Description
1. Introduction
The preformed nasal tracheal tube with cuff and Murphy eye, ID 6.5 mm (Ref. ETPN65C02) is the reference solution for nasal airway management in anaesthetic operating theatres, intensive care units, emergency departments and resuscitation. Its preformed geometry reproduces the anatomical nasopharyngeal curvature, facilitating insertion, reducing tissue trauma and keeping the surgical field clear in head and neck procedures, oral, maxillofacial, ENT and ophthalmic surgery. Manufactured by DEAS S.r.l. (Deaflux Respiratory Products, Italy) and distributed in Spain by Cahnos, it is supplied in a box of 10 individually packaged, sterile units, ready for immediate use.
2. Technical explanation of the device and procedure
In nasotracheal intubation, the tube is inserted through a nostril, advances through the nasopharynx, passes through the glottis and is positioned in the trachea, where it conducts gases from the ventilator or anaesthetic circuit to the lungs. The preformed south-facing curvature (curved towards the chin) follows the anatomy of the upper airway: it facilitates fixation of the tube on the forehead, minimises the risk of kinking and prevents the circuit from interfering with the operative field.
The low-pressure, high-volume cuff, inflated through the pilot balloon with self-sealing valve, seals the trachea and allows positive-pressure ventilation without significant leaks, with lower pressure on the mucosa than high-pressure cuffs. The Murphy eye, a distal lateral orifice smoothly formed and without sharp edges, guarantees an alternative gas passage if the tip of the tube becomes obstructed by secretions or by resting against the tracheal wall. The longitudinal radiopaque line, the double depth marker and the graduation every centimetre allow correct placement to be verified radiographically and clinically. The atraumatic, rounded tip minimises damage during intubation and extubation. It is a single-use device: it must be discarded after each patient to avoid the risk of cross-infection.
3. Technical Specifications
Technical data verified against the official technical data sheet of the ETPN range from manufacturer DEAS S.r.l. (Deaflux Respiratory Products, Italy), document ref. 04906-3e.doc Rev.3, and the ETPN range catalogue from Cahnos.
| Parameter | Specification |
|---|---|
| Reference | ETPN65C02 (Cahnos’s own code) |
| Type | Preformed nasal tracheal tube (RAE/AGT type) with cuff and Murphy eye |
| Access route | Nasal (nasopharyngeal), preformed south-facing design |
| Internal diameter (ID) | 6.5 mm |
| External diameter (OD) | 8.7 mm |
| Total length | 295 mm |
| Inflated cuff diameter | 20 mm |
| Sleeve | Sealing, low pressure and high volume, with pilot balloon and self-sealing valve |
| Murphy eye | Yes, smoothly formed, distal lateral safety orifice |
| Material | Transparent, thermosensitive medical-grade polyvinyl chloride (PVC) |
| Connector | 15 mm universal connector conforming to ISO 5356-1 |
| Marking | Longitudinal radiopaque line, double depth marker and graduation every cm |
| Latex | Natural rubber latex-free |
| Phthalates | DEHP/DOP-free |
| Sterility | Sterile, individually packaged (ethylene oxide per EN 11135-1; residues per ISO 10993-7) |
| Use | Single-use, non-reusable |
| CE classification | Medical device class IIa (Directive 93/42/EEC, notified body TÜV SÜD) |
| Indication | Positive-pressure ventilation and administration of anaesthesia when a ventilation face mask is not viable |
| Presentation | Box of 10 individually packaged units |
ETPN range – preformed nasal tracheal tubes with cuff (same design, presentation and manufacturer, Cahnos references):
| Reference | ID (mm) | OD (mm) | Length (mm) | Cuff Ø (mm) |
|---|---|---|---|---|
| ETPN40C02 | 4.0 | 5.3 | 210 | 14 |
| ETPN45C02 | 4.5 | 6.0 | 225 | 14 |
| ETPN50C02 | 5.0 | 6.7 | 245 | 17 |
| ETPN55C02 | 5.5 | 7.3 | 275 | 17 |
| ETPN60C02 | 6.0 | 8.0 | 285 | 20 |
| ETPN65C02 (this product) | 6.5 | 8.7 | 295 | 20 |
| ETPN70C02 | 7.0 | 9.3 | 310 | 25 |
| ETPN75C02 | 7.5 | 10.0 | 315 | 25 |
| ETPN80C02 | 8.0 | 10.7 | 330 | 26 |
Note: the range of diameters, lengths and cuff diameters in the table comes from the manufacturer’s official technical data sheet (DEAS S.r.l. / Deaflux) and the ETPN range catalogue from Cahnos; do not extend these values to other sizes without technical confirmation. This category of tracheal tubes is manufactured in accordance with the requirements of standard ISO 5361 (tracheal tubes and connectors) and the biocompatibility standards applicable to airway devices; verify the manufacturer’s declaration of conformity and CE marking on each supplied batch.
4. Equivalences and Cross-References
The reference ETPN65C02 is Cahnos’s own code for the range (manufacturer DEAS S.r.l. / Deaflux Respiratory Products, Italy); the family is officially documented as ETPN40C02–ETPN80C02 according to the manufacturer’s technical data sheet (ref. 04906-3e.doc Rev.3). It must not be confused with the official part number of any third-party manufacturer: there is no manufacturer-validated equivalence table for this code; consult Cahnos’s technical department before using it in specifications or tenders.
Comparable at category level (not 1:1 interchangeable): preformed nasal tubes with cuff and Murphy eye from lines such as Rüsch® AGT/RAE nasal and Sheridan® (Teleflex), Shiley™ RAE nasal with TaperGuard™ cuff (Medtronic), Portex™ (SOFT-SEAL™, Smiths Medical / ICU Medical) or BOENMED nasal preformed tubes respond to the same general clinical indication and a RAE nasal-type geometry. Any specific reference in size 6.5 mm must be individually checked against the manufacturer’s official technical data sheet before proposing stock substitution, as cuff technology, curve geometry, balloon material and connection system may differ.
Comparables by category within the Cahnos catalogue (same ID 6.5 mm, other configurations):
| Reference | Description |
|---|---|
| ETAN65C02 | Oral endotracheal tube (Magill type) with cuff and Murphy eye |
| ETPO65C02 | Preformed oral endotracheal tube with cuff and Murphy eye |
| ETSL65C02 | Oral endotracheal tube with cuff, subglottic suction lumen and Murphy eye |
All are comparable at category level, not 1:1 interchangeable without individual technical validation. There is no equivalence table validated by the manufacturer for this reference; consult the technical department before use in tendering.
5. Selection tips and variants of interest
Patient selection (usual use in clinical practice, not manufacturer specification): the ID 6.5 mm size is used in tall paediatric patients, adolescents or adults with narrow airway and nasal cavity; the final choice must always be based on the service protocol and anaesthetic assessment (age, weight, estimated airway and nasal cavity size).
Cuff pressure: inflate through the valve with a syringe and maintain a sealing pressure of 20–30 cmH₂O (usual clinical practice range) to minimise tracheal mucosal damage; monitor with a manometer if available in the service. Before intubation, check the integrity of the cuff and the patency of the pilot balloon and valve.
Position verification: check by bilateral auscultation and, if possible, radiography using the radiopaque line and the tube’s double depth marker.
Preferred indications: oral, maxillofacial, otorhinolaryngological and ophthalmic surgery, where it is important to keep the circuit out of the operative field, as well as mechanical ventilation in intensive care.
Variants of interest in the catalogue: if you need another size, the ETPN series covers from 4.0 mm to 8.0 mm (ETPN40C02 to ETPN80C02). For oral intubation, consider the ETAN (Magill type), ETPO (preformed oral) and ETSL (with subglottic suction lumen) ranges; for intubation requiring kink resistance or frequent repositioning, consider the reinforced tubes in the Cahnos intubation section. Contact the Cahnos technical department for samples, volume conditions and hospital supply lead times.




Reviews