Description
1. Introduction
Vagistop is a medical device designed for the prevention and treatment of vaginal bleeding in the gynaecological and obstetric setting. Its use effectively replaces traditional vaginal packing with gauze swabs, offering a safer, controlled, atraumatic and painless solution for the patient. It is the reference device in the Ginram® range from RI.MOS (pregnancy and childbirth area) and is supplied in a box of 5 units with individual packaging.
2. Technical explanation of the device
Vagistop consists of a vaginal tamponade device with a balloon that is inflated with air using a syringe connected to its valve. Once inserted, progressive inflation makes the balloon adhere completely to the vaginal walls, exerting an internal pressure of approximately 50 mmHg, which exceeds the pelvic capillary pressure (21 to 48 mmHg) and allows an effective and controlled haemostatic compressive effect.
- In gynaecology: indicated for the prevention and treatment of vaginal bleeding and postoperative haematoma after vaginal surgery or trauma. Recommended inflation volumes between 60 and 180 ml depending on vaginal size.
- In obstetrics: effective treatment of bleeding from postpartum vaginal tears and prevention of postpartum vaginal haematoma. Recommended inflation volumes from 240 to 480 ml, adjusted to vaginal size.
- Safety limit: the balloon has been tested and validated up to a maximum inflation of 500 ml of air, ensuring structural integrity under extreme conditions.
3. Technical Specifications
| Parameter | Specification |
| Reference | 720262 (official RI.MOS factory code, Ginram® range) |
| Device type | Inflatable vaginal tamponade device with air |
| Length | 46.3 cm |
| Shaft diameter | 4.5 mm (13.5 Fr) |
| Gynaecological indication | Prevention and treatment of vaginal bleeding and haematoma after vaginal surgery or trauma |
| Inflation volume in gynaecology | 60 to 180 ml depending on vaginal size |
| Obstetric indication | Treatment of bleeding from postpartum vaginal tears and prevention of postpartum vaginal haematoma |
| Inflation volume in obstetrics | 240 to 480 ml depending on vaginal size |
| Internal adhesion pressure | Approximately 50 mmHg (higher than the pelvic capillary pressure of 21 to 48 mmHg) |
| Tested inflation limit | 500 ml of air |
| Replaces | Vaginal packing with gauze swabs |
| Use | Disposable, single-use |
| Presentation | Box of 5 units, each individually packaged |
| Sterilisation | Ethylene oxide (EtO) |
| Marking | CE 0051 (notified body) |
4. Equivalences and Cross-References
The reference 720262 is the official factory code for the VAGISTOP device from RI.MOS, verified against the manufacturer’s official catalogue and its technical data sheet (length 46.3 cm, Ø 4.5 mm / 13.5 Fr, box of 5 units, EtO sterilisation and CE 0051 marking). No official equivalence table validated by the manufacturer has been found that cross-references this code 1:1 with other brands; the following alternatives are classified as comparable at category level, not interchangeable 1:1, and must be checked against each manufacturer’s technical data sheet and instructions for use before use in tenders:
| Manufacturer / Brand | Model / Family | Reference / SKU | Equivalence level | Technical differences |
| RI.MOS (Ginram®) | VAGISTOP | 720262 | Own verified reference (manufacturer’s official catalogue) | Vaginal tamponade device with air, 46.3 cm and 13.5 Fr |
| Cook Medical | Bakri® Postpartum Balloon | J-SOSR-100500 (G24237) | Comparable by category | Silicone balloon for postpartum uterine tamponade, inflated with saline solution up to 500 ml; uterine approach, not vaginal tamponade |
| Utah Medical Products | BT-Cath® Balloon Tamponade Catheter | BTC-ESY / BTC1000 | Comparable by category | Double-lumen catheter with intrauterine drainage and EasyFill™ system; indicated for uterine haemorrhage, not for vaginal tamponade |
| Pergo | Postpartum Balloon (uterine tamponade) | 05TD0206 (Cahnos catalogue) | Comparable by category | 100% silicone pear-shaped with capacity up to 5000 ml of saline; high-volume uterine tamponade, different indication and design |
| RI.MOS (Ginram®) | METRASTROP | 720264 (Cahnos catalogue) | Comparable by category | Double-balloon catheter (air and saline) for postpartum uterine haemorrhage; same manufacturer family but uterine indication |
There is no equivalence table validated by the manufacturer for this reference; consult the technical department at Cahnos before using this reference as an equivalence in specifications or tenders.
5. Selection tips and variants of interest
- Volume adjustment: start inflation progressively and respect the manufacturer’s ranges (60–180 ml in gynaecology and 240–480 ml in obstetrics) according to vaginal size, without exceeding the tested limit of 500 ml.
- Haemostasis confirmation: inflate until the balloon adheres completely to the vaginal walls; the internal pressure of approximately 50 mmHg exceeds the pelvic capillary pressure and ensures effective compression of the bleeding site.
- Contraindication for shared use: it is a single-use device; maintain permanent stock in delivery rooms, labour wards and gynaecological surgery operating theatres to ensure immediate availability.
- Variants in the Cahnos catalogue: for postpartum uterine haemorrhage, the METRASTROP catheter (Ref. 720264, double balloon) and the silicone Postpartum Balloon (Ref. 05TD0206, up to 5000 ml) are available; for the exclusively vaginal approach, Vagistop (Ref. 720262) is the specific option.





Reviews