Description
1. Introduction
The Luer Lock non-return valve Ref. VALARB (SKU 0322035) is the venous access accessory that provides safety and control to your centre’s infusion circuits. It allows injection and withdrawal through needle-free closed systems, with direct and reliable access to the venous line, minimising the risk of contamination and accidental needlestick injuries for healthcare staff. With a gravity opening pressure of only 0.3 PSI, it is the ideal solution for gravity infusion lines and conventional fluid therapy. It is supplied in a box of 200 units, in immediate stock for hospital distribution and replenishment.
2. Technical explanation of the device
This valve acts as a needle-free access point with a minimum opening pressure of 0.3 PSI, allowing fluid to flow by simple gravity without the need for external pressure. When a syringe or a device with a male luer is connected, the silicone septum opens and allows injection or withdrawal; upon removal of the device, the septum automatically seals the line, maintaining the integrity of the closed system and reducing manipulations on the circuit.
The Luer Lock connection ensures a firm and leak-proof coupling with extension lines, three-way stopcocks, catheters and standard infusion equipment on the market. It is recommended to disinfect the septum with an appropriate antiseptic before each connection and to respect the maximum usage time indicated in the device instructions.
3. Technical Specifications
The product specifications are summarised in the following table:
| Parameter | Specification |
| Reference | VALARB (SKU 0322035) |
| Valve type | Non-return, single-use |
| Opening pressure | 0.3 PSI (gravity) |
| Connection | Luer Lock |
| Sealing mechanism | Self-sealing silicone septum |
| Operating mode | Needle-free closed systems: injection and withdrawal |
| Main function | Direct venous access and automatic sealing upon device removal |
| Presentation | Box of 200 units (carton of 10 x 200) |
The VALARB valve is part of the Cahnos family of Luer Lock valves for needle-free closed systems:
| Reference | Type | Connection | Pressure / Characteristic | Presentation |
| VALARB | Non-return | Luer Lock | Gravity opening 0.3 PSI | Box of 200 units |
| VALASB | Anti-siphon | Luer Lock | Minimum opening 3.5 PSI | Box of 200 units |
| VALBDDN | Bidirectional | Luer Lock | Working pressure 60 PSI | Box of 200 units |
4. Equivalences and Cross-References
Within the Cahnos catalogue, the range of needle-free Luer Lock valves offers three features differentiated by their opening pressure: the non-return (0.3 PSI by gravity, the subject of this datasheet), the anti-siphon (3.5 PSI) and the bidirectional (60 PSI). At market level, there are needle-free connectors from other manufacturers that share the same general clinical indication (venous access in a closed system), although with opening pressures, materials and geometries that may differ, so they should be considered comparable by category and not interchangeable 1:1:
| Manufacturer / Brand | Family / Model | Code / SKU | Equivalence level | Technical differences / Remarks |
| Cahnos | Luer Lock non-return valve | VALARB (SKU 0322035) | Exact reference equivalence | Reference covered by this datasheet; it matches the official nomenclature of the manufacturer’s catalogue for the infusion range. |
| Cahnos | Luer Lock anti-siphon valve | VALASB | Comparable by category | Same family of needle-free closed system; the opening pressure differs (0.3 PSI by gravity versus 3.5 PSI). |
| Cahnos | Luer Lock bidirectional valve | VALBDDN | Comparable by category | It allows injection and withdrawal, but is designed for a working pressure of 60 PSI. |
| BD | Q-Syte™ needle-free connector (silicone septum) | Consult BD’s official catalogue | Comparable by category | Needle-free connector with septum; opening pressure and material may differ. |
| B. Braun | Safsite® / Caresite® ranges | Consult B. Braun’s official catalogue | Comparable by category | Needle-free access points compatible with Luer Lock; different pressure characteristics. |
| Vygon | Vadsite | Consult Vygon’s official catalogue | Comparable by category | Closed system connector for infusion and withdrawal; verify opening pressure. |
Important notice: there is no equivalence table validated by the manufacturer that allows this valve to be interchanged 1:1 with other connectors on the market. The above references are offered exclusively as a functional comparison by category. Consult our technical department before using this information in tendering processes or replacement of material in use.
5. Selection tips and variants of interest
- Select the non-return VALARB valve (0.3 PSI) when your protocol uses low-pressure gravity infusion and you need a needle-free access that opens without resistance.
- If you administer drugs via an injection pump and need to prevent gravity drip in the event of equipment failure, consider the anti-siphon VALASB valve (3.5 PSI).
- If you need an access that supports high working pressures and bidirectional flow, the bidirectional VALBDDN valve (60 PSI) is the indicated variant.
- Disinfect the silicone septum with an appropriate antiseptic before each connection and respect the maximum usage time indicated in the device instructions for central or peripheral lines.
- Check compatibility with your usual circuit: the Luer Lock connection is standard and fits extension lines, three-way stopcocks and conventional infusion equipment.
- For volume purchases, periodic replenishment or configuration of a complete mix of valves for your centre, our technical-commercial team will advise you and validate compatibility with your equipment before formalising the order.