Description
1. Introduction
The Cardioline ECG200L Glasgow (ref. 54204) electrocardiograph is the resting electrocardiography solution designed for private practice, cardiology consultations, health centres and functional diagnostic units. It simultaneously records the 12 standard leads and offers a very intuitive user experience thanks to its 7\” colour touch screen and dedicated quick-operation keys.
The device guides the operator step by step during acquisition: electrode placement, signal quality control, recording, printing and storage. It also incorporates the Glasgow interpretation algorithm, capable of diagnosing the anomalies commonly recognisable in the ECG (myocardial infarction, ventricular hypertrophy, ST-T changes, arrhythmias and conduction disorders or QT interval prolongation), with specific modules for adults and paediatric patients, including STEMI detection.
2. Technical explanation of the equipment
The ECG200L simultaneously acquires the 12 electrocardiographic leads (I, II, III, aVR, aVL, aVF and V1-V6) with a high-performance input stage compliant with the ANSI/AAMI IEC 60601-2-25:2011 standard. The signal is sampled at 32,000 samples per second per channel at the input stage, with an ECG resolution of 5 µV/LSB and 500 samples/s, ensuring faithful and stable reproduction of the trace.
It allows switching, with just the touch of a key, between acquisition modes automatic (12 leads), manual (3/6/12), Stat (12) and rhythm (1/3 leads), in standard or Cabrera configuration. The new rhythm ECG function allows analysing up to 3 minutes of ECG, including heart rate trend and its variability.
The device incorporates a digital linear-phase diagnostic high-pass filter, an adaptive 50/60 Hz mains interference filter and digital low-pass filters of 25/40 Hz for display and printing, as well as pacemaker detection with digital convolution filtering and defibrillation protection in accordance with AAMI/IEC standards. Recordings can be stored on the device (up to 50 ECGs) or exported via USB connection to a PC or a USB flash drive, in SCP-PDF format. It includes switchable multilingual internal software in 16 languages.
3. Technical Specifications
3.1. Product data table
| Brand | CARDIOLINE |
|---|---|
| Model | ECG200L GLASGOW (reference 54204) |
| Sales unit | 1 pc |
| Device type | Medical device |
| Device class | II A |
| CND | Z12050302 |
| GMDN | 16231 |
| EAN | 8056732655027 |
3.2. ECG acquisition
| ECG channels | 12 leads (I, II, III, aVR-L-F, V1-6) |
|---|---|
| Patient cable | Standard 15D, 10 wires |
| CMRR | >100 dB |
| Input impedance | 100 MOhm |
| Input stage sampling frequency | 32,000 samples/second/channel |
| ECG resolution | 5 µV/LSB; 500 samples/s |
| Dynamic range | ±325 mV |
| Band width | Performance equivalent to 0.05-150 Hz |
| Pacemaker detection | Hardware detection combined with digital convolution filtering |
| Filters | Digital linear-phase diagnostic high-pass filter (per 60601-2-25 2nd ed.), adaptive 50/60 Hz mains interference filter, digital low-pass filters of 25/40 Hz (display and printing only) |
| Defibrillation protection | AAMI/IEC standards |
| Input stage performance | ANSI/AAMI IEC 60601-2-25:2011 |
| Acquisition modes | Automatic (12), Manual (3/6/12), Stat (12), Rhythm (1/3) |
| Configuration | Standard or Cabrera |
| Disconnected electrode detection | Independent on all leads |
| Heart rate meter | 30-300 bpm |
| ECG measurements | All leads, average and corrected; HR; mean RR; PR interval, QRS duration, QT interval and QTc (Hodges, Bazett and Fridericia formulas); Sokolow-Lyon index (max R in V5/V6 and S in V1); P, R, T axes |
| ECG interpretation | Glasgow analysis programme for adults, paediatrics and STEMI |
| ECG storage | 50 ECG |
| Export format | SCP-PDF |
3.3. Display, printer and physical characteristics
| Display | 7\” backlit colour TFT LCD with capacitive touch panel |
|---|---|
| Screen resolution | 800×480 |
| On-screen data | 3/6/12 real-time leads |
| Thermal printer printhead | 216 mm |
| Print resolution | 8 dots/mm |
| Equipment dimensions | 413x295x80 mm |
| Weight | 4.17 kg |
3.4. Power supply and safety
| Power supply | 100-240 V, 50/60 Hz |
|---|---|
| Protection rating of the power supply | IP20 |
| Battery type | NiMH |
| Battery life | More than 500 ECGs; more than 6 hours |
| Weight-bearing time | 4 hours to 85% of capacity |
| Regulatory standard | MDD 93/42/EEC Class IIa; IEC 60601-1-2 (electromagnetic compatibility) Group 1 – Class B; IEC 60601-1 (electrical safety) |
| Protection against electric shock | IP (internally powered equipment), Class I (with external AC/DC power supply) |
| Water protection | IXP 0 |
3.5. Included accessories
- Patient cable
- Battery
- Pack of 180 sheets of paper
- Power cable
- Adapter for 4 mm pressure connector (box of 10)
- Glasgow interpretation
- Disposable tab electrodes (box of 100)
- Manual on CD (GB, FR, IT, ES, PT, DE, NL, CZ, HR, HU, PL, RO, RU)
4. Equivalences and Cross-References
Reference 54204 (SKU Cahnos GM54204) corresponds to the Cardioline ECG200L GLASGOW electrocardiograph with 7″ touch screen from the manufacturer’s catalogue. There is no manufacturer-validated equivalence table cross-referencing this reference with third parties on a 1:1 basis; within the ECG200 family itself from Cardioline, the manufacturer documents variants 54200, 54201, 54202 and 54205 with the same functional design (5″ screen in the 54200-2 series and 7″ in the 54204-5 series), while the Glasgow algorithm is available in references 54201 to 54205. On the market, equivalent equipment by category can be found (interpretive 12-lead resting ECG), such as the Cardioline ECG100S / ECG200S or the Bionet Cardio7, which can serve as a technical comparison in tender specifications, but each substitution must be validated field by field (channels, screen, print format, algorithm and connectivity) with our technical department before use in tendering.
5. Selection tips and variants of interest
Before finalising the purchase, check the print format your centre requires: this equipment prints with a thermal head and supports the most common recording formats (3, 3+1, 3+3, 6 and 12 channels in automatic mode; 3, 6 or 12 in continuous mode, in addition to the rhythm strip). 12-channel recording is only available in the 7″ screen versions (54204-5).
If your activity requires digital management of ECGs, note that the ECG EasyApp application for PC is included with references 54202 and 54205 and is optional for 54200-1 and 54204: it receives acquisitions in SCP format via USB, allows printing, modifying patient data and customising the design of rhythm strips (speed, gain, filters). Contact our technical-commercial team at Cahnos to confirm the exact reference, software options and the formalised quotation under B2B conditions.