We use our own and third-party cookies to analyse the use of the website and show you advertising related to your preferences, based on a profile built from your browsing habits (for example, pages visited). More information

Cahnos.com Compare invoice

Disposable foam electrosurgical plate for adults 202 x 101 mm (118 cm²), with REM system, without cable and TAB connection – Box of 25 units (Ref. F7820)

15,00 €

Disposable electrosurgical foam plate for adults, with active surface of 118 cm², REM system and TAB (tab) connection. Without cable: use the F7902/F7903 connection cables according to your electrosurgical unit. Box of 25 units with individual packaging. Class IIb medical device, latex-free, low impedance (<13 Ω) and high safety against electrical burns.

International shipping. Import duties and taxes are payable by the recipient.

Datasheet (opens in a new tab)

Reference
F7820

Description

The electrosurgical plate F7820 is the reference disposable neutral electrode for monopolar electrosurgery procedures in adult patients. Its foam design with conductive hydrogel offers excellent adhesion, uniform current distribution and effective protection against burns in the return area. It is supplied in a box of 25 units with individual packaging, designed for operating theatres, major ambulatory surgery, minor surgery clinics and sterilisation departments that require continuous stock.

Technical explanation

In monopolar electrosurgery, the high-frequency current generated by the electrosurgical unit exits through the active electrode (handle/pencil), passes through the patient’s tissue and must close the circuit through a neutral electrode or return plate. This plate incorporates the REM system (Return Electrode Monitoring): the conductive surface is divided into two halves, allowing the electrosurgical generator to continuously monitor the quality of contact between the plate and the skin. If the contact impedance increases due to partial detachment or drying of the gel, the system stops the energy emission and prevents the risk of burns.

As it is a plate without cable and with TAB (tab) connection, the professional must use a reusable adapter cable (F7902 universal 6.3 mm, F7903 Valleylab type, F7904 for ERBE, etc.) that connects the plate’s tab to the electrosurgical unit. This configuration reduces the cost per procedure by reusing the cable and facilitates warehouse management.

Technical specifications

Product data

Reference F7820
Measurements 202 x 101 mm
Active surface 118 cm²
Support Foam (polyethylene foam)
Conductive adhesive Hydrogel
REM system Yes – divided conductive surface
Cable Not included (tab connection)
Connection TAB (tab)
Presentation Box of 25 units with individual packaging
Patient weight Adults, ≥ 15 kg
Impedance < 13.0 Ω
Max. operating temperature < 3.5 °C @ 700 mA
Latex-free Yes
Classification Class IIb according to Directive 93/42/EEC

Adult range of the foam family with REM

Reference Dimensions Support Active surface REM Cable Connection
F7820 202 x 101 mm Foam 118 cm² Yes No TAB (tab)
F7820W/V 202 x 101 mm Foam 118 cm² Yes 300 cm Valleylab
F7820W/6.3 202 x 101 mm Foam 118 cm² Yes 300 cm Bipolar Ø 6.3 mm

The three references share the same hydrogel and foam base, the same active surface of 118 cm² and the same REM system. The difference lies exclusively in the connection system to the generator.

Equivalences and cross-references

There is no equivalence table validated by the manufacturer; consult the technical department before use in tenders. After verifying in official catalogues and regulatory databases (GUDID/EUDAMED), no product from another brand has been found that matches all verifiable attributes (dimensions 202 x 101 mm, active surface 118 cm², foam support, hydrogel, REM system, TAB connection and without cable).

Homologous brands and families that must be evaluated with the technical data sheet in hand before any substitution:

  • B.Braun / AESCULAP – range of disposable divided neutral plates (Patient Plates).
  • Medtronic-Covidien / Valleylab – PolyHesive cordless REM series (e.g. E7508/E7509): they require the generator’s REM system and a specific forceps; their support and geometry differ from the foam plate.
  • GIMA – disposable bipartite REM plates for electrosurgical units.
  • Taktvoll – Patient Return Electrode GB900, split, without cable, adult.
  • Other manufacturers of disposable neutral electrodes present on MedicalExpo (EMED, Promed, Richmedi, Medmount, Upspring Medical) with adult bipolar models without cable.

In all cases, a field-by-field check must be carried out: exact dimensions, active surface in cm², support and type of hydrogel, REM system, physical TAB connector and compatibility with the installed electrosurgical generator. Do not assume equivalence based solely on the generic name of the product.

Selection tips and variants of interest

  • Check the electrosurgical unit connector: this cableless F7820 plate needs an adapter cable (F7902 universal 6.3 mm, F7903 Valleylab type, F7904 for ERBE) that is reused between patients.
  • If you prefer a plate with integrated cable, choose the F7820W/V (Valleylab connector) or the F7820W/6.3 (bipolar connector Ø 6.3 mm).
  • For paediatric and neonatal patients there are versions F7820P (148 x 90 mm, 72 cm²) and F7820N (88 x 74 mm, 36 cm²) with the same REM technology and TAB connection.
  • If the generator does not have REM control, the alternative without REM in the same size is the F7805 (202 x 101 mm, 132 cm²), TAB connection and without cable.
  • Always apply on dry, intact, hairless skin, on a well-irrigated muscle area and away from pacemakers, metal implants or prominent bone areas. Do not reuse or cut the plate.
  • For institutional volumes, the box of 25 units allows optimal stock management in sterilisation departments and operating theatre stores.

Additional information

Material

Foam (polyethylene foam) with hydrogel adhesive

Measurements

202 x 101 mm

Active surface

118 cm²

REM system

Yes – divided conductive surface

Connection

TAB (tab)

Cable

Without cable

Presentation

Box of 25 units with individual packaging

Patient use

Adults, weight ≥ 15 kg

Impedance

< 13.0 Ω

Max service temperature

< 3.5 °C @ 700 mA

Regulatory classification

Class IIb (Directive 93/42/EEC)

Manufacturer reference

F7820

Latex

Latex-free

Reviews

There are no reviews yet.

Be the first to review “Disposable foam electrosurgical plate for adults 202 x 101 mm (118 cm²), with REM system, without cable and TAB connection – Box of 25 units (Ref. F7820)”

Your email address will not be published. Required fields are marked *