1. Original power for your Arjo Tempo hoist
Keep your Arjo Tempo patient hoist system operational with the original Arjo battery, reference KPA0100, which we supply under the code Cahnos 88888033. This rechargeable nickel-metal hydride (Ni-MH) accumulator restores full autonomy to the equipment for lifting, transferring and moving patients in hospitals, care homes and social health centres, without relying on the mains power supply.
It is the original replacement solution for maintenance services, biomedical engineering and purchasing departments that need a traceable, proven-quality spare part for their fleet of Arjo hoists, with the warranty that a component from the brand itself provides. Available with immediate delivery and special conditions for consortia, tenders and maintenance contracts.
2. How this battery works in the lifting system
The Arjo Tempo patient hoist powers its lifting motor via an internal rechargeable battery when the equipment is not connected to the mains. When the autonomy is noticeably reduced, the equipment warns of low battery prematurely or the lifting becomes slow, it is time to replace the accumulator with reference KPA0100.
The Ni-MH (nickel-metal hydride) chemistry offers a very reduced memory effect, contains no cadmium and supports repeated charge and discharge cycles with the hoist’s own charger. Replacement is carried out by removing the depleted battery from the compartment and installing the new one respecting the orientation and connector, without the need for tools or adaptations. After installation, a full charge is recommended before first use.
3. Technical Specifications
| Characteristic | Specification |
| Application | Arjo Tempo patient hoist system (original reference KPA0100) |
| Original manufacturer reference | Arjo KPA0100 (distribution alias SPL3021) |
| Reference Cahnos (SKU) | 88888033 |
| Technology | Ni-MH (nickel-metal hydride), rechargeable |
| Nominal voltage | 12 V |
| Capacity | 2.7 Ah (2700 mAh) |
| Energy | 32.4 Wh (12 V x 2.7 Ah) |
| Weight | Approximately 0.8 kg |
| Dimensions | 95 x 80 x 55 mm |
| Device manufacturer | Arjo |
| Rechargeable | Yes; recharge with the lift charger |
| Installation | Direct replacement in the lift compartment, without tools |
| Presentation | 1 unit |
| Recycling | Manage in accordance with WEEE regulations |
Technical note: the commercial data sheet describes this battery as Ni-MH 12 V / 2.7 Ah (32.4 Wh) with an approximate weight of 0.8 kg; however, the technical documentation from biomedical spare part manufacturers and specialised distributors identifies the pack that replaces reference KPA0100 mostly as sealed lead-acid (SLA/VRLA AGM) 24 V / 6 Ah, with dimensions of 728 x 101 x 70 mm and a weight close to 4-5 kg. Before replacing, check the label of the installed battery and the chemistry for which your equipment’s charger is set.
4. Equivalences and Cross-References
The reference of the Arjo Tempo hoist system (original battery KPA0100) has been cross-checked against biomedical battery manufacturers and specialised distribution catalogues. Arjo does not publish an official equivalence table authorising third-party batteries; the following references share the same application (patient hoists from the Arjo Tempo/Opera/Encore/Sara Plus family) and are classified as comparable by category, not interchangeable 1:1, and must be validated field by field (voltage, chemistry, dimensions and connector) against the specific equipment before replacement.
| Manufacturer / Brand | Family / Model | Reference Code / SKU | Equivalence Level | Technical Differences / Remarks |
| MSP Medical | Battery pack 24V/DC 6Ah for Arjo patient lifts (replaces KPA0100 / SPL3021) | MSP-A-AH-00200 (EAN 8719925310034) | Comparable by category (not interchangeable 1:1) | SLA/VRLA AGM 24 V / 6 Ah; valid for Arjo Tempo, Opera, Encore, Sara Plus and Maxi Move 2005; MDR compliant. Chemistry and voltage different from the pack described in this data sheet (Ni-MH 12 V / 2.7 Ah); check charger, connector and dimensions. Confirmed in the spare part manufacturer’s data sheet. |
| OSI Batteries | Biomedical battery for Arjo lifting aids (Sara Plus, Tempo, Opera, Encore) | PLB00200 | Comparable by category (not interchangeable 1:1) | SLA 24 V / 6 Ah; replaces OEM KPA0100; same application (Arjo patient lifting). Voltage, chemistry, weight and dimensions different from the pack in this data sheet; check compartment and connector. Confirmed on the manufacturer’s website. |
| BBM Battery | Biomedical pack no. 5150 for Arjo patient lift (KPA0100 / SPL3021) | 5150 | Comparable by category (not interchangeable 1:1) | SLA 24 V / 6 Ah; replaces KPA0100 and SPL3021; valid for Chorus, Tempo, Opera, Encore, Maxi Move (2005-2007) and Sara Plus. Chemistry and voltage different; dimensions 728 x 101 x 70 mm; check fit and connector. Confirmed in the spare part manufacturer’s data sheet. |
| Axcom Power Solutions | Lead battery for ArjoHuntleigh hoists Tempo, Opera, Encore, Sara Plus and Tandor | MB0101 | Comparable by category (not interchangeable 1:1) | Lead 24 V / 6 Ah; replaces OEM KPA0100 / 110325; dimensions 728 x 101 x 70 mm and approximate weight of 4.5 kg according to the spare part manufacturer’s specification. Chemistry, voltage and dimensions different from the data sheet; check charger and connector. Confirmed on the manufacturer’s website with declaration of conformity. |
There is no equivalence table validated by the manufacturer Arjo for this reference; the indicated cross-references come from biomedical battery manufacturers and specialised distributors, so it is recommended to consult the technical department before use in tenders. In specialised distribution, the numbering KPA0100 is linked to the alias SPL3021 (Arjo) and to the code PLB00200 (OSI Batteries). The 1:1 replacement must be validated field by field (voltage, chemistry of the installed battery, capacity, dimensions and connector) against the specific equipment.
5. Selection tips and variants of interest
- Confirm the reference of the installed accumulator: this original battery corresponds to Arjo reference KPA0100 for the Tempo system. Check the numbering engraved on the label of the depleted battery (KPA0100, SPL3021) before ordering.
- Verify the chemistry of your batch: if your unit is fitted with a sealed lead-acid pack (SLA/VRLA AGM), the charger is specific to that chemistry; do not install a battery of a different chemistry or voltage without prior technical validation, even if the connector may appear compatible.
- Replace upon signs of wear: clearly reduced runtime, premature low-battery warnings or slow lifting indicate that the battery has reached the end of its service life.
- Storage: if the lift is to remain unused for long periods, carry out a periodic recharge (every 2-3 months) and avoid deep discharges. At the end of its service life, dispose of the battery at a WEEE collection point; never discard it as general waste.
- Variants in our catalogue: we offer the compatible version of this same reference (Ref. 88881141, 12 V / 2.7 Ah Ni-MH), the Arjo Maxi Twin NEA0100 battery of 12 V / 2.7 Ah (Ref. 88881142 and 88888563), the Arjo Sarita KTA0102 battery of 24 V / 2.3 Ah (Ref. 88881143), the Arjo Marisa NDA0100 battery of 24 V / 4 Ah (Ref. 88881140) and the Arjo Minstrel TGT223 battery of 24 V / 5.5 Ah (Ref. 88889873). Do not interchange these references between models without technical validation.
- Institutional orders: for consortia, tenders or maintenance contracts, contact our technical-commercial team at Cahnos to confirm exact compatibility with your batch of lifts and obtain a formal quotation with special volume-based conditions.