Description
1. Product Presentation
We offer you the reusable electrosurgery electrode Ref. F4068/4 from FIAB, the 4 mm Ø ball active tip that your operating theatre needs to achieve reliable, reproducible and precise coagulation in every procedure. Its 4 mm Ø ball concentrates high-frequency energy on a rounded, compact surface, ideal for working on soft tissue, bleeding beds and small vessels with control, less carbonisation and reduced lateral thermal damage compared with needle or blade tips.
Being reusable and autoclavable, the same electrode can be used for multiple sterilisation cycles with consistent performance, significantly reducing the cost per procedure in hospitals, private clinics, minor surgery practices and ambulatory surgery centres. It is supplied in a box of 5 units, the ideal replenishment format for equipping the electrosurgery cart and keeping the operating theatre stock always available.
2. Technical explanation of the equipment and procedure
In a monopolar electrosurgery system, the electrosurgical generator (ESU) sends high-frequency current through the pencil (electrosurgical unit) to the active electrode, which concentrates the energy at the point of contact with the tissue; the circuit is closed through the neutral (dispersive) plate placed on the patient. Depending on the tip geometry, the predominant effect is cutting, coagulation or mixed.
The 4 mm Ø ball tip is designed for contact coagulation, desiccation and controlled fulguration: its rounded shape distributes the current over a contained surface, achieving effective and precise haemostasis with very reduced lateral thermal damage. Its 52 mm length offers comfortable reach for shallow surgical fields and precision manoeuvres at the end of the procedure.
This electrode mounts on a 4 mm Ø shaft, the specific configuration for pencils and handpieces with 4 mm receptacle electrode holders from the FIAB range. Its stainless steel construction allows sterilisation and reuse over numerous cycles, provided the sterilisation department protocols are followed.
3. Technical Specifications
| Characteristic | Value |
|---|---|
| Reference FIAB | F4068/4 |
| Electrode size | Ø 4 mm |
| Packaging | Box of 5 individual units |
| Units per box (web) | 5 |
| Characteristic | Specification |
|---|---|
| Brand / manufacturer | FIAB SpA (Vicchio, Florence – Italy) |
| Product type | Reusable stainless steel monopolar electrosurgery electrode |
| Tip shape | Sphere (ball / sphere) |
| Ball diameter | Ø 4 mm |
| Electrode length | 52 mm |
| Shaft diameter | 4 mm |
| Material | Stainless steel |
| Type of use | Reusable · autoclavable |
| Compatibility | Pencils and electrode holders with 4 mm shaft receptacle |
| Presentation | Box of 5 units (non-sterile) |
4. Equivalences and Cross-References
The reference F4068/4 corresponds to the official catalogue of the manufacturer FIAB (reusable electrode with 4 mm shaft, 52 mm length and 4 mm Ø ball tip, box of 5 pieces in individual packaging) and Cahnos markets it without altering characters. No table of equivalences validated by the manufacturer that cross-references this exact configuration (4 mm Ø ball + 52 mm + 4 mm shaft + reusable) with 1:1 codes from other brands has been found in public sources, regulatory databases or official catalogues. Before any substitution, check field by field: ball diameter (4 mm Ø), length (52 mm), shaft diameter (4 mm) and compatibility with your pencil or electrode holder; consult the technical department of Cahnos before use in tenders.
5. Selection tips and variants of interest
Choose the tip geometry according to the desired effect and the tissue to be treated: the 4 mm Ø ball of this electrode is the contact coagulation and medium-surface haemostasis option, ideal for general, dermatological and minor surgery. If your pencil has a universal 2.38 mm (3/32″) receptacle, the same reusable FIAB family offers ball tips with that shaft (e.g., references F4068 and F4064), as well as the longer-reach variants of the family. Always check the shaft diameter against the receptacle of your handle before placing the order.
As a reusable product, follow your centre’s internal cleaning, disinfection and sterilisation protocols and discard the electrode at any sign of deterioration of the tip or shaft. As complementary accessories, the sterile tip cleaner (Tip Cleaner F7520) and the sterile pencil sheath (F4545) are recommended. For institutional volumes, hospital supply agreements or validation of equivalents from other manufacturers, contact the technical-commercial team of Cahnos.





Reviews