Description
K15 Touchscreen multiparameter monitor: complete clinical monitoring in a single device
The multiparameter monitor K15 Touchscreen (2 IBP + EtCO2 configuration) is the bedside solution for hospitals, clinics and healthcare centres that require continuous, accurate, real-time monitoring of the patient’s vital signs. Designed for monitoring adult, paediatric and neonatal patients, it integrates in a single unit the functions of ECG, heart rate (HR), non-invasive blood pressure (NIBP), oxygen saturation (SpO2), respiration (RESP) and temperature (TEMP), together with invasive blood pressure (IBP) measurement and capnography (EtCO2).
Its 15-inch TFT touchscreen with a Windows-style interface (9 waveforms on screen) and a 3-level visual and audible alarm system ensures simple, ergonomic and intuitive operation for healthcare staff, with ample data storage and the possibility of software transfer and updates via USB.
How it works and what it offers technically
The K15 records and displays essential physiological parameters in real time, allowing a complete analysis of the patient’s clinical condition. In this configuration it adds the invasive blood pressure (IBP) module with 2 channels using strain-gauge transducers and EtCO2 capnography by infrared method (sidestream or mainstream mode), for advanced haemodynamic and ventilatory monitoring.
- Ergonomic design with intuitive user interface and touchscreen.
- USB connector for data transfer and software update.
- 9 waveforms on screen.
- Windows-style display, easy to operate.
- Visual and audible alarm for physiological and technical alarms.
- Large data storage capacity.
- Suitable for use with original Nellcor sensors.
- Protection against defibrillator discharge, resistance to electrosurgical unit interference, detection and inhibition of pacemaker beats.
Multilingual internal software: GB, DE, FR, PT, ES, IT, PL, TR, RU. PC software in GB, IT, ES, compatible with Windows XP, 7, 8, 10 and 11.
Technical specifications
ECG
| Parameter | Specification |
|---|---|
| Dynamic input range | ±(0.5~5 mVp) |
| Differential input impedance | 10 MΩ |
| Band width | 0.05~150 Hz (Diagnostic), 0.5~40 Hz (Monitoring), 1~20 Hz (Operation) |
| CMRR | 90 dB (Diagnostics) / 105 dB (Monitoring and Operation) |
| Sensitivity selection | ×1/4, ×1/2, ×1, ×2, ×4 and Automatic |
| Sweep speed | 6.25 mm/s, 12.5 mm/s, 25 mm/s, 50 mm/s |
| HR measurement range | 15~350 bpm ±1% or ±2 bpm (whichever is greater) |
| Functions | Pacemaker beat detection and rejection; 5-lead ECG |
RESP (Respiration)
| Parameter | Specification |
|---|---|
| Measurement range | 0~120 rpm ±5% or ±2 rpm (whichever is greater) |
TEMP (Temperature)
| Parameter | Specification |
|---|---|
| Measurement range | 21.0~50.0 °C ±0.2 °C (from 25~45 °C) |
NIBP (Non-invasive blood pressure)
| Parameter | Specification |
|---|---|
| Technique | Oscillometric method |
| Typical measurement time | <30 seconds (adult cuff) |
| SYS range | 40~275 mmHg (Adult) / 40~200 mmHg (Paediatric) / 40~135 mmHg (Neonate) |
| DIA range | 10~210 mmHg (Adult) / 10~150 mmHg (Paediatric) / 10~95 mmHg (Neonate) |
| MAP range | 20~230 mmHg (Adult) / 20~165 mmHg (Paediatric) / 20~110 mmHg (Neonate) |
| Measurement modes | Manual, Automatic, STAT, Multi-cycle |
| Accuracy | Mean difference ±5 mmHg; standard deviation 8 mmHg |
| Automatic intervals | 1-480 min |
SpO2 (Oxygen saturation)
| Parameter | Specification |
|---|---|
| Technique | Dual-wavelength optical method |
| Measurement range | 0%~100% ±2% for SpO2 in range 70~100% |
| Pulse range (PR) | 30~250 bpm ±2 bpm or ±2% (whichever is greater) |
| Low perfusion | Up to 0.3% |
| Sensors | Compatible with original Nellcor sensors |
EtCO2 (Capnography) – requires modules 33796 or 33797
| Parameter | Specification |
|---|---|
| Technique | Infrared optical method |
| Sampling mode | Sidestream or Mainstream |
| Measurement range | 0~150 mmHg |
| Accuracy | 0~40 mmHg ±2 mmHg; 41~70 mmHg ±5%; 71~100 mmHg ±8%; 101~150 mmHg ±10% |
| Flow rate | 50 ml/min ±10 ml/min (Sidestream) |
IBP (Invasive blood pressure) – 2 channels
| Parameter | Specification |
|---|---|
| Technique | Strain gauge transducer |
| Input sensitivity | 5 µV/V/mmHg |
| Measurement range | -50~300 mmHg ±2% or ±4 mmHg (whichever is greater) |
| Measurement positions | ART, RAP, PA, LAP, CVP, ICP, AUXP1, AUXP2 |
| Calibration | Zero calibration |
Other specifications
| Parameter | Specification |
|---|---|
| Power supply | AC 100-240 V, 50/60 Hz, 60 VA |
| Battery | Integrated rechargeable lithium battery 11.1 V / 4400 mAh |
| Display | 15-inch touchscreen TFT (the model supports 12″ or 15″ TFT display) |
| Alarm system | 3-level audible and visible alarm |
| Red | Ethernet |
| Safety standards | IEC 60601-1 |
Included accessories
- Adult SpO2 catheter
- Adult NIBP cuff
- SpO2 extension cable
- ECG cable
- Skin temperature probe
- Power cable
- Rechargeable lithium battery
- Dust cover
- Disposable electrodes
- Manual: GB, FR, IT, ES
- Single-use IBP monitoring kit
- PC software: GB, IT, ES
- EtCO2 configuration
Available references
| Reference | Description | Sales unit |
|---|---|---|
| GM35310 | K15 Touchscreen 2 IBP + EtCO2 – 15″ multiparameter monitor, 5-lead ECG | 1 unit |
Product data
| Manufacturer | GIMA |
|---|---|
| Sales unit | 1 unit |
| Device type | Medical device |
| Device class | II B |
| CND | Z1203020202 |
| GMDN | 33586 |
| EAN | 8023279353105 |
Equivalences and cross-references
The K15 monitor (manufacturer reference 35310, SKU Cahnos GM35310) is part of the multiparameter monitor range. The EtCO2 capnography module requires the reference accessories/modules 33796 or 33797. The PC software is compatible with Windows XP, 7, 8, 10 and 11.
There is no official manufacturer table authorising direct 1:1 equivalences with models from other manufacturers; to replace or expand an existing monitoring fleet (e.g. Contec CMS series, Mindray MEC/iMEC, Philips IntelliVue or GE Healthcare Carescape) it is recommended to compare measured parameters, number of ECG leads, IBP channels and EtCO2 modules before deciding. Our technical-commercial department can advise you on the most suitable configuration and compatibility for your centre.
Selection tips and variants of interest
This configuration with 2 IBP channels and EtCO2 capnography is designed for departments requiring advanced haemodynamic and ventilatory monitoring: intensive care units (ICU), emergency departments, operating theatres, coronary care units and resuscitation rooms. If your department only needs basic parameters (ECG, SpO2, NIBP and temperature), you may consider simpler configurations from the same family; if you require more invasive pressure or ECG channels, please consult the expanded variants of the K15 catalogue.
To complete the equipment, it is recommended to have the EtCO2 modules (ref. 33796 or 33797), as well as the usual consumables: NIBP cuffs in the sizes for your population (adult, paediatric or neonatal), SpO2 catheters compatible with Nellcor sensors, ECG cables and a single-use IBP monitoring kit. Consult our technical-commercial team to configure the monitor to suit the needs of your department and bed capacity.







Reviews