Description
1. Introduction
The scalpel blade No. 36, manufactured in-house and distributed by Cahnos, is a detachable surgical blade made of carbon steel, sterile and single-use, designed for clean and controlled incisions in organic tissues. It is the reference tool for professionals who value a fine edge, immediate disposal after each procedure and the complete absence of cross-contamination risk.
Each blade is supplied in an individual peelable package with VCI anti-corrosion inner paper, ready to be fitted in seconds to the compatible scalpel handle. Its presentation in a box of 100 units makes it the usual choice for hospitals, clinics, emergency services, autopsy rooms, pathological anatomy laboratories and primary care centres.
2. Technical explanation
Detachable scalpel blades are identified by the classic Bard-Parker numbering, common to the main manufacturers on the market. The fit between blade and handle is standardised according to standard ISO 7740:1985, which guarantees a firm and consistent attachment with the corresponding handle. Blade No. 36 belongs to the group of large blades (18-36) and is mounted on handles with fitment No. 4 (handles 4 and 4L in the standard market numbering).
Carbon steel, conforming to hardness grade BS 2982:1992, achieves a very fine sharpening and a long-lasting edge, although it requires anti-corrosion protection, resolved by the VCI inner paper of the individual peelable package. The product is supplied sterile by gamma radiation (minimum 2.5 Mrad), is single-use, has the size marked on each blade and has a shelf life of 5 years from manufacture. Profile No. 36 is used as a large curved-edge blade in general surgery, incisions in organic tissues and incision and puncture procedures.
3. Technical Specifications
| Parameter | Value |
| Manufacture | Own (Cahnos) |
| Blade number | 36 |
| Material | Carbon steel (grade according to BS 2982:1992) |
| Type | Disposable, detachable, single-use scalpel blade |
| Sterilisation | Gamma radiation (minimum 2.5 Mrad) |
| Presentation | Individual peelable packaging with inner VCI protective paper |
| Units | Box of 100 units |
| Fit (ISO 7740:1985) | Fitment No. 4 — handles 4 and 4L (standard numbering) |
| Expiry | 5 years from date of manufacture |
| Applied standards | ISO 7740:1985 (fit), BS 2982:1992 (hardness), ISO 7153-1 (surgical steel) |
| SKU Cahnos | 0100711 |
| Classification | Sterile single-use medical device; consult the CE marking and manufacturer’s registration |
Blade and handle fit guide (ISO 7740):
| Fit | Compatible blades |
| Fitment No. 3 (handles 3, 3L, 7, 9) | Blades 9-17 in the standard reference numbering |
| Fitment No. 4 (handles 4, 4L) | Blades 18-36 in the standard reference numbering |
Packaging data (surgical blade range):
| Item | Value |
| Inner box | 8.5 x 8 x 2.7 cm |
| Units per box | 100 units |
| Outer boxes | 10 boxes x 100 units |
| Export carton | 47 x 39 x 17 cm |
4. Equivalences and Cross-References
Blade No. 36 corresponds to the standard market Bard-Parker numbering, with fit ISO 7740 in the large blade group (18-36). The following references have been compared property by property (blade number, material, sterility, fit and presentation) against official manufacturer catalogues and the FDA GUDID database:
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
| Swann-Morton (United Kingdom) | Blade No. 36, carbon steel, sterile (No. 4 Range) | 0236 (SM0236 / 00SM36); verified in GUDID (UDI-DI 05033955002367) | Comparable by category (not interchangeable 1:1) | Same blade number, carbon steel conforming to BS 2982, sterile, individual package with VCI paper and box of 100 units. Fitment No. 4 (handles 4, 4L, 4 graduated and 6B). As it is another manufacturer, verify the exact profile geometry before replacing. |
| B. Braun / Aesculap (Germany) | Scalpel Blades Fig. 36, carbon steel, sterile | BB536 (box of 100); verified in GUDID (UDI-DI 04038653020041) | Comparable by category (not interchangeable 1:1) | Blade figure 36, carbon steel, sterile and disposable, in a dispenser pack of 100 units; fitment for handle No. 4. Same indication of large blade; confirm geometry and fit tolerances with the manufacturer. |
There is no equivalence table validated by the manufacturer for reference No. 36; consult the technical department before using in a tender. As homologous families in the sector with sterile carbon steel scalpel blades, Feather (Japan), Aspen Surgical (Bard-Parker, USA) and Kai Medical (Japan) can also be cited; in all of them, the availability of size 36 and its exact dimensions must be verified in the official datasheet before any replacement. Note that the official conventional range of Feather does not publish size 36 (its catalogue goes up to No. 25 in conventional blades), so its existence should not be assumed without confirmation.
5. Selection tips and variants of interest
- Verify the fit: blade No. 36 is mounted on handles with fitment No. 4 (handles 4 and 4L) according to ISO 7740. Confirm the handle available in your department before integrating the blade into the operating theatre trolley or into kits.
- Carbon steel vs. stainless steel: carbon steel offers superior sharpening and a more durable edge, but requires the VCI protection of the package and must not be exposed to moisture. It is not interchangeable 1:1 with a stainless steel blade of the same number.
- Single use: always dispose of the blade after use and manage the sharps waste in accordance with current regulations. The size is marked on each blade to facilitate traceability.
- Cahnos variants of the same range: blades No. 18 (0100705), No. 20 (0100706), No. 21 (0100707), No. 22 (0100708), No. 23 (0100709) and No. 24 (0100710) for the same fitment No. 4; the complete disposable scalpel No. 36 (0100733) when an integrated single-use handle and blade is preferred.
- Storage: keep the box in a dry place at room temperature, respecting the expiry date of 5 years from manufacture. Check the integrity of the individual package before each use.



Reviews