Description
1. Continuous, safe infusion without electronic equipment
The ELASPUMP® 40 ml elastomeric infusion pump (Ref. 093000/8) is the most compact option in the Cahnos continuous flow range: a single-use, sterile device free from DEHP, PVC and latex, designed to administer drugs at a constant flow rate of 10 ml/hour without relying on electronic pumps, batteries or maintenance. With the full 40 ml volume it offers an approximate autonomy of 0 days + 4 hours, making it the ideal solution for short continuous infusion regimens, outpatient therapy, home hospitalisation and controlled analgesia, at a very competitive cost per treatment and with full product traceability.
2. Technical description and operating principle
The ELASPUMP® pump operates through the elastic recovery pressure of a silicone elastomer contained inside a sealed, rigid and transparent housing, which prevents accidental leaks and allows visual inspection of the drug level from the outside without having to open the device. This pressure drives the medication through the tubular system at a constant, factory-calibrated flow rate of 10 ml/hour, resulting in an approximate autonomy of 0 days + 4 hours with the full 40 ml volume.
- Silicone elastomer encapsulated in a sealed, rigid and transparent housing, with visual inspection of the drug level from the outside.
- One-way filling valve with closed circuit and flat bottom, compatible with a standard syringe.
- Hydrophobic filter on the male luer-lock connector cap, which allows the line to be primed without removing the cap and preserves internal sterility until actual use.
- In-line particle filter integrated into the circuit, which retains impurities and guarantees the purity of the administered drug.
- Removable sliding clamp, easy to remove if necessary.
- Colour-coded label with flow rate printed on it, for quick and unambiguous identification of the device.
- Patient label, designed for easy treatment personalisation.
- Full traceability: variable data and product code printed on both the outer housing and the pump body.
3. Technical Specifications
3.1. Product technical data sheet
| Characteristic | Specification |
| Brand | ELASPUMP® |
| Reference | 093000/8 |
| Capacity | 40 ml |
| Flow type | Fixed continuous (factory-calibrated) |
| Infusion flow rate | 10 ml/hora |
| Approximate duration | 0 days + 4 hours (approx. 4 hours at full volume) |
| Drive mechanism | Silicone elastomer in rigid and transparent casing |
| Filling system | Closed-circuit one-way valve, compatible with standard syringe |
| Connection | Male Luer-lock with cap and hydrophobic filter |
| Filtration | Hydrophobic filter + in-line particle filter |
| Routes of administration | Intravenous, arterial, epidural and subcutaneous |
| Clinical indications | Chemotherapy, pain therapy, antiviral and antibiotic treatments |
| Power supply | Does not require battery or electrical power source |
| Use | Single-use, sterile |
| Composition | DEHP-free, PVC-free, latex-free |
Indicative values calculated at full volume; actual duration may vary slightly depending on conditions of use and the characteristics of the drug.
3.2. Indicative infusion duration by flow rate (40 ml volume)
| Flow rate | Approximate duration |
| 1 ml/hora | 1 day + 16 hours (approx. 40 hours) |
| 2 ml/hora | 20 hours |
| 4 ml/hora | 10 hours |
| 5 ml/hora | 8 hours |
| 8 ml/hora | 5 hours |
| 10 ml/hour (current product) | 0 days + 4 hours |
3.3. ELASPUMP® continuous flow range available from Cahnos
| Reference | Capacity | Flow rate | Approximate duration |
| 093000/8 (current product) | 40 ml | 10 ml/hora | 0 days + 4 hours |
| 093001/4 | 60 ml | 2 ml/hora | 1 day + 6 hours |
| 093001/8/B | 60 ml | 10 ml/hora | 0 days + 6 hours |
| 093002/5 | 100 ml | 4 ml/hora | 1 day + 1 hour |
| 093004/8/B | 200 ml | 10 ml/hora | 0 days + 20 hours |
| 093005/8/B | 275 ml | 10 ml/hora | 1 day + 3 hours |
4. Equivalences and Cross-References
The ELASPUMP® 40 ml (Ref. 093000/8) belongs to the continuous flow elastomeric pump range of the ELASPUMP® family and complements the following references available from Cahnos:
- 093001/8/B: ELASPUMP® 60 ml with UV protection, continuous flow of 10 ml/hour (0 days + 6 hours), same technology with a larger volume.
- 093002/8/B: ELASPUMP® 100 ml with UV protection, continuous flow of 10 ml/hour (0 days + 10 hours).
- 093004/8/B: ELASPUMP® 200 ml with UV protection, continuous flow of 10 ml/hour (0 days + 20 hours).
- 093005/8/B: ELASPUMP® 275 ml with UV protection, continuous flow of 10 ml/hour (1 day + 3 hours).
- 093001/4: ELASPUMP® 60 ml continuous flow (2 ml/hour = 1 day + 6 hours), same family without UV protection.
- 093002/5: ELASPUMP® 100 ml continuous flow, DEHP-free (4 ml/hour = 1 day + 1 hour).
Thanks to its 40 ml capacity, pressurised elastomer system with luer-lock connector, hydrophobic filter and fixed factory-calibrated flow rate, this device is comparable to disposable continuous flow elastomeric pumps from other manufacturers used in outpatient infusion, controlled analgesia and short-term home intravenous therapy, and is interchangeable in most continuous infusion clinical protocols. Please consult our technical-commercial team to confirm exact compatibility with your usual reference supplier.
5. Selection tips and variants of interest
- Volume and duration: the 40 ml version at 10 ml/hour is ideal for short-duration therapies (approx. 4 hours at full volume). If your protocol requires longer autonomy, versions of 60 ml (Refs. 093001/4 and 093001/8/B), 100 ml (Ref. 093002/5) and up to 275 ml (Ref. 093005/8/B) are available with the same operating principle.
- Required flow rate: if your protocol is fixed, this continuous flow version at 10 ml/hour offers the best cost-benefit ratio. If you need to adjust the rate during treatment, consider the ELASPUMP® variable flow versions (Refs. 093004/M1-M3 and 093005/M1-M3).
- Photosensitive drugs: for light-sensitive medications, choose the ELASPUMP® versions with UV protection (Refs. 093001/8/B, 093002/8/B, 093004/8/B and 093005/8/B).
- Recommended accessories: combine the pump with 1-way or 3-way infusion extension lines with stopcock to adapt the system to the patient’s needs.
- Purchasing conditions: contact our technical-commercial department to request samples, technical data sheets or special conditions for tenders and hospital purchases.