We use our own and third-party cookies to analyse the use of the website and show you advertising related to your preferences, based on a profile built from your browsing habits (for example, pages visited). More information

Cahnos.com Compare invoice

EcoSpin III Benchtop Centrifuge – 300-4,500 rpm, Rotor 12×15 ml, RCF 2,400 g | IVDR | Ref. 60.59-9010

876,00 €

Benchtop centrifuge EcoSpin III with 12-position angle rotor for tubes up to 15 ml, speed range from 300 to 4,500 rpm and max. RCF of 2,400 g. Quiet brushless motor (<56 dB), digital time and speed control, quick-spin function, electromagnetic lid lock and IVDR certification. Ideal for separation of serum, plasma and urine in practices and clinical laboratories. Includes adapter tubes, power cable and mounting keys.

International shipping. Import duties and taxes are payable by the recipient.

Datasheet (opens in a new tab)

Reference
60.60-9010

Description

1. Introduction

The EcoSpin III benchtop centrifuge (manufacturer ref. 60.59-9010) is a compact and quiet device designed for routine separation of clinical samples such as serum, plasma and urine in medical practices, clinical analysis laboratories and small-volume care points. Its IVDR certification (Regulation EU 2017/746) guarantees compliance with European safety and suitability requirements for in vitro diagnostic use.

2. Technical explanation of the equipment

The EcoSpin III uses a brushless motor with electronic speed control in 100 rpm increments, allowing precise adjustment of the relative centrifugal force applied to each tube. The 12-position angle rotor accepts tubes up to 15 ml, and the unit incorporates a quick-spin function for rapid spins, a digital preselector for time and speed, and automatic memory of the last values used. For operator safety, the lid is locked electromagnetically during operation and is released automatically at the end of the cycle.

3. Technical Specifications

Parameter Value
Manufacturer reference 60.59-9010 (medical ECONET)
Rotor 12-position angle rotor for tubes up to 15 ml
Speed range 300 – 4,500 rpm (increments of 100 rpm)
Maximum RCF 2,400 g
Motor Brushless with electronic speed control
Control Digital: time and speed preselector; quick-spin function
Noise < 56 dB(A) at maximum speed
Power supply 110 – 240 V AC, 50/60 Hz
Dimensions (W × D × H) 255 × 245 × 140 mm
Net weight 6 kg
Safety Electromagnetic lid lock with automatic release; resistant plastic housing and hygienic membrane keypad
Certification IVDR (Regulation EU 2017/746) – device for in vitro diagnostics
Included Adapter tubes, power cable, screwdriver and Allen key

4. Equivalences and Cross-References

The code 60.60-9010 is the internal SKU of Cahnos; the official manufacturer reference (medical ECONET GmbH) is 60.59-9010, verified in the manufacturer’s official catalogue and in the product technical data sheet. There is no cross-reference equivalence table validated by medical ECONET for other brands; the following references have been checked against official sources from each manufacturer and are offered as a functional comparison:

Manufacturer / Brand Family / Model Reference Code / SKU Equivalence Level Technical Differences / Remarks
medical ECONET GmbH EcoSpin III 60.59-9010 Exact reference equivalence (same manufacturer) Same centrifuge; SKU 60.60-9010 is Cahnos’s internal code for this same unit.
DLAB Scientific DM0412 (rotor A12-10P) 9134002122 Comparable by category (not interchangeable 1:1) Same range 300–4,500 rpm and RCF 2,490 g (vs 2,400 g); rotor A12-10P accepts 12×10/7/5 ml or 8×15 ml (not 12×15 ml); larger dimensions (354×304×215 mm); similar weight (6 kg).
Hettich EBA 200 1800 Comparable by category (not interchangeable 1:1) Fixed rotor 8×15 ml (not 12 positions); higher speed (6,000 rpm / 3,461 g); IVDR certification; larger dimensions and weight (261×353×228 mm; 9 kg).

For public tender purposes, it is recommended to consult Cahnos’s technical department before replacing the EcoSpin III with any of the comparable units, since the exact reference assignment can only be confirmed against the original manufacturer.

5. Selection tips and variants of interest

  • Working volume: if the laboratory mainly processes 10–15 ml tubes, the EcoSpin III offers the greatest capacity with 12 positions; for 50 ml tubes, a higher-category unit is required (e.g. centrifuge with rotor A6-50P).
  • Centrifugal force: check that the RCF required for your protocol (e.g. serum separation at 1,800–2,000 g or plasma at 2,400 g) is within the maximum range of 2,400 g.
  • Certification: for in vitro diagnostics, verify that the unit has CE-IVDR marking; the EcoSpin III includes it.
  • Accessories: confirm the included adapters (10 ml and 5/2 ml tubes) and request additional adapters if you work with tubes of smaller diameter.
  • Quietness and space: with 255 × 245 × 140 mm and <56 dB, it is ideal for practices with limited space.

Additional information

Manufacturer

medical ECONET GmbH (Germany)

Manufacturer reference

60.59-9010

Speed range

300–4,500 rpm (increments of 100 rpm)

Maximum RCF

2,400 g

Rotor

12-position angle rotor for tubes up to 15 ml

Motor

Brushless with electronic control

Control

Digital: time, speed, quick-spin and settings memory

Power supply

110–240 V AC, 50/60 Hz

Dimensions

255 × 245 × 140 mm (W × D × H)

Net weight

6 kg

Lid lock

Electromagnetic with automatic release

Certifications

IVDR (Regulation EU 2017/746)

Includes

Adapter tubes, power cable, screwdriver and Allen key

Reviews

There are no reviews yet.

Be the first to review “EcoSpin III Benchtop Centrifuge – 300-4,500 rpm, Rotor 12×15 ml, RCF 2,400 g | IVDR | Ref. 60.59-9010”

Your email address will not be published. Required fields are marked *