Description
1. Commercial Introduction
The elastomeric infusion pump ELASPUMP® 100 ml (Ref. 093002/7) is the ideal solution for the constant and controlled administration of medications in hospitals, clinics, home hospitalisation units and outpatient care centres. It is a single-use, sterile device that is free from DEHP, PVC and latex, which allows drugs to be infused at a calibrated flow rate of 8 ml/hour without relying on electronic pumps, batteries or maintenance, with a very competitive cost per treatment and no risk of cross-contamination between patients. Its ease of handling makes it the practical choice for both healthcare professionals and patients themselves in short-duration infusion regimens.
2. Technical description and operating principle
The ELASPUMP® pump operates through the elastic recovery pressure of a silicone elastomer contained within a hermetic, rigid and transparent housing, which prevents accidental leaks and allows visual inspection of the drug level from the outside without having to open the device. This pressure drives the medication through the tubular system at a constant flow rate calibrated at the factory of 8 ml/hour, which translates into an approximate autonomy of 0 days + 12 hours (about 12 hours) with the full 100 ml volume. Its technical features include:
- Silicone elastomer: encapsulated in a hermetic, rigid and transparent housing, with visual level inspection and no accidental leaks.
- Closed-circuit one-way filling valve: with a flat bottom, compatible with a standard syringe, which facilitates elastomer loading.
- Hydrophobic filter on male luer-lock cap: allows the tubular system to be primed once the elastomer is filled without removing the cap, ensuring greater internal sterility until actual use.
- In-line particle filter: retains impurities and guarantees the purity of the administered drug.
- Removable sliding clamp: easy to remove when necessary.
- Colour-coded label: with the flow rate printed for immediate and simple identification of the patient access flow.
- Patient label: designed for easy personalisation of the treatment.
- Full traceability: variable data and product code printed on the external housing and on the pump body.
3. Technical Specifications
| Characteristic | Specification |
|---|---|
| Brand | ELASPUMP® |
| Reference | 093002/7 |
| Capacity | 100 ml |
| Flow type | Fixed continuous (factory-calibrated) |
| Infusion flow rate | 8 ml/hora |
| Approximate duration | 0 days + 12 hours (approx. 12 hours at full volume) |
| Drive mechanism | Silicone elastomer in rigid and transparent casing |
| Filling system | Closed-circuit one-way valve, compatible with standard syringe |
| Connection | Male Luer-Lock connector with cap and hydrophobic filter |
| Filtration | Hydrophobic filter + in-line particle filter |
| Routes of administration | Intravenous, arterial, epidural and subcutaneous |
| Clinical indications | Chemotherapy, pain therapy, antiviral and antibiotic treatments |
| Power supply | Does not require battery or electrical power source |
| Use | Single-use, sterile |
| Composition | DEHP-free, PVC-free, latex-free |
Indicative values calculated at full volume; actual duration may vary slightly depending on usage conditions and drug characteristics.
Indicative infusion duration by flow rate (100 ml volume):
| Flow rate | Approximate duration |
|---|---|
| 2 ml/hora | 2 days + 2 hours |
| 4 ml/hora | 1 day + 1 hour |
| 5 ml/hora | 0 days + 20 hours |
| 8 ml/hour (current product) | 0 days + 12 hours |
| 10 ml/hora | 0 days + 10 hours |
ELASPUMP® continuous flow range of 100 ml available at Cahnos:
| Reference | Capacity | Flow rate | Approximate duration |
|---|---|---|---|
| 093002/4/B | 100 ml | 2 ml/hora | 2 days + 2 hours |
| 093002/5/B | 100 ml | 4 ml/hora | 1 day + 1 hour |
| 093002/6/B | 100 ml | 5 ml/hora | 0 days + 20 hours |
| 093002/7 (current product) | 100 ml | 8 ml/hora | 0 days + 12 hours |
| 093002/8/B | 100 ml | 10 ml/hora | 0 days + 10 hours |
4. Equivalences and Cross-References
The ELASPUMP® 100 ml continuous flow (Ref. 093002/7) belongs to the range of elastomeric pumps of the ELASPUMP® family, being complementary to the following references available at Cahnos:
- 093002/7/B: same capacity and flow rate (100 ml, 8 ml/hour) with UV protection for photosensitive drugs; direct equivalent of this model in protected version.
- 093002/5/B: ELASPUMP® 100 ml with UV protection, continuous flow of 4 ml/hour (1 day + 1 hour), same capacity with lower flow rate.
- 093002/6/B: ELASPUMP® 100 ml with UV protection, continuous flow of 5 ml/hour (0 days + 20 hours).
- 093002/8/B: ELASPUMP® 100 ml with UV protection, continuous flow of 10 ml/hour (0 days + 10 hours), same capacity with higher flow rate.
- 093004/7: ELASPUMP® 200 ml continuous flow (8 ml/hour = 1 day + 1 hour), same flow rate with larger volume.
- 093005/7: ELASPUMP® 275 ml continuous flow (8 ml/hour = 1 day + 10 hours), same flow rate with larger volume.
- 093001/4: ELASPUMP® 60 ml continuous flow (2 ml/hour = 1 day + 6 hours), for smaller volumes.
- 093002/M1 · M2 · M3: ELASPUMP® 100 ml variable flow (4 and 8 speeds), DEHP-free, for protocols with adjustable flow rate.
Due to its 100 ml capacity, its pressurised elastomer system with luer-lock connector, hydrophobic filter and fixed flow rate calibrated at the factory, this device is comparable to disposable continuous-flow elastomeric pumps from other manufacturers used in ambulatory infusion, controlled analgesia and home intravenous therapy, being interchangeable in most clinical protocols for continuous infusion and patient-controlled analgesia (PCA). Please consult our technical-commercial team to confirm exact compatibility with your usual reference supplier.
5. Selection tips and variants of interest
- Volume to infuse: the 100 ml version at 8 ml/hour is ideal for short-duration therapies (approx. 12 hours at full volume), such as outpatient chemotherapy regimens, pain therapy or antibiotic therapy. For smaller volumes, the 60 ml version is available (Ref. 093001/4) and for longer therapies, the 200 ml and 275 ml versions (Refs. 093004/7 and 093005/7).
- Required flow rate: if your protocol is fixed, this continuous flow version at 8 ml/hour offers the best cost-benefit ratio. If you need to adjust the speed during treatment, consider the 100 ml variable flow versions (Refs. 093002/M1 and 093002/M3).
- Photosensitive drugs: if the medication is light-sensitive, choose the ELASPUMP® version with UV protection (Ref. 093002/7/B) or the rest of the UV-protected range in 100, 200 and 275 ml.
- Recommended accessories: combine the pump with 1-way or 3-way infusion extension lines with stopcock to adapt the system to the patient’s needs.
- Purchasing conditions: contact our technical-commercial department to request samples, technical data sheets or special conditions for tenders and hospital purchases.