Description
1. Introduction
The Dixion BabyGuard I-1125 neonatal transport incubator is the reference professional solution for neonatal intensive care units (NICUs), neonatology departments and medical transport teams that need to maintain a stable and controlled thermal environment for the newborn during any type of transfer. Its specific transport design, with internal battery and alternative AC/DC power supply, allows the patient to be transferred with full continuity of care both within the hospital and in an ambulance, without interruptions to the microclimate or monitoring. At Cahnos we have this incubator available with technical and commercial advice for institutional purchases, tenders and projects for equipping new neonatal units.
2. Technical explanation of the equipment
The BabyGuard I-1125 functions as a precision environmental control chamber for the neonate. A microcomputer manages two thermal regulation modes: air mode, which regulates the air temperature inside the hood (25-37 °C), and baby mode, which automatically adjusts the temperature according to the patient’s actual skin temperature using a sensor with an accuracy of ±0.2 °C. The double-walled acrylic-glass hood improves thermal insulation and reduces fluctuations, and its side door allows the mattress tray to be removed for easy access to the patient. The equipment incorporates passive air humidification, an oxygen supply system with cylinders, an observation lamp and alternative power supply via alternating current (AC) or direct current (DC), with a rechargeable internal battery that guarantees continuous operation for up to 90 minutes, sufficient for intra-hospital and inter-hospital transfers. The height-adjustable stand, IV pole and mattress complete a package designed for the daily work of healthcare staff.
3. Technical Specifications
The official technical values from the manufacturer Dixion for the BabyGuard I-1125 transport incubator (SKU I-1125) are listed below:
| Parameter | Specification |
| Set type | Closed neonatal transport incubator for intensive care |
| Control mode | Air mode and baby mode controlled by microcomputer |
| Air temperature control range | 25-37 °C (override mode >37 °C) |
| Baby temperature control range | 34-37 °C (override mode up to 37-37.5 °C) |
| Skin temperature sensor accuracy | ±0.2 °C |
| Temperature uniformity | ≤1.5 °C |
| Noise level inside the hood | ≤52 dB |
| Humidification | Passive air humidification in the chamber |
| Oxygen system | Oxygen supply with cylinders included |
| Hood | Double-walled acrylic-glass with side door (mattress tray removal) |
| AC power supply | 220-230 V / 50 Hz or 110-120 V / 50-60 Hz (400 VA) |
| DC power supply | 12 V / 10 A or 24 V / 6 A |
| Rechargeable internal battery | 90 minutes of continuous operation (36 °C set, ambient 15 °C) |
| Dimensions (main body) | 1090 x 655 x 760 mm |
| Dimensions (cabinet) | 1380 x 610 x 400 mm |
| Weight | 80 kg (main body) + 53 kg (cabinet) |
| Unit height | Adjustable |
| Indications | Set temperature, air temperature, baby temperature and battery capacity, displayed separately |
Safety alarms
| Alarm | Description |
| Power failure | Power cut warning |
| Overtemperature | Overtemperature protection |
| Sensor deviation | Temperature sensor deviation |
| Fan motor failure | Forced air flow failure |
| Sensor failure | Temperature sensor failure indication |
Standard equipment and optional configuration
| Standard equipment | Main body (hood, controller, mattress tray, internal battery and observation lamp), oxygen cylinders and supply system, skin temperature sensor, IV pole, mattress and adjustable stand |
| Optional configuration | Temperature override mode >37 °C and disposable skin temperature sensor; ambulance trolley available on request |
4. Equivalences and Cross-References
The BabyGuard I-1125 belongs to the Dixion family of closed intensive care incubators within the Cahnos catalogue, alongside references I-1103 (passive humidification), I-1107 (optional humidity and oxygen concentration control) and I-1120 (independent servo-control of air temperature, skin and humidity). As open resuscitation systems, the same series offers the radiant warmers W-1140, W-1140-02 and W-1145 for delivery rooms. In the clinical concept of neonatal transport, the I-1125 competes functionally with transport incubators from other international manufacturers such as Dräger (Isolette transport series), GE Healthcare (neonatal transport systems), Fanem or Atom Medical, being interchangeable in its role of thermal support, oxygen administration and continuity of care during transfers. Complete the equipment with the gel positioners for incubator tray available at Cahnos (ref. GBSK0022-PS and GBSK0023-PS). For budgets and tenders, the technical-commercial department of Cahnos can provide you with the detailed comparison with the reference your centre currently uses.
5. Selection tips and variants of interest
Assess the type of care your unit provides: if you carry out intra-hospital, inter-hospital or ambulance transfers of high-risk neonates, the I-1125 is the variant of the closed series with internal battery, AC/DC power supply and optional trolley for ambulance, ensuring thermal stability throughout the journey. If your centre cares for stable neonates requiring a closed environment with maximum servo-control of humidity and oxygen in the NICU, consider the I-1120 or the I-1107; and if you require open resuscitation in delivery rooms with rapid access to the newborn, consider the W-1140, W-1140-02 or W-1145 systems. Always check the electrical compatibility of the centre (220-230 V / 50 Hz or 110-120 V / 50-60 Hz), the condition of the rechargeable internal battery —with a typical service life of about 3 years— and the space available in the ambulance or destination room. For institutional purchases, supply times or tender requirements, consult your technical-commercial advisor at Cahnos.





Reviews