Description
1. Introduction
The Rapid Gonorrhoea Detection Test in cassette (ref. IGO-502) is an in vitro diagnostic device intended for the qualitative detection of the antigen of Neisseria gonorrhoeae, the causative agent of gonorrhoea, in cervical (female) and urethral (male) swab samples. Manufactured by ALLTEST Biotech within its JusChek® range, it provides a reliable result in just 10 minutes, with visual reading and simple interpretation, without the need for laboratory instrumentation. It is the ideal solution for STI clinics, gynaecology and urology services, health centres, clinical laboratories and screening programmes for sexually transmitted infections. Gonorrhoea is one of the most prevalent bacterial sexually transmitted infections and, in many cases, is asymptomatic (up to 80% in women and 10% in men), so having a rapid screening at the point of care facilitates early diagnosis and early clinical intervention.
2. Technical explanation and procedure
The test is a lateral flow chromatographic immunoassay that uses highly specific antibodies against the antigen of Neisseria gonorrhoeae, immobilised on the test region (T) of the cassette membrane. The sample obtained with the swab is processed with the reagent solutions included in the blood collection tube: 5 drops of reagent 1 (NaOH 0.15 M, approx. 300 µl) are added, the swab is inserted and rotated 15 times, and left to stand for 2 minutes; then 4 drops of reagent 2 (HCl 0.2 M, approx. 200 µl) are added and the swab is rotated again 15 times until the solution clears. The extracted antigen reacts with the antibody conjugate and the mixture migrates along the membrane, generating a coloured line in the T zone if the result is positive. The appearance of the control line (C) confirms the validity of the assay (adequate sample volume and correct migration). For reading, 3 drops of the extracted solution (approx. 100 µl) are placed in the sample well and the result is interpreted at 10 minutes (do not interpret after 30 minutes). The procedure requires no additional equipment, which speeds up clinical decision-making at the point of care.
3. Technical Specifications
3.1. General characteristics
| Parameter | Value |
|---|---|
| Product reference | IGO-502 |
| Manufacturer | ALLTEST Biotech – JusChek® range |
| Method | Lateral flow chromatographic immunoassay |
| Analyte detected | Neisseria gonorrhoeae antigen |
| Sample type | Cervical swab (female) and urethral swab (male) |
| Format | Visual reading cassette |
| Time to result | 10 minutes (do not interpret after 30 min) |
| Certification | Class C In Vitro Medical Device according to IVDR 2017/746 (certificate available on request) |
| Storage temperature | 2 – 30 °C (do not freeze) |
| Shelf life | 24 months (2 years) from the date of manufacture |
| Use | For professional in vitro diagnostic use only |
3.2. Clinical data (reference method: culture
| Sample type | Sensitivity | Specificity | Accuracy |
|---|---|---|---|
| Cervical smear (female) | 94.4 % | 96.9 % | 95.9 % |
| Urethral smear (male) | 91.6 % | 97.1 % | 94.3 % |
Clinical study conducted with patient samples from STI clinics. Intra-assay and inter-assay precision verified with three different lots, with correct identification in more than 99% of determinations.
3.3. Box contents
| Component | Quantity |
|---|---|
| Test cassettes | 20 |
| Blood collection tubes with dropper cap | 20 |
| Reagent buffer solution 1 (0.15 M NaOH) | 1 |
| Reagent buffer solution 2 (0.2 M HCl) | 1 |
| Workstation | 1 |
| Sterile swabs for cervical collection in women | 20 |
| Instructions for use | 1 |
Required material not included: sterile male urethral swabs and timer.
3.4. Packaging
| Detail | Value |
|---|---|
| Units | 20 tests per box |
| Boxes per carton | 40 boxes (800 tests per carton) |
| Carton dimensions | 55 × 39.5 × 46 cm |
| Weight | 13 kg per box |
3.5. Additional data of interest
| Parameter | Value |
|---|---|
| Assay volume | 5 drops reagent 1 (≈300 µl) + 4 drops reagent 2 (≈200 µl); 3 drops of extract (≈100 µl) in the well |
| Swab preservation after collection | 4–6 h at room temperature (15–30 °C) or 24 h refrigerated (2–8 °C) in a dry transport tube; do not freeze |
| Cross-reactions | Evaluated with 10⁷ CFU/test against Chlamydia trachomatis, Gardnerella vaginalis, Candida albicans, Staphylococcus aureus, Klebsiella pneumoniae, among others; negative result |
| Expected reactivity | Detects N. gonorrhoeae antigen from viable and non-viable organisms; does not allow assessment of therapeutic success or failure |
4. Equivalences and Cross-References
This product belongs to the family of rapid STI screening tests from ALLTEST Biotech (JusChek® range). The same gonorrhoea antigen detection assay on swab is marketed under the manufacturer’s reference IGO-502, also available in white-label format ACRO IGO-502; processed samples can additionally be read with the LF Portable Reader Plus reader from the same company, allowing objective reading and recording of results. There is also a variant intended for self-diagnosis (IGO-N502H) and a combined presentation of gonorrhoea and chlamydia in a single cassette. In the Cahnos catalogue, this test is complemented by the rest of the STI screening line: the Rapid Chlamydia Detection Test in cassette (ref. ICH-502), the Rapid Syphilis Test JusChek® IgG/IgM (ref. ISY-N402) and the Rapid HIV 1/2 Test Kit (ref. IHI-402), all manufactured by ALLTEST Biotech and intended for professional diagnosis of sexually transmitted infections.
5. Selection tips and variants of interest
Professional use: indicated exclusively for healthcare centres, clinical staff and laboratories; it is not a self-test for home use. Sample type: in women, cervical collection is used with the included sterile swab (avoid collection during the menstrual period and remove excess mucus beforehand with a swab); in men, urethral collection requires a sterile swab not included, and the patient is recommended not to urinate for at least 1 hour before collection. Quality control: check the control line (C) on each determination; if it does not appear, the result is invalid and the test must be repeated. Precautions: excess blood on the swab may cause false positives; store the cassettes in their original sealed packaging at 2–30 °C, protected from humidity and direct sunlight, without freezing and within the expiry date. Volume of use: the box of 20 tests is the standard presentation; for high-volume screening programmes, the full carton (800 tests) offers the best cost per test ratio. Combined STI screening: for a complete battery in a single campaign, combine this test with references ICH-502 (chlamydia), ISY-N402 (syphilis) and IHI-402 (HIV 1/2).





Reviews