Description
Dräger Oxylog 2000 Battery – Ni-Cd 7.2 V 1 Ah (Ref. 8411599)
1. Introduction
This replacement battery is designed for the transport and emergency ventilator Dräger Oxylog 2000, one of the most widely used ventilation devices in pre-hospital care, emergencies and intra-hospital transport. It replaces the original manufacturer reference 8411599 (also documented as 8411599-05), offering the same electrical behaviour: nickel-cadmium (Ni-Cd) technology, nominal voltage of 7.2 V and capacity of 1 Ah. It is the ideal replacement solution to keep your fleet of ventilators operational with a very competitive price-quality ratio, without sacrificing the reliability required for life support.
2. Technical explanation
The Dräger Oxylog 2000 is a pneumatic-electronic emergency ventilator that, when not connected to the mains or an external 12 V source, relies on its internal accumulator to power the control electronics, monitoring and alarms. This Ni-Cd battery of 7.2 V and 1 Ah (7.2 Wh) provides the necessary energy for autonomous operation during transport or power cuts, a critical factor in rescue ventilation. The nickel-cadmium chemistry provides stable current delivery and robust behaviour over a wide temperature range, characteristics required by a device that often operates in a pre-hospital environment. As stated in the official Dräger spare parts list (spare parts list 5503.160, position 8: accupack 1000 mAh), the original reference for this accumulator is 8411599, validated for versions 8412000 and 8413950 (Oxylog 2000 SW 3.0). Like any rechargeable battery, its autonomy depends on the state of charge, cycle history and ambient temperature; it is recommended to keep it charged and periodically check its residual capacity as part of the technical service maintenance plan.
3. Technical Specifications
| Parameter | Specification |
| Application | Dräger Oxylog 2000 Ventilator (and Microvent Family) |
| Original manufacturer reference | 8411599 (documented variant: 8411599-05) |
| Technology | Rechargeable Nickel-Cadmium (Ni-Cd) |
| Nominal voltage | 7.2 V |
| Nominal capacity | 1 Ah (1,000 mAh) |
| Energy | 7.2 Wh |
| Cell format | AA |
| Weight | Approximately 0.120 kg |
| Indicative dimensions | Approx. 60 × 30 × 40 mm |
| Device manufacturer | Dräger |
| SKU Cahnos | 88888150 |
4. Equivalences and Cross-References
Reference 8411599 is confirmed as the original battery for the Oxylog 2000 in the official Dräger spare parts list (5503.160). There is no equivalence table validated by Dräger itself that cross-references this reference with third-party brands; the cross references listed below come from battery manufacturers and independent spare parts distributors and have been verified against their product datasheets. Always confirm the chemistry, voltage and connector before replacing.
| Manufacturer / Brand | Family / Model | Code / SKU | Equivalence level | Technical differences / Remarks |
| Dräger (original manufacturer) | Accupack Oxylog 2000 (spare parts list 5503.160) | 8411599 / 8411599-05 | Exact reference equivalence (same manufacturer) | Original Ni-Cd battery 7.2 V 1.0 Ah; position 8 of the official spare parts list of the device. |
| Surgicall | 7.2 V battery for Oxylog 2000 / Microvent | 43.888.8150 | Comparable by category (not interchangeable 1:1) | Same Ni-Cd chemistry and voltage 7.2 V; capacity 1.1 Ah (slightly higher than 1 Ah). |
| AKKUmed | Ni-MH battery for Dräger Oxylog 2000 | 110146-N | Comparable by category (not interchangeable 1:1) | Ni-MH chemistry (not Ni-Cd); 7.2 V, 1.1 Ah, dimensions approx. 60×30×40 mm, CE declaration of the replacement manufacturer. |
| OSI Batteries | Biomedical battery Oxylog 2000 Ventilator | B11376 | Comparable by category (not interchangeable 1:1) | Ni-Cd 7.2 V 1.0 Ah; replacement that substitutes original part 8411599-05. |
| Partopia Medical | Oxylog 2000 Ventilator Battery | PM0321 | Comparable by category (not interchangeable 1:1) | Ni-Cd 7.2 V 1.0 Ah; documented aliases 8411599-05 and 5938/5938-I. |
Furthermore, on the spare parts market the cross codes BATT/110146, EE050305, OM11376, P-100AASJ/A1, PA-A111C-GC and CS-DRX200MD (the latter in Ni-MH versions of 7.2 V and 2 Ah) frequently appear associated with this same accumulator; not all have been verified against the official replacement manufacturer’s datasheet, so they should be treated as indicative references pending confirmation with the supplier before use. In case of tender, consult with the technical department of Cahnos to validate the exact equivalence with the reference your centre has in use.
5. Selection tips and variants of interest
Always check the exact version of your device: reference 8411599 is the internal battery of the Oxylog 2000; if your fleet includes other Dräger ventilators, note that each family uses its own accumulator (for example, Evita 2, Evita 4 or Babylog Carina do not share this battery). For mixed fleets with Microvent devices, confirm connector compatibility. Regarding maintenance, Ni-Cd batteries suffer from memory effect and self-discharge: avoid repeated partial cycles without full discharge and keep the accumulator charged when the device is in storage. If your centre values higher capacity, there are Ni-MH replacements of 7.2 V with more ampere-hours (up to 2 Ah), but before opting for them confirm with the technical service that the device’s charging circuit is compatible with that chemistry. For institutional purchases, supply deadlines or tender needs, contact your technical-commercial advisor at Cahnos.





Reviews
There are no reviews yet.