Description
1. Commercial Introduction
The Elastomeric Infusion Pump ELASPUMP® 100 ml at 10 ml/hour (Ref. 093002/8) is the ideal solution for the continuous and controlled administration of medications in hospitals, clinics, home hospitalisation units and outpatient care centres. It is a single-use, sterile device free of DEHP, PVC and latex, which allows drugs to be infused at a constant, calibrated flow rate of 10 ml/hour without relying on electronic pumps, batteries or maintenance, with a very competitive cost per treatment and no risk of cross-contamination between patients. Its ease of handling makes it a practical option for both healthcare professionals and patients themselves, especially in short-duration regimens of approximately 10 hours.
2. Technical description and operating principle
The ELASPUMP® pump consists of a silicone elastomer contained inside a hermetic, rigid and transparent housing, which prevents accidental leaks and allows visual inspection of the drug level from the outside without having to open the device. Its operating principle is based on the elastic recovery pressure of the elastomer, which propels the medication through the tubular system at a constant flow rate calibrated in the factory of 10 ml/hour, resulting in an approximate autonomy of 0 days + 10 hours with the full volume of 100 ml.
- Silicone elastomer: encapsulated in a hermetic, rigid and transparent housing, leak-free and with visual inspection of the level.
- One-way filling valve: closed-circuit with flat bottom, compatible with a standard syringe, which facilitates loading of the elastomer.
- Hydrophobic filter: on the male luer-lock connector cap, allows priming of the line without removing the cap and preserves internal sterility until actual use.
- In-line particle filter: retains impurities and guarantees the purity of the administered drug.
- Removable sliding clamp: easy to remove when necessary.
- Colour-coded label: with flow rate printing for quick and immediate identification of the flow.
- Patient label: designed for easy personalisation of the treatment.
- Full traceability: variable data and product code printed on the external housing and on the pump body.
3. Technical Specifications
General device features:
| Characteristic | Specification |
| Brand | ELASPUMP® |
| Reference | 093002/8 |
| Capacity | 100 ml |
| Flow type | Fixed continuous (factory-calibrated) |
| Infusion flow rate | 10 ml/hora |
| Approximate duration | 0 days + 10 hours (approx. 10 hours at full volume) |
| Drive mechanism | Silicone elastomer in rigid and transparent casing |
| Filling system | Closed-circuit one-way valve, compatible with standard syringe |
| Connection | Male Luer-lock with cap and hydrophobic filter |
| Filtration | Hydrophobic filter + in-line particle filter |
| Routes of administration | Intravenous, arterial, epidural and subcutaneous |
| Clinical indications | Chemotherapy, pain therapy, antiviral and antibiotic treatments |
| Power supply | Does not require battery or electrical power source |
| Use | Single-use, sterile |
| Composition | DEHP-free, PVC-free, latex-free |
ELASPUMP® continuous flow range of 100 ml available at Cahnos (the /B references incorporate UV protection for photosensitive drugs):
| Reference | Capacity | Flow rate | Approximate duration |
| 093002/8 (current product) | 100 ml | 10 ml/hora | 0 days + 10 hours |
| 093002/5 | 100 ml | 4 ml/hora | 1 day + 1 hour |
| 093002/6 | 100 ml | 5 ml/hora | 0 days + 20 hours |
| 093002/7 | 100 ml | 8 ml/hora | 0 days + 12 hours |
| 093002/8/B | 100 ml | 10 ml/hora | 0 days + 10 hours |
| 093001/4 | 60 ml | 2 ml/hora | 1 day + 6 hours |
Indicative values calculated at full volume; actual duration may vary slightly depending on usage conditions and drug characteristics.
4. Equivalences and Cross-References
The ELASPUMP® 100 ml at 10 ml/hour (Ref. 093002/8) belongs to the manufacturer’s family of continuous flow elastomeric pumps and is complementary to the following references available at Cahnos:
| Reference | Description |
| 093002/8/B | ELASPUMP® 100 ml with UV protection, same flow rate of 10 ml/hour, for photosensitive drugs |
| 093002/5 | ELASPUMP® 100 ml, continuous flow of 4 ml/hour (1 day + 1 hour), same capacity with lower flow rate |
| 093002/7 | ELASPUMP® 100 ml, continuous flow of 8 ml/hour (0 days + 12 hours), same capacity with intermediate flow rate |
| 093001/4 | ELASPUMP® 60 ml continuous flow (2 ml/hour = 1 day + 6 hours), for smaller volumes |
| 093002/M1 · M3 | ELASPUMP® 100 ml variable flow (4 and 8 speeds), DEHP-free, for protocols with adjustable flow rate |
| 093004/8/B | ELASPUMP® 200 ml with UV protection, 10 ml/hour (0 days + 20 hours), same flow rate with larger volume |
| 093005/8/B | ELASPUMP® 275 ml with UV protection, 10 ml/hour (1 day + 3 hours), for longer therapies |
Due to its 100 ml capacity, its pressurised elastomer system with luer-lock connector, hydrophobic filter and fixed calibrated flow rate, this device is comparable to disposable continuous flow elastomeric pumps from other manufacturers used in ambulatory infusion, controlled analgesia and home intravenous therapy, being interchangeable in most clinical protocols for continuous infusion and patient-controlled analgesia (PCA). Consult our technical-commercial team to confirm exact compatibility with your usual reference supplier.
5. Selection tips and variants of interest
- Volume to infuse: the 100 ml at 10 ml/hour version is ideal for short-duration therapies (approx. 10 hours at full volume). For smaller volumes, the 60 ml version (Ref. 093001/4) is available, and for longer therapies, the 200 ml and 275 ml versions.
- Required flow rate: if your protocol is fixed, this continuous flow version at 10 ml/hour offers the best cost-benefit ratio. If you need to adjust the speed during treatment, consider the variable flow versions of 100 ml (Refs. 093002/M1 and 093002/M3).
- Photosensitive drugs: for light-sensitive medications, choose the ELASPUMP® version with UV protection (Ref. 093002/8/B), with identical capacity and flow rate.
- Recommended accessories: combine the pump with 1-way or 3-way infusion extension lines with stopcock to adapt the system to the patient’s needs.
- Purchase conditions: contact our technical-commercial department to request samples, technical data sheets or special conditions for tenders and hospital purchases.