1. Replacement battery for the Stryker Flyte system
At Cahnos we make available the rechargeable lithium-ion battery for the Stryker Flyte system (original manufacturer reference 0408-660-000), the “Power Pack Extended Life” accumulator with which Stryker equips its Flyte surgical helmet (Casque Chirurgical Flyte) and its Flyte personal protection system.
It is the spare part required by medical electronics services, operating theatre maintenance, sterilisation units and hospital purchasing departments to restore full autonomy to protection equipment with integrated airflow, with the reliability of a traceable component and at a very competitive cost compared with the official replacement via the manufacturer.
2. Technical explanation of the equipment
The Stryker Flyte is a surgical helmet and personal protection system (GMDN: Surgical helmet system; Stryker UDI registration in GUDID, primary DI 04546540591654, catalogue No. 0408660000) that integrates cap, visor, frontal light and an airflow ventilation system that helps maintain a sterile field for the surgical team. The battery is housed in the power pack (Power Pack) and powers the system without cables or dependence on the mains supply.
This battery uses lithium-ion chemistry in a 2S configuration (two cells in series) to deliver 7.4 V nominal with 5.0 Ah capacity (≈37 Wh) and incorporates integrated protection electronics against overcharge, overdischarge and overcurrent. It works with the Stryker Steri-Shield 8-bay charger (family 0408-655). When the helmet’s autonomy is noticeably reduced or the equipment does not complete the full operating theatre day, it is time to replace the battery with this reference. Installation is direct in the original pack compartment, without tools or adaptations.
3. Technical Specifications
It replaces the original reference 0408-660-000 with the same performance and compatibility as the pack supplied with the new equipment.
| Parameter | Specification |
|---|---|
| Device manufacturer | Stryker Corporation |
| Original manufacturer reference | 0408-660-000 (Power Pack, Extended Life) |
| UDI catalogue number (GUDID) | 0408660000 — primary DI 04546540591654 |
| SKU Cahnos | 88881180 |
| Battery type | Rechargeable, with integrated protection electronics |
| Technology | Lithium-ion (Li-Ion), 2S configuration |
| Nominal voltage | 7.4 V |
| Capacity | 5.0 Ah (5000 mAh) |
| Energy | 37 Wh (Stryker GUDID UDI record: 39 Wh) |
| Dimensions (length x width x height) | 79 x 41 x 43 mm |
| Approximate weight | 0.221 kg |
| Application | Stryker Flyte surgical helmet (Casque Chirurgical Flyte) and Flyte personal protection system |
| Compatible charger | Stryker Steri-Shield 8-bay, family 0408-655 (230 V version: 0408-655-001) |
| Regulatory classification | FDA GUDID Class I (GMDN: Surgical helmet system); non-implantable |
| Presentation | 1 unit |
Note: reference 0408-660-000 corresponds to a single battery pack (unit), not a batch. The catalogue data sheet states 37 Wh (5.0 Ah x 7.4 V); the Stryker UDI record in GUDID identifies the pack as “Power Pack, Extended Life 7.4V 39Wh”. Always verify the reference printed on the removed pack before placing the order.
4. Equivalences and Cross-References
Reference 0408-660-000 is the 7.4 V Power Pack Extended Life that Stryker fits to the Flyte surgical helmet and the Flyte personal protection system. The third-party alternatives cited below have been checked against battery manufacturers’ data sheets and specialist distribution and are classified as comparable by category, not 1:1 interchangeable: they share chemistry (Li-Ion), voltage (7.4 V) and target platform, but come from independent manufacturers and may vary slightly in declared capacity and dimensions.
| Manufacturer / Brand | Family / Model | Reference code / SKU | Equivalence level | Technical differences / Remarks |
|---|---|---|---|---|
| Cameron Sino | Li-Ion medical battery for Stryker Casque Chirurgical Flyte / Flyte Personal Protection System | CS-SKR408MD (EAN 4894128164777) | Comparable by category (not interchangeable 1:1) | Direct replacement, Li-Ion 7.4 V / 5200 mAh / 38.48 Wh; dimensions 79.70 x 41.50 x 44.00 mm, grey. Capacity slightly higher than the original (5200 vs 5000 mAh) and differences of tenths of a mm; consistently documented across several specialist distributors. Verify fit and BMS behaviour during the first cycle. |
On the market there are compatible products from other brands (FOUNCY, SPANN, BELLV, XSPLENDOR, KingSener, MedicalBatteryDirect, among others) that cover reference 0408-660-000 with 5200 mAh cells, but as they do not have a uniform, verifiable public reference code they are not included in the table.
There is no equivalence table validated by the manufacturer (Stryker) that authorises third-party packs for reference 0408-660-000 or that cross-references this battery with other systems; the above alternatives come from battery manufacturers and specialist distribution and are offered as comparable by category. Consult our technical department before using this reference in tenders or maintenance contracts.
5. Selection tips and variants of interest
- Identify the exact reference: check the code 0408-660-000 and the model of your equipment (Casque Chirurgical Flyte or Flyte Personal Protection System) on the label of the removed pack. Always order by SKU Cahnos 88881180 to avoid confusion.
- Do not confuse with other Stryker operating theatre batteries: Stryker cordless surgical motors and drills (e.g. 4112, 4115, 7126, 7222 or 6212) use voltages of 9.6–10.8 V with different Ni-Cd, Ni-MH or Li-Ion chemistries and are not interchangeable with the Flyte pack (7.4 V).
- Correct charger: use the Stryker Steri-Shield charger (0408-655 family; European version 0408-655-001, 230 V, 50-60 Hz, 0.9 A) and complete the charge cycle before first use. If a third-party replacement is used, a full charge-discharge cycle is recommended so that the equipment BMS recalibrates.
- Li-Ion battery maintenance: avoid repeated deep discharges, recharge it when the equipment indicates low battery and keep it in a cool, dry place (ideally at 40–60% charge) if it is to be stored.
- Plan rotation stock in PPE fleets: with two packs per unit (one in use and one charging), stoppages during long operating theatre sessions are eliminated. For mixed fleets or institutional orders, contact our technical-commercial department to confirm compatibility and a formal quotation.