Description
Continuous, safe infusion without electronic equipment
The ELASPUMP® 275 ml elastomeric pump with UV protection is the ideal solution for continuous drug administration in hospitals, clinics, home hospitalisation units and outpatient centres. As it does not rely on batteries, electricity or maintenance, it offers a very competitive cost per treatment and eliminates the risk of cross-contamination between patients. Its UV protection makes it especially recommended for photosensitive drugs, guaranteeing the integrity of the drug at all times during infusion.
Technical description of the device
The ELASPUMP® pump consists of a silicone elastomer housed inside a rigid, transparent, hermetic casing that prevents accidental leaks and allows visual inspection of the drug level from the outside. Its operating principle is based on the elastic recovery pressure of the elastomer, which drives the medication through the tubular system at a constant, calibrated flow rate of 8 ml/hour, ensuring homogeneous infusion over the approximately 1 day and 10 hours of autonomy.
- Silicone elastomer in a rigid, transparent, hermetic casing, with visual inspection of the drug level.
- One-way filling valve with closed circuit and flat bottom, compatible with a standard syringe.
- Hydrophobic filter on the capsule of the male luer-lock connector, which allows priming the tubular system without removing the capsule, preserving internal sterility until actual use.
- In-line particle filter, which retains impurities and guarantees the purity of the administered drug.
- Removable sliding clamp, easy to remove if necessary.
- Colour-coded label with flow rate imprint, for immediate identification of the flow.
- Patient label, for easy personalisation of the treatment.
- Full traceability: variable data and product code printed on the outer casing and on the pump body.
Technical specifications
| Characteristic | Specification |
|---|---|
| Capacity | 275 ml |
| Flow type | Fixed continuous (factory-calibrated) |
| Infusion flow rate | 8 ml/hora |
| Approximate duration | 1 day + 10 hours (at full volume) |
| Protection | UV protection for photosensitive drugs |
| Drive mechanism | Silicone elastomer in rigid and transparent casing |
| Filling system | Closed-circuit one-way valve, compatible with standard syringe |
| Connection | Male Luer-lock with cap and hydrophobic filter |
| Filtration | Hydrophobic filter + in-line particle filter |
| Routes of administration | Intravenous, arterial, epidural and subcutaneous |
| Clinical indications | Chemotherapy, pain therapy, antiviral and antibiotic treatments |
| Power supply | No battery or power source |
| Use | Single-use, sterile |
| Composition | DEHP-free, PVC-free, latex-free |
| Brand | ELASPUMP® |
| Reference | 093005/7/B |
Indicative duration of the ELASPUMP® 275 ml range with UV protection:
| Reference | Flow rate | Approximate duration |
|---|---|---|
| 093005/65/B | 7 ml/h (continuous, UV) | 1 day + 15 hours |
| 093005/7/B (current product) | 8 ml/h (continuous, UV) | 1 day + 10 hours |
| 093005/8/B | 10 ml/h (continuous, UV) | 1 day + 3 hours |
| 093002/5/B | 4 ml/h (continuous, UV, 100 ml) | 1 day + 1 hour |
Approximate values calculated at full volume; actual duration may vary slightly depending on use conditions and drug characteristics.
Equivalences and cross-references
The ELASPUMP® 275 ml with UV protection and continuous flow (Ref. 093005/7/B) belongs to the range of elastomeric pumps of the ELASPUMP® family, and is complemented by the following references available at Cahnos:
- 093005/65/B – ELASPUMP® 275 ml with UV protection, continuous flow of 7 ml/h (1 day + 15 hours).
- 093005/8/B – ELASPUMP® 275 ml with UV protection, continuous flow of 10 ml/h (1 day + 3 hours).
- 093002/5/B – ELASPUMP® 100 ml with UV protection, continuous flow of 4 ml/h (1 day + 1 hour).
- 093005/M1 · M2 · M3 – ELASPUMP® 275 ml variable flow (4 and 8 speeds), same capacity with adjustable flow rate.
- 093005/125 and 093005/250 – ELASPUMP® 275 ml high continuous flow (125 and 250 ml/h) for rapid infusions.
- 093001/4 – ELASPUMP® 60 ml continuous flow (2 ml/h).
Due to its pressurised elastomer system with luer-lock connector, hydrophobic filter, calibrated flow rate and UV protection, this device is comparable to disposable 275 ml continuous flow elastomeric pumps from other manufacturers used in controlled analgesia and ambulatory infusion, and is interchangeable in most clinical protocols for continuous infusion and patient-controlled analgesia (PCA). Consult our technical-commercial team to confirm exact compatibility with your usual reference supplier.
Selection tips and variants of interest
- Volume to infuse: the 275 ml version is ideal for medium and long-duration therapies. For smaller volumes, the 60 ml (093001/4) and 100 ml (093002/5/B) versions are available.
- Required flow rate: if your protocol is fixed, this continuous flow model at 8 ml/h offers an economical and reliable solution. If you need to adjust the speed, consider the 275 ml variable flow versions (093005/M1, M2 and M3).
- Photosensitive drugs: for light-sensitive medications, always choose the versions with UV protection (093005/65/B, 093005/7/B, 093005/8/B and 093002/5/B).
- Accessories: combine the pump with 1-way or 3-way extension sets with stopcock to adapt the system to the patient’s needs.
Contact our technical-commercial department to request samples, technical data sheets or special conditions for tenders and hospital purchases.