Description
1. Introduction
The oral endotracheal tube with cuff, subglottic suction lumen and Murphy eye ID 8.0 mm (ref. ETSL80C02) is an airway management device designed for orotracheal intubation of adult patients. It is the reference solution in intensive care units, anaesthesia services and emergency departments where a sealed, stable airway with continuous suction capability is required during mechanical ventilation or general anaesthesia. It is supplied in a box of 10 units, individually packaged and sterile, ready for immediate use.
2. Technical explanation of the device
The tube is inserted orally through the glottis into the trachea and is connected at its proximal end to a ventilator, an anaesthesia circuit or a resuscitation bag via the standard 15 mm connector (conforming to ISO 5356-1). The low-pressure, high-volume cuff, once inflated via the pilot balloon with Luer connection, creates an airtight seal against the tracheal wall that allows positive pressure ventilation and minimises the risk of leakage and aspiration of gastric or pharyngeal contents.
Its integrated suction lumen runs along the wall of the tube and opens into a proximal port that can be connected to a vacuum system, allowing subglottic secretions accumulated above the cuff to be aspirated without the need to extubate the patient. This function reduces bacterial colonisation of the lower respiratory tract and is a recognised measure for preventing ventilator-associated pneumonia (VAP). The Murphy eye, a lateral opening at the distal end, ensures airflow even if the tip of the tube becomes occluded by contact with the tracheal wall. The Magill-type curve, the rounded atraumatic tip, the radiopaque line and the depth marks facilitate safe placement and verification.
3. Technical Specifications
| Parameter | Specification |
|---|---|
| Reference | ETSL80C02 |
| Material | Thermosensitive polyvinyl chloride (PVC), latex-free and DEHP-free |
| Type | Oral, with cuff and Magill curve conforming to ISO/EN |
| Internal diameter (ID) | 8.0 mm |
| Outer diameter (OD) | 12.2 mm |
| Total length | 330 mm |
| Sleeve | Low pressure and high volume (balloon diameter: 26 mm) |
| Suction lumen | Yes, drainage/suction of the subglottic area |
| Murphy eye | Yes, gently bevelled |
| Connector | Universal 15 mm (ISO 5356-1), Luer connection for pilot balloon |
| Marking | Centimetric graduation, double depth marker, longitudinal radiopaque line |
| Use | Single-use (disposable), not reusable |
| Sterilisation | Ethylene oxide (ETO), individual blister packaging |
| Presentation | Box of 10 units |
| Indication | Orotracheal intubation, positive pressure ventilation, anaesthesia |
ETSL range – oral endotracheal tubes with cuff and suction lumen (references Cahnos):
| Reference | ID (mm) | OD (mm) | Length (mm) | Cuff Ø (mm) |
|---|---|---|---|---|
| ETSL60C02 | 6.0 | 9.6 | 285 | 20 |
| ETSL65C02 | 6.5 | 10.2 | 295 | 20 |
| ETSL70C02 | 7.0 | 10.8 | 310 | 25 |
| ETSL75C02 | 7.5 | 11.5 | 315 | 25 |
| ETSL80C02 (this product) | 8.0 | 12.2 | 330 | 26 |
| ETSL85C02 | 8.5 | 12.9 | 330 | 26 |
| ETSL90C02 | 9.0 | 13.6 | 330 | 28 |
4. Equivalences and Cross-References
Functional category equivalence (not interchangeable 1:1). This tube belongs to the family of endotracheal tubes with subglottic suction lumen, a category represented by manufacturers such as Teleflex (Rüsch® Guardian™ with evacuation lumen), Medtronic (Shiley™ Evac with TaperGuard™) and ICU Medical (Portex SACETT™ / Blue Line™). All share the same general clinical indication —drainage of secretions above the cuff to reduce VAP—, but differ in cuff material (ultra-thin polyurethane vs. PVC), geometry and connection system, so they should not be treated as direct equivalents without individual verification.
The reference ETSL80C02 corresponds to the nomenclature of the ETSL range of the DEAFLUX line from DEAS S.r.l., a pattern verified in the manufacturer’s official catalogue for this family of tracheal tubes with suction line; however, this specific calibre could not be independently verified in public sources, so it remains pending confirmation with the manufacturer/distributor for use in tenders or stock replacement.
There is no reference equivalence table validated by the manufacturer for this code; consult the technical department of Cahnos before using in tenders. Comparables by category within the Cahnos catalogue in ID 8.0 mm: ETAN80C02 (with cuff and Murphy eye, without suction lumen), ETA080C02A2 (reinforced with cuff) and ETAM80C02A1 (reinforced with cuff and guide).
5. Selection tips and variants of interest
ID 8.0 mm is a standard size commonly used in adult patients; as a general reference, tubes of 7.0–9.0 mm are used depending on body build, always selecting the largest size compatible with the patient’s anatomy. Before intubation, check the integrity of the cuff, the patency of the suction lumen and the pilot balloon, and control the sealing pressure (usual recommended cuff pressure: 20–30 cmH₂O; common market practice, verify the centre’s protocol).
If your centre works with prolonged intubations or patients at high risk of VAP, the ETSL range with suction lumen is the indicated option. For procedures requiring resistance to kinking or frequent repositioning, consider the reinforced tubes of the ETA series (ETA080C02A2, ETAM80C02A1). For head and neck surgery or nasal route, consider preformed tubes. Contact the technical department of Cahnos for samples, volume conditions and hospital supply lead times.