Description
1. Introduction
The MININJECTOR (Ref. 770314) is a single-use uterine injector specifically designed for sonohysterography, also called saline infusion sonohysterography (SIS). It allows controlled and atraumatic instillation of physiological saline solution into the uterine cavity, improving ultrasound visualisation of the cavity and endometrial structures. It is presented in a box of 10 units with individual sterile packaging, ready for use in gynaecology clinics, assisted reproduction units and diagnostic radiology services.
2. Technical explanation of the equipment and procedure
Sonohysterography combines transvaginal ultrasound with instillation of physiological saline into the uterine cavity: the fluid distends the cavity and acts as an ultrasound contrast medium, allowing identification of polyps, submucosal fibroids, adhesions or endometrial thickening. To this end, the MININJECTOR incorporates a transparent catheter with radiopaque line along its entire length and a measurement scale in centimetres that helps the professional control the depth of placement. The tip, soft and with a single distal orifice, allows atraumatic outflow of the saline solution. The set is completed with a rigid introducer that facilitates passage through the cervix and an adjustable positioner made of soft plastic material that acts as a stop on the external cervical os, preventing fluid reflux and ensuring stable fixation throughout the examination. The catheter ends in a female luer-lock connector for secure connection of the syringe.
3. Technical Specifications
| Parameter | Specification |
| Device type | Uterine injector for sonohysterography (saline solution infusion) |
| Reference | 770314 |
| Catheter calibre | 5 Fr |
| Catheter | Transparent, with radiopaque line along its entire length and measurement scale in centimetres for depth control |
| Tip | Soft, with a distal orifice for the outflow of the saline solution |
| Introducer | Rigid, facilitates introduction through the cervix |
| Positioner | Adjustable, made of soft plastic material; acts as a fixation stop and prevents reflux |
| Connection | Female luer-lock connector for syringe |
| Compatible medium | Physiological saline solution (saline) |
| Application | Sonohysterography / saline infusion under transvaginal ultrasound |
| Sterilisation | Sterile, single-use, individually packaged |
| Presentation | Box of 10 units |
4. Equivalences and Cross-References
There is no official equivalence table validated by the manufacturer that cross-references reference 770314 with those of other brands. The references listed below belong to the same functional segment (5 Fr catheters and injectors for sonohysterography/hysterosonography), but are not interchangeable 1:1: they differ mainly in the fixation system (distal balloon versus the MININJECTOR’s adjustable positioner) and in the configuration of the set.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
| Cook Medical | Goldstein Sonohysterography Catheter (5.3 Fr × 26 cm), adjustable acorn positioner | G17149 | Comparable by category (not interchangeable 1:1) | Balloon-free catheter with adjustable acorn positioner that minimises reflux at the external cervical os, similar to the MININJECTOR’s stop; closed rounded tip with oval lateral port. Verified on Cook Medical’s official website. |
| Bioteque | Bellota SonoCatheter with cervical cone, 5 Fr | UIC5FA | Comparable by category (not interchangeable 1:1) | Access to the uterine cavity for saline infusion in sonohysterography; includes placement sheath that avoids the use of tenaculum and sealing cervical cone. Verified on Bioteque’s official website. |
| CooperSurgical | H/S Catheter Set (hysterosonography and hysterosalpingography), 5 Fr, with integrated stylet | 61-5005 | Comparable by category (not interchangeable 1:1) | Catheter with distal latex-free balloon and integrated stylet; verified in CooperSurgical official datasheet and GUDID. Balloon fixation system versus the MININJECTOR’s adjustable positioner. |
| CooperSurgical | GyneCath Uterine Infusion Catheter, 5 Fr (1.5 cc syringe) | 19610 | Comparable by category (not interchangeable 1:1) | Flexible radiopaque catheter of 30 cm with 3 cc latex-free balloon and double lumen; verified on CooperSurgical official website. Balloon fixation configuration versus mechanical stop. |
| MedGyn Products | HSG Catheter with rigid insertion sheath, 5 Fr, latex-free | 022726 | Comparable by category (not interchangeable 1:1) | Catheter with wide balloon and rigid insertion sheath; includes syringe for balloon inflation; verified on MedGyn official website and authorised distributor catalogues. |
| Marina Medical Instruments | HSG-Sono Catheter SureFlex, 5 Fr, 5 cc balloon with 6 cc syringe (box of 10) | MM1-605 | Comparable by category (not interchangeable 1:1) | Flexible latex-free catheter with 5 cc balloon and 6 cc syringe; balloon volume and set components differ. Verified on Marina Medical Instruments official website. |
Other brands in the same functional segment are Argon Medical (HSG Catheters) and Bexen Medical (radiopaque polyurethane HSG catheter with soft latex-free balloon, available in 5 Fr and 7 Fr, whose reference code is not published in its official datasheet and must be requested from the manufacturer). There is no equivalence table validated by the manufacturer; consult the technical department before use in tenders or as a product substitute. Exact equivalences of gauge, length and fixation system must always be verified against the technical datasheets and instructions for use of each brand.
5. Selection tips and variants of interest
- Same family with stylet: if the cervical canal presents access difficulty or stenosis, the MININJECTOR CM (Ref. 880314) version incorporates a malleable plunger or mandrel that gives rigidity to the catheter during insertion while maintaining the same 5 Fr gauge.
- Hysterosalpingography (HSG) with contrast medium and fixation balloon: the SOFTINJECTOR family (Refs. 720240, 720226 and 720250 in 5, 6 and 7 Fr) offers a flexible double-lumen catheter with a 2.5 cc balloon, and the RIGIDINJECTOR family (Refs. 720239, 720225 and 720251) the rigid catheter version. The Injector Plus (Ref. 730226, 6 Fr semi-rigid) and the HANDLIN injector manipulator (Ref. 720199) complete the range for injection needs with mandrel or active uterine manipulation.
- Recommendations for use: always perform prior hysterometry to avoid contact with the uterine fundus, lubricate the tip before insertion, advance the positioner (stop) over the cervix to prevent reflux and connect the syringe to the luer-lock connector to inject the saline solution slowly. Verify connector compatibility with the department’s syringe.
- Single-use device: do not reuse; use only if the packaging is intact, use immediately after opening and destroy after use in accordance with current regulations. It must be used only by specialised medical personnel.