinforced tubes of the ETA series (ETA060C02A2, ETAM60C02A1). For head and neck surgery or nasal route, consider the nasal preformed tube (ETPN60C02). Contact the technical department of Cahnos for samples, volume conditions and hospital supply lead times.