Description
1. Product Presentation
The disposable electrosurgical unit Ref. F4798 (SKU 0565104) is a monopolar single-use electrosurgical pencil, ready to use in each procedure. It incorporates blade F4046 of stainless steel, a 320 cm cable, a 2.38 mm (3/32″) shaft and universal 3-pin Valleylab-type connector, which guarantees compatibility with virtually all electrosurgical generators on the market. Its presentation in a box of 50 units optimises logistics, stock control and cost per procedure in hospitals, clinics and ambulatory surgery centres, with individual heat-sealed packaging that ensures sterility until opened in the operating theatre.
2. Technical explanation of the equipment and procedure
The electrosurgical unit acts as active electrode of a monopolar electrosurgery system: the pencil connects to the generator (ESU) via the tripolar Valleylab-type connector and the circuit is closed through the dispersive (neutral) plate on the patient. Its manual double push button control allows activating the cut (CUT) and coagulation (COAG) functions without the need for a foot pedal, giving the surgeon ergonomic and direct control during dissection and haemostasis.
The F4046 blade concentrates high-frequency energy on a standard stainless steel edge that allows precise incisions and coagulation of small vessels with minimal carbonisation of adjacent tissue. The 2.38 mm (3/32″) shaft ensures rigidity and stability during work, while the 320 cm cable provides ample freedom of movement around the operating table. Being a sterile single-use device, the risk of cross-contamination and the costs of cleaning and re-sterilisation are eliminated.
3. Technical Specifications
Technical features of the electrosurgical unit F4798:
| Specification | Value |
| Product type | Electrosurgical unit (electrosurgical pencil) disposable with push button |
| Manufacturer reference | F4798 |
| SKU Cahnos | 0565104 |
| Electrode | Blade F4046 (stainless steel) |
| Shaft diameter | 2.38 mm (3/32″) |
| Cable length | 320 cm |
| Connector | Valleylab (3 pins) |
| Control | Double push button (cut and coagulation) |
| Sterility | Sterile, single-use |
| Material | High-resistance plastic · latex-free |
| Presentation | Box of 50 units with individual packaging |
Variants of the same family (same push-button pencil, 2.38 mm shaft and Valleylab connector):
| Reference | Electrode | Cable | Connector |
| F4797 | Blade F4046 | 320 cm | Valleylab |
| F4798/5 | Blade F4046 | 500 cm | Valleylab |
| F4797/A | Needle F4048 | 320 cm | Valleylab |
| F4797/L | Blade F4050 | 320 cm | Valleylab |
| F4797/P | Sphere F4044 | 320 cm | Valleylab |
| F4797/PL | Sphere F4067 | 320 cm | Valleylab |
| F4797/SE | Blade F4046/F4050 with smoke evacuator | 300 cm | Valleylab (box of 10 units) |
4. Equivalences and Cross-References
No official cross-reference table from the manufacturer has been found in public sources that validates an exact 1:1 equivalence of the F4798 pencil with pencils from other brands. The following comparisons are functional comparisons verified against public documentation from each cited manufacturer (official catalogues, technical data sheets, GUDID and authorised distribution) and should not be treated as 1:1 interchangeable without validation from the technical department.
| Manufacturer / Brand | Family / Model | Reference Code / SKU | Equivalence Level | Technical Differences / Remarks |
| Medtronic / Valleylab (Covidien) | Valleylab™ Electrosurgical Pencil with push button, stainless steel hex-locking blade, 10 ft (≈3 m) cable | E2516 (box of 50) | Comparable by category (not interchangeable 1:1) | Disposable push-button pencil with removable blade and standard tripolar connector; accepts 3/32″ (≈2.4 mm) shaft electrodes. The exact reference and blade designation (F4046) differ. Code and attributes verified in Medtronic documentation, GUDID and authorised distribution. |
| Erbe | Disposable electrosurgical pencil with 2 buttons, 3 m cable, with spatula electrode and sheath | 20190-107 (box of 10) | Comparable by category (not interchangeable 1:1) | Disposable 2-button pencil for VIO/ICC/ACC units and non-Erbe equipment; international 3-pin connector. Spatula-type tip, not F4046 blade; presentation in box of 10 units. Verified in official Erbe catalogue and authorised distribution. |
| ConMed | GoldLine® Push Button, 1″ (2.54 cm) blade, 10 ft cable and sheath | 130309A (box of 40) | Comparable by category (not interchangeable 1:1) | Disposable push-button pencil with preloaded blade, standard 3-pin connector and hexagonal locking; the blade geometry/designation and presentation format (box of 40 units) differ. Code and attributes verified in ConMed catalogue and authorised distribution. |
| Bovie | Disposable Electrosurgical Push Button Pencil with blade | ESP1 (box of 50) | Comparable by category (not interchangeable 1:1) | Disposable push-button pencil with blade and standard 3-pin connector, individual sterile packaging, box of 50; the brand and exact blade reference differ. Verified in Bovie authorised distribution catalogues. |
There is no equivalence table validated by the manufacturer; consult the technical department before use in tendering. Before any substitution, check field by field: electrode geometry and length, shaft diameter (3/32″ ≈ 2.4 mm), connector type (3-pin Valleylab versus other systems), cable length, presentation and regulatory classification of the actual product.
5. Selection tips and variants of interest
The choice of electrode determines the effect on tissue: the F4046 blade of this pencil is the versatile option for cutting and coagulation in open surgery; the F4048 needle (F4797/A) provides maximum precision for pinpoint incisions and haemostasis of small bleeding points; the F4050 blade (F4797/L) offers very similar performance; and the F4044/F4067 ball electrodes (F4797/P, F4797/PL) are indicated for coagulation of large surfaces. If more reach around the operating table is needed, request the F4798/5 or F4797/5 variant (F4046 blade with 500 cm cable); and for procedures with surgical smoke generation, the F4797/SE version incorporates a smoke evacuator.
Always check that your electrosurgical generator has a 3-pin Valleylab-type connection (most Medtronic/Covidien and compatible equipment): if you use an Erbe generator, request the direct connection variant F4797/ERB or the adapters F4832 and F4833 for Erbe equipment, and F4831 for Berchtold and Martin. As complementary accessories from the electrosurgery catalogue, the sterile sheath F4545 (holster) and the sterile tip cleaner F7520 are recommended to maintain electrode performance during the procedure. This is a single-use product: do not re-sterilise or reuse.





Reviews