We use our own and third-party cookies to analyse the use of the website and show you advertising related to your preferences, based on a profile built from your browsing habits (for example, pages visited). More information

Cahnos.com Compare invoice

Preformed Nasal Tracheal Tube with Low-Pressure Cuff and Murphy Eye – ID 6.0 mm (Ø ext. 8.0 mm) | Ref. ETPN60C02 | Box of 10 units

36,45 €

Preformed nasal tracheal tube with low-pressure cuff and Murphy eye, with an internal diameter (ID) of 6.0 mm and an external diameter of 8.0 mm. Manufactured from thermosensitive medical-grade polyvinyl chloride (PVC), radiopaque, latex-free and phthalate-free; incorporates a universal 15 mm connector (ISO 5356-1), double depth marker and pilot balloon. Indicated for positive-pressure ventilation and administration of anaesthesia when the use of a face mask is not possible, keeping the surgical field clear in head and neck surgery. Sterile by ethylene oxide, presentation in a box of 10 individually packaged units. Reference Cahnos ETPN60C02.

International shipping. Import duties and taxes are payable by the recipient.

Reference
ETPN60C02

Description

1. General description

The preformed nasal tracheal tube with cuff and Murphy eye (reference ETPN60C02) is the reference solution for airway management via the nasal route in intensive care units, anaesthesia operating theatres, emergency departments and resuscitation. Its preformed design reproduces the nasopharyngeal anatomical curvature, which facilitates insertion, reduces tissue trauma and keeps the surgical field clear in head and neck procedures, oral or dentomaxillary surgery. It is supplied in a box of 10 individually packaged and sterile units, ready for immediate use.

2. Technical explanation of the device

The tube is inserted through a nostril and advances through the nasopharynx to the trachea thanks to its preformed curvature, which also minimises the risk of kinking and displacement during ventilation. At the proximal end it connects to the ventilator, the anaesthesia circuit or the resuscitation bag via the universal 15 mm connector in accordance with standard ISO 5356-1.

The low-pressure, high-volume cuff, inflated via the pilot balloon with Luer-type connection and one-way valve, creates an airtight seal against the tracheal wall that allows positive-pressure ventilation and reduces the risk of leakage and aspiration of pharyngeal or gastric contents, with lower pressure on the mucosa than high-pressure cuffs. The Murphy eye, a lateral opening at the distal end, ensures airflow even if the tube tip becomes occluded by contact with the tracheal wall. The longitudinal radiopaque line, the double depth marker and the graduation every centimetre facilitate position verification by radiological and clinical control. The atraumatic, rounded tip minimises damage during intubation and extubation.

3. Technical Specifications

Technical data verified against the official technical data sheet of the ETPN range from Cahnos (manufacturer DEAS S.r.l. / Deaflux):

Parameter Specification
Access route Nasal (nasopharyngeal), preformed design
Internal diameter (ID) 6.0 mm
External diameter (OD) 8.0 mm
Total length 285 mm
Inflated cuff diameter 20 mm
Sleeve Sealing, low pressure and high volume, with pilot balloon and Luer connection
Murphy eye Yes, smoothly formed, distal lateral safety orifice
Material Transparent, thermosensitive medical-grade polyvinyl chloride (PVC)
Connector Universal 15 mm (ISO 5356-1)
Marking Longitudinal radiopaque line, double depth marker and graduation every cm
Latex Natural rubber latex-free
Phthalates DEHP/DOP-free
Sterility Sterile, individually packaged (ethylene oxide, according to EN 11135-1; residues according to ISO 10993-7)
Use Single-use, non-reusable
CE classification Medical device class IIa (Directive 93/42/EEC, TÜV SÜD certified)
Presentation Box of 10 individually packaged units (carton of 100 units)

ETPN range – preformed nasal tracheal tubes with cuff (references Cahnos):

Reference ID (mm) OD (mm) Length (mm) Cuff Ø (mm)
ETPN40C02 4.0 5.3 210 14
ETPN45C02 4.5 6.0 225 14
ETPN50C02 5.0 6.7 245 17
ETPN55C02 5.5 7.3 275 17
ETPN60C02 (this product) 6.0 8.0 285 20
ETPN65C02 6.5 8.7 295 20
ETPN70C02 7.0 9.3 310 25
ETPN75C02 7.5 10.0 315 25
ETPN80C02 8.0 10.7 330 26

4. Equivalences and Cross-References

Functional equivalence of category (not interchangeable 1:1). This tube belongs to the family of preformed nasal tracheal tubes with cuff, a category represented by manufacturers such as Teleflex (Rüsch® line of preformed nasal tubes with cuff, e.g. reference 111781060 for ID 6.0 according to the manufacturer’s web catalogue) and Medtronic (Shiley™ nasal line with TaperGuard™ cuff, e.g. reference 96360 for ID 6.0 according to authorised distribution catalogues). They all share the same general clinical indication, but differ in cuff technology, the geometry of the preformed curve and the connection system, so they should not be treated as direct equivalents without individual verification of each reference.

The reference ETPN60C02 is a proprietary code of the Cahnos range (manufacturer DEAS S.r.l. / Deaflux Respiratory Products, Italy) and could not be compared with an official third-party manufacturer code in public sources: there is no equivalence table validated by the manufacturer; consult the technical department of Cahnos before use in tenders or as a replacement for stock of other brands.

Comparables by category within the Cahnos catalogue (same ID 6.0 mm):

Reference Description
ETAN60C02 Oral endotracheal tube (Magill type) with cuff and Murphy eye
ETPO60C02 Preformed oral endotracheal tube with cuff and Murphy eye
ETSL60C02 Oral endotracheal tube with cuff, subglottic suction lumen and Murphy eye

5. Selection tips and variants of interest

The ID 6.0 mm is reserved for tall paediatric patients or small-built adults; as a general guide, in adults tubes of 7.0–9.0 mm are used according to build, always selecting the largest size compatible with the patient’s anatomy and the diameter of the nostril. Before intubation, check the integrity of the cuff, the patency of the pilot balloon and valve, and control the sealing pressure (usual recommended cuff pressure: 20–30 cmH₂O; common market practice, verify the centre’s protocol).

If your centre performs head and neck surgery, oral surgery or any procedure in which the surgical field must be kept clear, the preformed nasal tube of the ETPN range (sizes 4.0–8.0 mm) is the indicated option. For prolonged intubations with a high risk of ventilator-associated pneumonia, consider the ETSL range with subglottic suction lumen (ETSL60C02). Contact the technical department of Cahnos for samples, volume conditions and hospital supply lead times.

Additional information

Material

Thermosensitive medical-grade polyvinyl chloride (PVC)

Access route

Nasal (nasopharyngeal), preformed design

Inner diameter

6.0 mm

External diameter od

8.0 mm

Inflated cuff diameter

20 mm

Sleeve

Low pressure and high volume, with pilot balloon and Luer connection

Connector

Universal 15 mm (ISO 5356-1)

Radiopaque line

Yes, longitudinal

Marking

Double depth marker and graduation every cm

Latex

Latex-free

Phthalates

DEHP/DOP-free

Sterilisation

Ethylene oxide (ETO), in accordance with EN 11135-1

Use

Single-use (disposable)

CE classification

Class IIa (Directive 93/42/EEC, TÜV SÜD)

Presentation

Box of 10 individually packaged units

Shelf life

59 months from the date of production

Manufacturer

DEAS S.r.l. (Deaflux Respiratory Products), Italy

Longitud

28.5 cm

Ojo murphy

With Murphy's eye

Reviews

There are no reviews yet.

Be the first to review “Preformed Nasal Tracheal Tube with Low-Pressure Cuff and Murphy Eye – ID 6.0 mm (Ø ext. 8.0 mm) | Ref. ETPN60C02 | Box of 10 units”